The Jonas study: Evaluation of the retrievability of the Cordis OptEase Inferior Vena Cava Filter

被引:67
作者
Oliva, VL
Szatmari, F
Giroux, MF
Flemming, BK
Cohen, SA
Soulez, G
机构
[1] CHU Montreal, Dept Radiol, Montreal, PQ H2L 4M1, Canada
[2] Csolnoky Ferenc Cty Hosp Veszprem, Dept Radiol, Veszprem, Hungary
[3] Queen Elizabeth II Med Ctr, Dept Radiol, Halifax, NS, Canada
[4] Cordis Corp, Miami, FL USA
关键词
IN-VIVO EVALUATION; ANIMAL-MODEL; FOLLOW-UP; EXPERIENCE; THROMBOSIS; PLACEMENT; REGISTRY; TRIAL; VITRO;
D O I
10.1097/01.RVI.0000171699.57957.C7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate the success, safety, and efficacy of the retrieval of the OptEase Permanent/Retrievable Vena Cava Filter (Cordis, Warren, NJ), when implanted for temporary protection against venous thromboembolism. MATERIALS AND METHODS: This prospective, multicenter, non-randomized study enrolled 27 patients who needed temporary protection against pulmonary embolism in whom the intent at the time of filter insertion was retrieval of the OptEase filter. Patients presented with deep venous thrombosis (n = 17), pulmonary embolism (PE) (n = 6), and high risk for PE without thromboembolic disease (n = 4). Assessments included duplex sonography of the access site performed within 24 hours of device implantation and retrieval. All patients underwent cavography before and after filter placement and filter retrieval. Contrast-enhanced computed tomography (CT) of the abdomen and clinical follow-tip was performed at 1 month after device retrieval. RESULTS: Of the 27 enrolled patients, 21 patients (77.8%) met the criteria for retrieval and all 21 patients (100%) had filters successfully retrieved with no associated adverse events. Retrieval was not attempted in six patients as a result of ongoing contraindication to anticoagulation (n = 3), large trapped thrombi within the filter (n = 2), and poor patient prognosis (n = 1). Time to retrieval ranged from 5 to 14 days with a mean implantation time of 11.1 days +/- 1.82. No symptomatic pulmonary embolism, vena cava wall injury, vena cava stenosis, significant bleeding, filter fracture, or filter migration was observed. Nineteen of the 21 patients (90.5%) were followed for 1-month post-retrieval with no device-related adverse events or symptomatic PE. Caval patency was documented in 18 of these 19 patients with CT. Two patients were lost to follow-up, and one patient refused to undergo CT examination. CONCLUSION: The OptEase permanent/retrievable vena cava filter can be safely and successfully retrieved up to 14 days in patients who no longer require inferior vena cava filter protection against pulmonary embolism.
引用
收藏
页码:1439 / 1445
页数:7
相关论文
共 26 条
[1]   Initial experience in humans with a new retrievable inferior vena cava filter [J].
Asch, MR .
RADIOLOGY, 2002, 225 (03) :835-844
[2]   Inferior vena cava filter removal after 317-day implantation [J].
Binkert, CA ;
Bansal, A ;
Gates, JD .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2005, 16 (03) :395-398
[3]   A new optional vena cava filter: Retrieval at 12 weeks in an animal model [J].
Brountzos, EN ;
Kaufman, JA ;
Venbrux, AC ;
Brown, R ;
Harry, J ;
Kinst, TF ;
Kleshinski, S ;
Ravenscroft, AC .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (06) :763-772
[4]   Incorporation of the Gunther temporary inferior vena cava filter into the caval wall [J].
Burbridge, BE ;
Walker, DR ;
Millward, SF .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1996, 7 (02) :289-290
[5]   Evaluation of the LGM Vena-Tech infrarenal vena cava filter in an ovine venous thromboembolism model [J].
Crochet, D ;
Grossetête, R ;
Bach-Lijour, B ;
Larguier, L ;
Le Nihouannen, JC .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 12 (06) :739-745
[6]   Animal experience in the Gunther Tulip retrievable inferior vena cava filter [J].
de Gregorio, MA ;
Gimeno, MJ ;
Tobio, R ;
Lostalé, F ;
Mainar, A ;
Beltrán, JM ;
Madariaga, B ;
Alfonso, ER ;
Medrano, J ;
Viloria, A .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 24 (06) :413-417
[7]   A clinical trial of vena caval filters in the prevention of pulmonary embolism in patients with proximal deep-vein thrombosis [J].
Decousus, H ;
Leizorovicz, A ;
Parent, F ;
Page, Y ;
Tardy, B ;
Girard, P ;
Laporte, S ;
Faivre, R ;
Charbonnier, B ;
Barral, FG ;
Huet, Y ;
Simonneau, G .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 338 (07) :409-415
[8]   In vivo evaluation of a retrievable vena cava filter - The Dibie-Musset filter: Experimental results [J].
Dibie, A ;
Musset, D ;
Heissler, M ;
Fournet, JC ;
Palau, R ;
Laborde, F .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1998, 21 (02) :151-157
[9]   PERCUTANEOUS INFERIOR VENA-CAVAL FILTERS - FOLLOW-UP OF 7 DESIGNS IN 320 PATIENTS [J].
FERRIS, EJ ;
MCCOWAN, TC ;
CARVER, DK ;
MCFARLAND, DR .
RADIOLOGY, 1993, 188 (03) :851-856
[10]  
Grassi Clement J, 2003, J Vasc Interv Radiol, V14, pS271