Diagnostic reference levels in angiography and interventional radiology: A Belgian multi-centre study

被引:23
作者
Bleeser, F. [1 ]
Hoornaert, M. -T. [1 ]
Smans, K. [2 ,3 ]
Struelens, L. [3 ]
Buls, N. [4 ]
Berus, D. [4 ]
Clerinx, P. [4 ]
Hambach, L. [5 ]
Malchair, F. [5 ]
Bosmans, H. [2 ]
机构
[1] CH Jolimont Lobbes, B-7100 Haine St Paul, Belgium
[2] Univ Hosp Gasthuisberg, B-3000 Louvain, Belgium
[3] CEN SCK, B-2400 Mol, Belgium
[4] Vrije Univ Brussels Hosp, B-1090 Brussels, Belgium
[5] CHU Sart Tilman, B-4000 Liege, Belgium
关键词
D O I
10.1093/rpd/ncn028
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The purpose of this study was to determine diagnostic reference levels (DRLs) for common angiographic and interventional procedures in Belgium. Dose Area Product (DAP) measurements were performed on 21 systems, ( 13 angiography and 4 vascular surgery centres). Type of procedure, total DAP, patient weight and height were collected on a daily basis during 1 y. The 75th percentile of the distribution of DAP values was defined as DRL. Preliminary DRLs were calculated for the three most frequent procedures for the whole population, for a weight class of patients (65-80 kg) and normalised to the standard size patient. Among them, the DRL for angiography of the lower limbs (30% of the procedures) from the whole population was 74.6 and 63.2 Gycm(2) for the size corrected. The mean DAP values of each room was then compared to these DRLs.
引用
收藏
页码:50 / 55
页数:6
相关论文
共 11 条
  • [1] AROUA A, 2004, 1104 I U RAD APPL
  • [2] A PHANTOM BASED METHOD FOR DERIVING TYPICAL PATIENT DOSES FROM MEASUREMENTS OF DOSE-AREA PRODUCT ON POPULATIONS OF PATIENTS
    CHAPPLE, CL
    BROADHEAD, DA
    FAULKNER, K
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (814) : 1083 - 1086
  • [3] Dosimetry Working Party of the Institute of physical sciences in medicine, 1992, NAT PROT PAT DOS MEA
  • [4] *EC, 1997, 91 EC
  • [5] *INT COMM RAD UN M, 1975, ICRP PUBL, V23
  • [6] THE REFERENCE MAN IN DIAGNOSTIC-RADIOLOGY DOSIMETRY
    LINDSKOUG, BA
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (773) : 431 - 437
  • [7] Diagnostic reference levels in interventional radiology
    Marshall, NW
    Chapple, CL
    Kotre, CJ
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (12) : 3833 - 3846
  • [8] A study of patient radiation doses in interventional radiological procedures
    McParland, BJ
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (842) : 175 - 185
  • [9] Preliminary reference levels in interventional cardiology
    Neofotistou, V
    Vano, E
    Padovani, R
    Kotre, J
    Dowling, A
    Toivonen, M
    Kottou, S
    Tsapaki, V
    Willis, S
    Bernardi, G
    Faulkner, K
    [J]. EUROPEAN RADIOLOGY, 2003, 13 (10) : 2259 - 2263
  • [10] PATIENT DOSE VALUES IN INTERVENTIONAL RADIOLOGY
    VANO, E
    GONZALEZ, L
    FERNANDEZ, JM
    GUIBELALDE, E
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (815) : 1215 - 1220