Clinical requirements for extracellular vesicle assays

被引:70
作者
Ayers, Lisa [1 ,2 ]
Pink, Ryan [2 ]
Carter, David Raul Francisco [2 ]
Nieuwland, Rienk [3 ,4 ]
机构
[1] Oxford Univ Hosp NHS Fdn Trust, Dept Clin & Lab Immunol, Oxford OX3 7LE, England
[2] Oxford Brookes Univ, Fac Hlth & Life Sci, Dept Biol & Med Sci, Oxford, England
[3] Univ Amsterdam, Lab Expt Clin Chem, Amsterdam UMC, Amsterdam, Netherlands
[4] Univ Amsterdam, Vesicle Observat Ctr, Amsterdam UMC, Amsterdam, Netherlands
关键词
Extracellular vesicles; clinical assays; accreditation; ISO; 15189; standardisation; validation; verification; NANOPARTICLE TRACKING ANALYSIS; FLOW-CYTOMETRY; CIRCULATING MICROPARTICLES; PREANALYTICAL PARAMETERS; BLOOD-PLASMA; SIZE; STANDARDIZATION; QUANTIFICATION; ENUMERATION; LIMITATIONS;
D O I
10.1080/20013078.2019.1593755
中图分类号
Q2 [细胞生物学];
学科分类号
071009 ; 090102 ;
摘要
The scientific and clinical interest in extracellular vesicles (EV) has grown exponentially during the past 15 years. As most research indicates that EVs can be utilised in diagnostics, prognostics and therapeutics, we may be on the brink of establishing the clinical utility of EV measurement, but how can we make this a reality? If we are to introduce EVs as biomarkers into clinical laboratories, it will be necessary to offer fully validated, International Organization for Standardization (ISO) standard 15189 assays. ISO 15189 defines the quality management system requirements particular to medical laboratories and is used internationally to determine accreditation. In order for a clinical laboratory to offer an accredited test for EVs, this assay must have been subjected to a thorough assay validation process. This process requires the generation of data related to defined performance characteristics, to ensure that an assay is performing in accordance with the needs of its clinical users. Each of the defined performance characteristics will be discussed in this review, along with the issues that specifically affect EV analysis. Accreditation is increasingly important for all clinical laboratories and the standards required to achieve this are becoming more and more stringent. Therefore, as companies seek to develop the best assays to detect EVs and their molecular contents for clinical utility, and as we move rapidly towards our goal of offering EV analysis in the diagnosis and monitoring of disease, it is timely to highlight the requirements for the clinical accreditation of such assays. It is essential to consider these parameters to ensure that we develop the highest quality assays possible and ultimately the best outcomes for patients.
引用
收藏
页数:10
相关论文
共 65 条
[1]   Extracellular vesicle characteristics in stored red blood cell concentrates are influenced by the method of detection [J].
Almizraq, Ruqayyah J. ;
Seghatchian, Jerard ;
Holovati, Jelena L. ;
Acker, Jason P. .
TRANSFUSION AND APHERESIS SCIENCE, 2017, 56 (02) :254-260
[2]   Microparticles facilitate neutrophil/platelet crosstalk [J].
Andrews, Robert K. ;
Berndt, Michael C. .
BLOOD, 2008, 112 (06) :2174-2175
[3]   Verification of examination procedures in clinical laboratory for imprecision, trueness and diagnostic accuracy according to ISO 15189:2012: a pragmatic approach [J].
Antonelli, Giorgia ;
Padoan, Andrea ;
Aita, Ada ;
Sciacovelli, Laura ;
Plebani, Mario .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2017, 55 (10) :1501-1508
[4]   Pre-Analytical Handling Conditions and Small RNA Recovery from Urine for miRNA Profiling [J].
Armstrong, David A. ;
Dessaint, John A. ;
Ringelberg, Carol S. ;
Hazlett, Haley F. ;
Howard, Louisa ;
Abdalla, Moemen A. K. ;
Barnaby, Roxanna L. ;
Stanton, Bruce A. ;
Cervinski, Mark A. ;
Ashare, Alix .
JOURNAL OF MOLECULAR DIAGNOSTICS, 2018, 20 (05) :565-571
[5]   Procoagulant and platelet-derived microvesicle absolute counts determined by flow cytometry correlates with a measurement of their functional capacity [J].
Ayers, Lisa ;
Harrison, Paul ;
Kohler, Malcolm ;
Ferry, Berne .
JOURNAL OF EXTRACELLULAR VESICLES, 2014, 3 (01)
[6]   Measurement of circulating cell-derived microparticles by flow cytometry: Sources of variability within the assay [J].
Ayers, Lisa ;
Kohler, Malcolm ;
Harrison, Paul ;
Sargent, Ian ;
Dragovic, Rebecca ;
Schaap, Marianne ;
Nieuwland, Rienk ;
Brooks, Susan A. ;
Ferry, Berne .
THROMBOSIS RESEARCH, 2011, 127 (04) :370-377
[7]   The impact of various preanalytical treatments on the phenotype of small extracellular vesicles in blood analyzed by protein microarray [J].
Baek, Rikke ;
Sondergaard, Evo K. L. ;
Varming, Kim ;
Jorgensen, Malene M. .
JOURNAL OF IMMUNOLOGICAL METHODS, 2016, 438 :11-20
[8]   Uncertainty of measurement: an immunology laboratory perspective [J].
Beck, Sarah C. ;
Lock, Robert J. .
ANNALS OF CLINICAL BIOCHEMISTRY, 2015, 52 (01) :7-17
[9]   Exosome-Based Detection of EGFR T790M in Plasma from Non-Small Cell Lung Cancer Patients [J].
Castellanos-Rizaldos, Elena ;
Grimm, Dominik G. ;
Tadigotla, Vasisht ;
Hurley, James ;
Healy, John ;
Neal, Patricia L. ;
Sher, Mia ;
Venkatesan, Raajdeep ;
Karlovich, Chris ;
Raponi, Mitch ;
Krug, Anne ;
Noerholm, Mikkel ;
Tannous, Jihane ;
Tannous, Bakhos A. ;
Raez, Luis E. ;
Skog, Johan K. .
CLINICAL CANCER RESEARCH, 2018, 24 (12) :2944-2950
[10]   A new microparticle size calibration standard for use in measuring smaller microparticles using a new flow cytometer [J].
Chandler, W. L. ;
Yeung, W. ;
Tait, J. F. .
JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2011, 9 (06) :1216-1224