Thermally imprinted microcone structure-assisted lateral-flow immunoassay platforms for detecting disease marker proteins

被引:16
作者
Aoyama, Shuhei [1 ,2 ]
Akiyama, Yuto [1 ]
Monden, Kenji [1 ]
Yamada, Masumi [2 ]
Seki, Minoru [2 ]
机构
[1] Denka Co Ltd, Denka Innovat Ctr, 3-5-1 Asahi Machi, Machida, Tokyo 1948560, Japan
[2] Chiba Univ, Grad Sch Engn, Dept Appl Chem & Biotechnol, Inage Ku, 1-33 Yayoi Cho, Chiba 2638522, Japan
关键词
C-REACTIVE PROTEIN; SURFACE MODIFICATION; POINT; PAPER; INFLUENZA; SYSTEM; DEVICE; ASSAY;
D O I
10.1039/c8an01903g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Although various types of on-site immunoassay platforms have been developed, facile and reliable sample-to-answer immunoassay systems are still under development. In this study, we proposed a lateral-flow immunoassay system utilizing a polymer sheet with microcone array structures fabricated by thermal imprinting. By adding a surfactant to the running/washing buffer, we were able to dispense with complicated chemical modification protocols, which are usually necessary to enhance antibody adsorption. We investigated three types of polymeric materials and confirmed that polycarbonate is most suitable as an imprinted polymer substrate. Experiments using microcone arrays with different distances revealed that the increased surface area with nanometer-scale surface roughness was key to achieving stable immobilization of antibodies. To assess the applicability of this assay to clinical diagnosis, C-reactive protein in a pure buffer and in a serum sample was analyzed as a model, with a similar to 0.1 mu g mL(-1) lower limit of detection. The presented approach would open a new path to the development of immunoassay-based on-site point-of-care testing by virtue of its simplicity of operation, high sensitivity, and versatility.
引用
收藏
页码:1519 / 1526
页数:8
相关论文
共 45 条
[1]  
Busa LSA, 2016, ANALYST, V141, P6598, DOI [10.1039/c6an01475e, 10.1039/C6AN01475E]
[2]   Micro-injection moulding of polymer microfluidic devices [J].
Attia, Usama M. ;
Marson, Silvia ;
Alcock, Jeffrey R. .
MICROFLUIDICS AND NANOFLUIDICS, 2009, 7 (01) :1-28
[3]   Surface modification for enhancing antibody binding on polymer-based microfluidic device for enzyme-linked immunosorbent assay [J].
Bai, Yunling ;
Koh, Chee Guan ;
Boreman, Megan ;
Juang, Yi-Je ;
Tang, I-Ching ;
Lee, L. James ;
Yang, Shang-Tian .
LANGMUIR, 2006, 22 (22) :9458-9467
[4]  
Banerjee R, 2018, ANALYST, V143, P1970, DOI [10.1039/c8an00307f, 10.1039/C8AN00307F]
[5]   Microfluidic immunosensor systems [J].
Bange, A ;
Halsall, HB ;
Heineman, WR .
BIOSENSORS & BIOELECTRONICS, 2005, 20 (12) :2488-2503
[6]   A critical insight into the development pipeline of microfluidic immunoassay devices for the sensitive quantitation of protein biomarkers at the point of care [J].
Barbosa, Ana I. ;
Reis, Nuno M. .
ANALYST, 2017, 142 (06) :858-882
[7]   Polymer microfabrication technologies for microfluidic systems [J].
Becker, Holger ;
Gaertner, Claudia .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2008, 390 (01) :89-111
[8]  
Becker H, 2017, METHODS MOL BIOL, V1547, P3, DOI 10.1007/978-1-4939-6734-6_1
[9]   Single and competitive protein adsorption on polymeric surfaces [J].
Bhaskar, Thanga Bhuvanesh Vijaya ;
Roch, Toralf ;
Romero, Oscar ;
Ma, Nan ;
Kratz, Karl ;
Lendlein, Andreas .
POLYMERS FOR ADVANCED TECHNOLOGIES, 2015, 26 (12) :1387-1393
[10]   Fluorescence lateral flow immunoassay based point-of-care nanodiagnostics for orthopedic implant-associated infection [J].
Borse, Vivek ;
Srivastava, Rohit .
SENSORS AND ACTUATORS B-CHEMICAL, 2019, 280 :24-33