Purpose. To determine if free electron laser (FEL) energy can be delivered to a small space to perform optic nerve sheath fenestration with minimal acute nerve damage. Methods. A 530 mm hollow waveguide probe was designed. Optic nerve sheath fenestration (1.0 mm diameter) was performed in 8 rabbits using either the FEL (4 eyes, 6.45mm, 10 Hz, 2 mJ) or a knife (4 eyes). Within 2 hours following surgery, the animals were perfused with aldehyde fixative. The integrity of the optic nerve and glial response at the site of fenestration were evaluated on tissue sections with H&E, andantibodies to S100 beta or GFAP. Results. Surgery using the FEL probe was found to be technically superior to the knife. The glial reaction was limited to a zone adjacent to the fenestration and was similar in both the FEL and knife incisions. Conclusions. The FEL appears capable of efficiently performing an optic nerve sheath fenestration in a small space with minimal acute damage. Both the FEL and knife incisions result in a rapid glial response at the site of fenestration even when optic nerve integrity is not compromised.