On the verification of the incident energy fluence in tomotherapy IMRT

被引:24
作者
Kapatoes, JM
Olivera, GH
Ruchala, KJ
Mackie, TR
机构
[1] TomoTherapy Inc, Middleton, WI 53562 USA
[2] Univ Wisconsin, Dept Med Phys, Madison, WI 53706 USA
[3] Univ Wisconsin, Dept Human Oncol, Madison, WI 53706 USA
关键词
D O I
10.1088/0031-9155/46/11/313
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For any radiotherapy verification technique, it is desirable that issues with the accelerator, multileaf collimator and patient position be detected. In previous works, an effective method for this level of verification was presented. This paper identifies second-order issues affecting the part of the process in which the incident energy fluence is verified. These problems will affect any rotational intensity-modulated radiotherapy delivery that divides each rotation or arc into projections; however the solutions offered in this paper are specific to the method previously developed. The issues affecting the energy fluence verification method include leaf bouncing, delivery implementation and leaf latency. All three matters were found to introduce small errors in the verified energy fluence values for a small fraction of leaf states. The overall effect on the deposited dose over the course of a rotational delivery involving thousands of beam pulses per rotation is negligible. Regardless, effective correction strategies are presented; these are utilized in order to characterize both the delivered energy fluence and deposited dose as accurately as possible.
引用
收藏
页码:2953 / 2965
页数:13
相关论文
共 36 条
  • [1] [Anonymous], 1988, NUMERICAL RECIPES C
  • [2] Characterization of the output for helical delivery of intensity modulated slit beams
    Balog, JP
    Mackie, TR
    Reckwerdt, P
    Glass, M
    Angelos, L
    [J]. MEDICAL PHYSICS, 1999, 26 (01) : 55 - 64
  • [3] Relative profile and dose verification of intensity-modulated radiation therapy
    Chang, JW
    Mageras, GS
    Chui, CS
    Ling, CC
    Lutz, W
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (01): : 231 - 240
  • [4] Verification of segmented beam delivery using a commercial electronic portal imaging device
    Curtin-Savard, AJ
    Podgorsak, EB
    [J]. MEDICAL PHYSICS, 1999, 26 (05) : 737 - 742
  • [5] In vivo dosimetry during external photon beam radiotherapy
    Essers, M
    Mijnheer, BJ
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (02): : 245 - 259
  • [6] Measurement of the response of Gd2O2S:Tb phosphor to 6 MV x-rays
    Glendinning, AG
    Hunt, SG
    Bonnett, DE
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (02) : 517 - 530
  • [7] The application of transit dosimetry to precision radiotherapy
    Hansen, VN
    Evans, PM
    Swindell, W
    [J]. MEDICAL PHYSICS, 1996, 23 (05) : 713 - 721
  • [8] Dual-beam imaging for online verification of radiotherapy field placement
    Jaffray, DA
    Chawla, K
    Yu, C
    Wong, JW
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05): : 1273 - 1280
  • [9] A radiographic and tomographic imaging system integrated into a medical linear accelerator for localization of bone and soft-tissue targets
    Jaffray, DA
    Drake, DG
    Moreau, M
    Martinez, AA
    Wong, JW
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (03): : 773 - 789
  • [10] Cone-beam computed tomography with a flat-panel imager: Initial performance characterization
    Jaffray, DA
    Siewerdsen, JH
    [J]. MEDICAL PHYSICS, 2000, 27 (06) : 1311 - 1323