Out-of-field photon dose following removal of the flattening filter from a medical accelerator

被引:110
作者
Kry, Stephen F. [1 ]
Vassiliev, Oleg N. [1 ]
Mohan, Radhe [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
关键词
MODULATED RADIATION-THERAPY; FREE CLINICAL ACCELERATOR; MONTE-CARLO MODEL; 6; MV; MULTILEAF COLLIMATOR; LINEAR-ACCELERATOR; 2ND CANCERS; FREE CLINAC; BEAMS; NEUTRON;
D O I
10.1088/0031-9155/55/8/003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of this paper is to determine the effect of removing the flattening filter from a linear accelerator on the out-of-field photon dose. A Monte Carlo model was used to simulate 6 MV and 18 MV photon beams from a Varian 2100 accelerator with the flattening filter in place and with it removed. The out-of-field photon doses and composition (head leakage, patient scatter and collimator scatter) were calculated from square open fields in a water tank as a function of distance from central axis, field size and depth. The out-of-field doses resulting from intensity-modulated radiation therapy to the prostate at 6 MV were also calculated, with and without the flattening filter, to sensitive organs in an anthropomorphic Rando phantom. Removal of the flattening filter reduced the out-of-field dose near the treatment field (<3 cm from the field edge) because of decreased collimator scatter. It increased the out-of-field dose at intermediate distances from the field edge (3-15 cm) because of increased patient scatter. At greater distances, the out-of-field dose was decreased because of reduced head leakage. For the clinical treatment examined, the out-of-field dose was generally reduced following removal of the flattening filter, particularly at large distances from the treatment field. Removal of the flattening filter may be advantageous by reducing the out-of-field dose to the patient. This could contribute to reducing the long-term risk of secondary malignancies. In general, however, the out-of-field dose depends on treatment and patient parameters, and a reduction may not always be achievable.
引用
收藏
页码:2155 / 2166
页数:12
相关论文
共 39 条
[11]   Calculation of effective dose from measurements of secondary neutron spectra and scattered photon dose from dynamic MLC IMRT for 6 MV, 15 MV, and 18 MV beam energies [J].
Howell, RM ;
Hertel, NE ;
Wang, ZL ;
Hutchinson, J ;
Fullerton, GD .
MEDICAL PHYSICS, 2006, 33 (02) :360-368
[12]   MEASUREMENTS OF DOSE FROM SECONDARY RADIATION OUTSIDE A TREATMENT FIELD [J].
KASE, KR ;
SVENSSON, GK ;
WOLBARST, AB ;
MARKS, MA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1983, 9 (08) :1177-1183
[13]   UNCERTAINTY ANALYSIS OF ABSORBED DOSE CALCULATIONS FROM THERMOLUMINESCENCE DOSIMETERS [J].
KIRBY, TH ;
HANSON, WF ;
JOHNSTON, DA .
MEDICAL PHYSICS, 1992, 19 (06) :1427-1433
[14]   Dosimetric characteristics of 6 and 10 MV unflattened photon beams [J].
Kragl, Gabriele ;
Wetterstedt, Sacha Af ;
Knaeusl, Barbara ;
Lind, Marten ;
McCavana, Patrick ;
Knoos, Tommy ;
McClean, Brendan ;
Georg, Dietmar .
RADIOTHERAPY AND ONCOLOGY, 2009, 93 (01) :141-146
[15]   Out-of-field photon and neutron dose equivalents from step-and-shoot intensity-modulated radiation therapy [J].
Kry, SF ;
Salehpour, M ;
Followill, DS ;
Stovall, M ;
Kuban, DA ;
White, RA ;
Rosen, II .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (04) :1204-1216
[16]   The calculated risk of fatal secondary malignancies from intensity-modulated radiation therapy [J].
Kry, SF ;
Salehpour, M ;
Followill, DS ;
Stovall, M ;
Kuban, DA ;
White, RA ;
Rosen, II .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (04) :1195-1203
[17]   Energy spectra, sources, and shielding considerations for neutrons generated by a flattening filter-free Clinac [J].
Kry, Stephen F. ;
Howell, Rebecca M. ;
Titt, Uwe ;
Salehpour, Mohammad ;
Mohan, Radhe ;
Vassiliev, Oleg N. .
MEDICAL PHYSICS, 2008, 35 (05) :1906-1911
[18]   A Monte Carlo model for out-of-field dose calculation from high-energy photon therapy [J].
Kry, Stephen F. ;
Titt, Uwe ;
Followill, David ;
Poenisch, Falk ;
Vassiliev, Oleg N. ;
White, R. Allen ;
Stovall, Marilyn ;
Salehpour, Mohammad .
MEDICAL PHYSICS, 2007, 34 (09) :3489-3499
[19]   Reduced neutron production through use of a flattening-filter-free accelerator [J].
Kry, Stephen F. ;
Titt, Uwe ;
Poenisch, Falk ;
Vassiliev, Oleg N. ;
Salehpour, Mohammad ;
Gillin, Michael ;
Mohan, Radhe .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 68 (04) :1260-1264
[20]   A Monte Carlo model for calculating out-of-field dose from a Varian 6 MV beam [J].
Kry, Stephen F. ;
Titt, Uwe ;
Ponisch, Falk ;
Followill, David ;
Vassiliev, Oleg N. ;
White, R. Allen ;
Mohan, Radhe ;
Salehpour, Mohammad .
MEDICAL PHYSICS, 2006, 33 (11) :4405-4413