Adaquate fixation of plates for stability during mandibular reconstruction

被引:59
作者
Kimura, Akiko
Nagasao, Tomohisa
Kaneko, Tsuyoshi
Tamaki, Tamotsu
Miyamoto, Junpei
Nakajima, Tatsuo
机构
[1] Keio Univ, Sch Med, Dept Plast & Reconstruct Surg, Tokyo, Japan
[2] Natl Ctr Child Hlth & Dev, Dept Plast & Reconstruct Surg, Tokyo, Japan
[3] Nippon Inst Technol, Dept Mech Engn, Saitama, Japan
关键词
finite element analysis(method); mandibular reconstruction; plate; stress;
D O I
10.1016/j.jcms.2006.01.003
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: To determine the most appropriate plate fixation for dispersing the stress around screws in mandibular reconstruction. Material and methods: Forty-eight three-dimensional reconstructed mandibular models with Central (C) or right Lateral (L) defects were created and divided into three groups, fixed with: (1) two screws on each side, (2) three screws on the left side with the third (middle one) located distally, or (3) three screws on the left side with the third (middle one) located proximally. A 300 N vertical load was applied to the left molar region. The maximum stress was calculated using a finite element method and statistically evaluated. Results: Stress was concentrated around the screw at the distal end of the mandibular halves on the loaded side ('crucial screw'). For the C defect, stress concentrated around the 'crucial screw' was greater when there were only two screws for fixation (20.55 MPa) than when there were three screws (16.17 MPa; p = 0.022). No significant difference was found between the two groups with three screws for fixation. For the L defect on the other hand, stress on the 'crucial screw' was relatively greater when there were three screws for fixation (18.9 MPa) than when there were only two screws for (12.83 MPa; p = 0.051). No significant difference was found between the two groups with three screws for fixation. As for the reconstruction plate, the stresses on the plates were not different among the various screw groups for fixation nor among the C and L defects. Conclusion: This paradoxical result is explained by plate bowing. Thus in large defects screw positions should take into consideration the potential for plate bowing. (C) 2006 European Association for Cranio-Maxillofacial Surgery.
引用
收藏
页码:193 / 200
页数:8
相关论文
共 26 条
[1]   Volume-length impact of lateral jaw resections on complication rates [J].
Arden, RL ;
Rachel, JD ;
Marks, SC ;
Dang, K .
ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 1999, 125 (01) :68-72
[2]  
Blackwell KE, 1996, ARCH OTOLARYNGOL, V122, P672
[3]   The bridging lateral mandibular reconstruction plate revisited [J].
Blackwell, KE ;
Lacombe, V .
ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 1999, 125 (09) :988-993
[4]   THE FREE-FLAP AND PLATE IN OROMANDIBULAR RECONSTRUCTION - LONG-TERM REVIEW AND INDICATIONS [J].
BOYD, JB ;
MULHOLLAND, RS ;
DAVIDSON, J ;
GULLANE, PJ ;
ROTSTEIN, LE ;
BROWN, DH ;
FREEMAN, JE ;
IRISH, JC .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1995, 95 (06) :1018-1028
[5]  
BOYD JB, 1993, PLAST RECONSTR SURG, V92, P1266
[6]   Computerized analysis of resorbable polymer plates and screws for the rigid fixation of mandibular angle fractures [J].
Cox, T ;
Kohn, MW ;
Impelluso, T .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2003, 61 (04) :481-487
[7]   Modeling of jaw biomechanics in the reconstructed mandibulectomy patient [J].
Curtis, DA ;
Plesh, O ;
Hannam, AG ;
Sharma, A ;
Curtis, TA .
JOURNAL OF PROSTHETIC DENTISTRY, 1999, 81 (02) :167-173
[8]  
Curtis DA, 1997, HEAD NECK-J SCI SPEC, V19, P287, DOI 10.1002/(SICI)1097-0347(199707)19:4<287::AID-HED7>3.0.CO
[9]  
2-X
[10]   EXPERIENCE WITH AO RECONSTRUCTION PLATES AFTER PARTIAL MANDIBULAR RESECTION INVOLVING ITS CONTINUITY [J].
FREITAG, V ;
HELL, B ;
FISCHER, H .
JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 1991, 19 (05) :191-198