On-line immunoaffinity column-liquid chromatography-tandem mass spectrometry method for trace analysis of diuron in wastewater treatment plant effluent sample

被引:17
作者
Zhang, Xiuli
Martens, Dieter
Kraemer, Petra M.
Kettrup, Antonius A.
Liang, Xinmiao
机构
[1] LUFA Speyer, D-67346 Speyer, Germany
[2] Chinese Acad Sci, Dalian Inst Chem Phys, Dalian 116023, Peoples R China
[3] GSF, Inst Ecol Chem, Natl Res Ctr Environm & Hlth, D-85764 Neuherberg, Germany
[4] Tech Univ Munich, Lehrstuhl Okol Chem & Umweltanalyt, D-85350 Freising Weihenstephan, Germany
关键词
immunoaffinity column; diuron; on-line; LC-MS-MS; matrix effect; wastewater;
D O I
10.1016/j.chroma.2006.08.007
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An on-line immunoaffinity column with liquid chromatography/tandem mass spectrometry (IAC-LC-MS/MS) method for the determination of diuron in water matrices was described. This method used a sol-gel immunoaffinity column (20 mm x 4 mm I.D.) for on-line sample cleanup and enrichment, a monolithic analytical column (100 mm x 4.6 mm I.D.) for separation, and a triple quadrupole mass spectrometer for quantitation. The major challenges for the on-line set-up were discussed. The optimized on-line protocol was emphasized by the fact that low limit of quantitation (LOQ) of 1.0 ng/L was achieved with only 2.5-mL sample. In addition, a satisfactory accuracy (similar to 90% of recovery) and precision (< 6% of relative standard deviation) at 50 ng/L concentration were also obtained. Due to the ability of the sol-gel immunoaffinity column to eliminate matrix effect, the on-line IAC-LC-MS/MS analysis method can reliably determine diuron in wastewater treatment plant effluent sample. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:112 / 118
页数:7
相关论文
共 45 条
[1]  
[Anonymous], 2002, Off. J. Eur. Comm, VL221, P8
[2]  
[Anonymous], OFFICIAL J EUROPEAN
[3]  
Ardrey R.E., 2003, LIQUID CHROMATOGRAPH
[4]   Trace determination of priority pesticides in water by means of high-speed on-line solid-phase extraction-liquid chromatography-tandem mass spectrometry using turbulent-flow chromatography columns for enrichment and a short monolithic column for fast liquid chromatographic separation [J].
Asperger, A ;
Efer, J ;
Koal, T ;
Engewald, W .
JOURNAL OF CHROMATOGRAPHY A, 2002, 960 (1-2) :109-119
[5]  
Bonfiglio R, 1999, RAPID COMMUN MASS SP, V13, P1175, DOI 10.1002/(SICI)1097-0231(19990630)13:12<1175::AID-RCM639>3.3.CO
[6]  
2-S
[7]   On-line coupling of immunosorbent and liquid chromatographic analysis for the selective extraction and determination of polycyclic aromatic hydrocarbons in water samples at the ng 1-1 level [J].
Bouzige, M ;
Pichon, V ;
Hennion, MC .
JOURNAL OF CHROMATOGRAPHY A, 1998, 823 (1-2) :197-210
[8]   Solid-phase extraction/high-performance liquid chromatography-electrospray mass spectrometry to determine endocrine disruptors in water samples [J].
Brossa, L ;
Pocurull, E ;
Borrull, F ;
Marcé, RM .
CHROMATOGRAPHIA, 2004, 59 (7-8) :419-423
[9]   On-line immunoaffinity extraction-coupled column capillary liquid chromatography tandem mass spectrometry: Trace analysis of LSD analogs and metabolites in human urine [J].
Cai, JY ;
Henion, J .
ANALYTICAL CHEMISTRY, 1996, 68 (01) :72-78
[10]   Rapid analysis of trace levels of antibiotic polyether ionophores in surface water by solid-phase extraction and liquid chromatography with ion trap tandem mass spectrometric detection [J].
Cha, JM ;
Yang, S ;
Carlson, KH .
JOURNAL OF CHROMATOGRAPHY A, 2005, 1065 (02) :187-198