In-tube solid-phase microextraction with molecularly imprinted polymer to determine interferon alpha 2a in plasma sample by high performance liquid chromatography

被引:38
|
作者
Chaves, Andrea Rodrigues [1 ]
Costa Queiroz, Maria Eugenia [1 ]
机构
[1] Univ Sao Paulo, Dept Quim, Fac Filos Ciencias & Letras Ribeirao Preto, BR-14040903 Ribeirao Preto, SP, Brazil
基金
巴西圣保罗研究基金会;
关键词
In-tube SPME; High performance liquid chromatography; Molecularly imprinted polymers; Interferon alpha 2a; MASS-SPECTROMETRY; RECEPTOR;
D O I
10.1016/j.chroma.2013.10.018
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A molecularly imprinted sol-gel polymer (MIP) based on protein (biopharmaceutical) template with a mild template removal condition using protease was synthetized and evaluated as stationary phase for in-tube solid phase microextraction (in-tube SPME) of the interferon alpha 2a from plasma samples, followed by high performance liquid chromatography analysis with fluorescence detection (HPLC-FD). The developed MIP exhibited high selectivity for the analyte in a complex matrix. The in-tube SPME variables such as draw/eject cycles, draw/eject volume, and desorption conditions were optimized to establish the equilibrium conditions in a short time. The MIP in-tube SPME/HPLC-FD method presented linear response over a dynamic range of 8-300 ng mL(-1), with a correlation coefficient of 0.997. The interassay precision presented coefficient of variation lower than 9.2%, and accuracy values between 92% and 98%. The developed MIP performed as well as other selective interferon alpha 2a stationary phases (e.g., immunosorbent and restricted access material), with the advantage that it is robust, easy to handle and cheap to synthesize, in a addition to requiring smaller sample volume (50 mu L). Based on the analytical validation results, the proposed method (MIP in-tube SPME/HPLC-FD) can be a useful tool to determine interferon alpha 2a in plasma samples from patients receiving therapeutic dosages. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:43 / 48
页数:6
相关论文
共 50 条
  • [1] In-tube molecularly imprinted polymer solid-phase microextraction for the selective determination of propranolol
    Mullett, WM
    Martin, P
    Pawliszyn, J
    ANALYTICAL CHEMISTRY, 2001, 73 (11) : 2383 - 2389
  • [2] Automated in-tube solid-phase microextraction coupled to high-performance liquid chromatography
    Eisert, R
    Pawliszyn, J
    ANALYTICAL CHEMISTRY, 1997, 69 (16) : 3140 - 3147
  • [3] In-tube solid-phase microextraction with a dummy molecularly imprinted monolithic capillary coupled to ultra-performance liquid chromatography-tandem mass spectrometry to determine cannabinoids in plasma samples
    Marchioni, Camila
    Vieira, Tatiana Manzini
    Miller Crotti, Antonio Eduardo
    Crippa, Jose Alexandre
    Costa Queiroz, Maria Eugenia
    ANALYTICA CHIMICA ACTA, 2020, 1099 : 145 - 154
  • [4] Determination of quercetin using a molecularly imprinted polymer as solid-phase microextraction sorbent and high-performance liquid chromatography
    Rahimi, Marzieh
    Bahar, Soleiman
    Heydari, Rouhollah
    Amininasab, S. Mojtaba
    MICROCHEMICAL JOURNAL, 2019, 148 : 433 - 441
  • [5] Temperature-response polymer coating for in-tube solid-phase microextraction coupled to high-performance liquid chromatography
    Yu, Qiong-Wei
    Ma, Qiao
    Feng, Yu-Qi
    TALANTA, 2011, 84 (04) : 1019 - 1025
  • [6] Monolithic silica column for in-tube solid-phase microextraction coupled to high-performance liquid chromatography
    Shintani, Y
    Zhou, X
    Furuno, M
    Minakuchi, H
    Nakanishi, K
    JOURNAL OF CHROMATOGRAPHY A, 2003, 985 (1-2) : 351 - 357
  • [7] Determination of telmisartan in rat tissues by in-tube solid-phase microextraction coupled to high performance liquid chromatography
    Nie, J
    Zhao, Q
    Huang, JF
    Xiang, BR
    Feng, YQ
    JOURNAL OF SEPARATION SCIENCE, 2006, 29 (05) : 650 - 655
  • [8] Automated in-tube solid-phase microextraction-high-performance liquid chromatography for carbamate pesticide analysis
    Gou, YN
    Eisert, R
    Pawliszyn, J
    JOURNAL OF CHROMATOGRAPHY A, 2000, 873 (01) : 137 - 147
  • [9] Fiber-in-tube solid-phase microextraction with molecularly imprinted coating for sensitive analysis of antibiotic drugs by high performance liquid chromatography
    Hu, Yuling
    Song, Chaoyong
    Li, Gongke
    JOURNAL OF CHROMATOGRAPHY A, 2012, 1263 : 21 - 27
  • [10] In-tube solid-phase microextraction coupled to liquid chromatography (in-tube SPME/LC) analysis of nontricyclic antidepressants in human plasma
    Goncalves Silva, Bruno Jose
    Lancas, Fernando Mauro
    Costa Queiroz, Maria Eugenia
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2008, 862 (1-2): : 181 - 188