Design improvement and dynamic finite element analysis of novel ITI dental implant under dynamic chewing loads

被引:11
作者
Cheng, Yung-Chang [1 ]
Lin, Deng-Huei [1 ]
Jiang, Cho-Pei [2 ]
Lee, Shyh-Yuan [3 ,4 ]
机构
[1] Natl Kaohsiung First Univ Sci & Technol, Dept Mech & Automat Engn, Kaohsiung, Taiwan
[2] Natl Formosa Univ, Dept Power Mech Engn, Huwei Township, Taiwan
[3] Natl Yang Ming Univ, Dept Dent, Taipei, Taiwan
[4] Taipei Vet Gen Hosp, Dept Stomatol, Taipei, Taiwan
关键词
New dental implant model; dynamic loading; micromotion; uniform design; STRESS-DISTRIBUTION; BONE;
D O I
10.3233/BME-151346
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The main aim of this article was to introduce the application of a uniform design for experimental methods to drop the micromotion of a novel ITI dental implant model under the dynamic loads. Combining the characteristics of the traditional ITI and Nano-Tite implants, a new implant with concave holes has been constructed. Compared to the traditional ITI dental implant model, the micromotion of the new dental implant model was significantly reduced by explicit dynamic finite element analysis. From uniform design of experiments, the dynamic finite element analysis method was applied to caluculated the maximum micromotion of the full model. Finally, the chief design in all the experiment simulations which cause the minimum micromotion is picked as the advanced model of the design. Related to the original design, which was associated with a micromotion of 45.11 mu m, the micromotion of the improved version was 31.37 mu m, for an improvement rate of 30.5%.
引用
收藏
页码:S555 / S561
页数:7
相关论文
共 10 条
  • [1] Finite element modelling of implant designs and cortical bone thickness on stress distribution in maxillary type IV bone
    Chou, I-Chiang
    Lee, Shyh-Yuan
    Wu, Ming-Chang
    Sun, Chia-Wei
    Jiang, Cho-Pei
    [J]. COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2014, 17 (05) : 516 - 526
  • [2] Fang K.T., 1993, Number-Theoretic Methods in Statistics, VVolume 51
  • [3] Histological evaluation of peri-implant bone at implants subjected to occlusal overload or plaque accumulation
    Isidor, F
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 1997, 8 (01) : 1 - 9
  • [4] The role of primary stability for successful immediate loading of dental implants. A literature review
    Javed, Fawad
    Romanos, George E.
    [J]. JOURNAL OF DENTISTRY, 2010, 38 (08) : 612 - 620
  • [5] Static, dynamic and fatigue behaviors of dental implant using finite element method
    Kayabasi, Oguz
    Yuezbasioglu, Emir
    Erzincanli, Fehmi
    [J]. ADVANCES IN ENGINEERING SOFTWARE, 2006, 37 (10) : 649 - 658
  • [6] Kim ES, 2013, J ADV PROSTHODONT, V5, P21
  • [7] Uniform design and its applications in chemistry and chemical engineering
    Liang, YZ
    Fang, KT
    Xu, QS
    [J]. CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2001, 58 (01) : 43 - 57
  • [8] Mammadzada Sakhavat, 2011, Mathematical & Computational Applications, V16, P546
  • [9] FRICTIONAL INTERFACE MICROMOTIONS AND ANISOTROPIC STRESS-DISTRIBUTION IN A FEMORAL TOTAL HIP COMPONENT
    RUBIN, PJ
    RAKOTOMANANA, RL
    LEYVRAZ, PF
    ZYSSET, PK
    CURNIER, A
    HEEGAARD, JH
    [J]. JOURNAL OF BIOMECHANICS, 1993, 26 (06) : 725 - 739
  • [10] TANG M, 2004, QUAL ENG, V16, P461