Using cavity theory to describe the dependence on detector density of dosimeter response in non-equilibrium small fields

被引:58
作者
Fenwick, John D. [1 ]
Kumar, Sudhir [2 ,3 ,4 ,5 ]
Scott, Alison J. D. [2 ,3 ]
Nahum, Alan E. [2 ,3 ]
机构
[1] Univ Oxford, Dept Oncol, Oxford OX3 7DQ, England
[2] Clatterbridge Canc Ctr, Dept Med Phys, Wirral CH63 4JY, Merseyside, England
[3] Univ Liverpool, Dept Phys, Liverpool L69 7ZE, Merseyside, England
[4] Bhabha Atom Res Ctr, CTCRS, Radiol Phys & Advisory Div, Bombay 400094, Maharashtra, India
[5] Homi Bhabha Natl Inst, Bombay 400085, Maharashtra, India
关键词
SPOT SIZE; BEAM;
D O I
10.1088/0031-9155/58/9/2901
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The dose imparted by a small non-equilibrium photon radiation field to the sensitive volume of a detector located within a water phantom depends on the density of the sensitive volume. Here this effect is explained using cavity theory, and analysed using Monte Carlo data calculated for schematically modelled diamond and Pinpoint-type detectors. The combined impact of the density and atomic composition of the sensitive volume on its response is represented as a ratio, F-w,F-det, of doses absorbed by equal volumes of unit density water and detector material co-located within a unit density water phantom. The impact of density alone is characterized through a similar ratio, P rho-, of doses absorbed by equal volumes of unit and modified density water. The cavity theory is developed by splitting the dose absorbed by the sensitive volume into two components, imparted by electrons liberated in photon interactions occurring inside and outside the volume. Using this theory a simple model is obtained that links P rho- to the degree of electronic equilibrium, s(ee), at the centre of a field via a parameter I-cav determined by the density and geometry of the sensitive volume. Following the scheme of Bouchard et al (2009 Med. Phys. 36 4654-63) F-w,F-det can be written as the product of P rho-, the water-to-detector stopping power ratio [(L) over bar (Delta)/rho](det)(w), and an additional factor Pfl-. In small fields [(L) over bar (Delta)/rho](det)(w) changes little with field-size; and for the schematic diamond and Pinpoint detectors Pfl- takes values close to one. Consequently most of the field-size variation in F-w,F-det originates from the P rho- factor. Relative changes in s(ee) and in the phantom scatter factor s(p) are similar in small fields. For the diamond detector, the variation of P rho- with s(ee) (and thus field-size) is described well by the simple cavity model using an I-cav parameter in line with independent Monte Carlo estimates. The model also captures the overall field-size dependence of P rho- for the schematic Pinpoint detector, again using an I-cav value consistent with independent estimates.
引用
收藏
页码:2901 / 2923
页数:23
相关论文
共 25 条
[1]   A new formalism for reference dosimetry of small and nonstandard fields [J].
Alfonso, R. ;
Andreo, P. ;
Capote, R. ;
Huq, M. Saiful ;
Kilby, W. ;
Kjaell, P. ;
Mackie, T. R. ;
Palmans, H. ;
Rosser, K. ;
Seuntjens, J. ;
Ullrich, W. ;
Vatnitsky, S. .
MEDICAL PHYSICS, 2008, 35 (11) :5179-5186
[2]  
Attix F. H., 1986, Introduction to Radiological Physics and Radiation Dosimetry, P172
[3]  
Berger MJ., 2005, ESTAR, PSTAR, and ASTAR: Computer programs for calculating Stopping-Power and range Tables for electrons. Protons, and Helium Ions (version 1.2.3)
[4]   Ionization chamber gradient effects in nonstandard beam configurations [J].
Bouchard, Hugo ;
Seuntjens, Jan ;
Carrier, Jean-Francois ;
Kawrakow, Iwan .
MEDICAL PHYSICS, 2009, 36 (10) :4654-4663
[5]   A GENERAL THEORY OF CAVITY IONISATION [J].
BURLIN, TE .
BRITISH JOURNAL OF RADIOLOGY, 1966, 39 (466) :727-&
[6]   The influence of small field sizes, penumbra, spot size and measurement depth on perturbation factors for microionization chambers [J].
Crop, F. ;
Reynaert, N. ;
Pittomvils, G. ;
Paelinck, L. ;
De Wagter, C. ;
Vakaet, L. ;
Thierens, H. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) :2951-2969
[7]   Small fields:: Nonequilibrium radiation dosimetry [J].
Das, Indra J. ;
Ding, George X. ;
Ahnesjo, Anders .
MEDICAL PHYSICS, 2008, 35 (01) :206-215
[8]   Beam characteristics and stopping-power ratios of small radiosurgery photon beams [J].
Ding, George X. ;
Ding, Frances .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (17) :5509-5521
[9]   Calculation of kQclin,Qmsrfclin,fmsr for several small detectors and for two linear accelerators using Monte Carlo simulations [J].
Francescon, P. ;
Cora, S. ;
Satariano, N. .
MEDICAL PHYSICS, 2011, 38 (12) :6513-6527
[10]   Application of Monte Carlo simulation to cavity theory based on the virtual electron source concept [J].
Fu, YC ;
Luo, ZM .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (17) :3263-3274