Comparison of the Biomechanical Effect of the FFX Device Compared With Other Lumbar Fusion Devices: A Finite Element Study

被引:2
作者
Simon, Lionel [1 ]
Millot, Fabrice [2 ]
Hoarau, Xavier [3 ]
Buttin, Romain [1 ]
Srour, Robin [4 ,5 ]
机构
[1] LSCF, Lyon, France
[2] Hop St Avold, St Avold, France
[3] Polyclin Navarre, Pau, France
[4] Hop Civils Colmar, Colmar, France
[5] Hop Civils Colmar, 39 Ave Liberte, F-68024 Colmar, France
关键词
lumbar facet joint; FFX; finite element analysis; lumbar spine; biomechanics; pedicle screws; DISC; FIXATION; RANGE;
D O I
10.14444/8355
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: Lumbar decompression can result in postsurgical instability and spondylolisthesis in patients with lumbar spinal stenosis (LSS). While pedicle screw (PS) constructs improve stability and support fusion, their use can lead to adjacent level degeneration due to rigidity and resultant overload of anatomical structures. The FFX device is a facet spacer designed to be a less invasive alternative for obtaining fusion compared with PS. Objective: The present study aimed to compare biomechanical performance of the FFX device to different lumbar spine procedures using the finite element (FE) method. Study Design: Comparative biomechanical study by FE method. Methods: An FE model for the lumbar spine was developed and validated to assess vertebral displacement and stress variations in the facet joints and discs following surgery. Modeled scenarios included a healthy spine as a reference model, laminectomy (LAM), and prior to/following L4 -L5 fusion for LAM + FFX and LAM + PS. Results: LAM increased displacement compared with the healthy spine and both instrumented spine procedures. Facet joint stress at adjacent levels for LAM + PS was significantly higher than with LAM + FFX prior to fusion (+13.5% for L3 -L4; +15.7% for L5 -S1). Adjacent level disc stress at L5 -S1 was 7.7% higher for LAM + PS vs LAM + FFX. Adjacent level facet joint and disc stresses for LAM + FFX were equivalent to LAM + PS once fusion occurred. Conclusions: Instrumented spine fixation prevents the risk of lumbar instability associated with LAM alone. Compared with PS, the FFX device is a less invasive alternative for the treatment of LSS, which potentially lowers the risk of adjacent segment degeneration prior to fusion that provides equivalent stability once fusion is achieved.
引用
收藏
页码:935 / 943
页数:10
相关论文
共 26 条
  • [1] Do facet screws provide the required stability in lumbar fixation? A biomechanical comparison of the Boucher technique and pedicular fixation in primary and circumferential fusions
    Agarwala, Amit
    Bucklen, Brandon
    Muzumdar, Aditya
    Moldavsky, Mark
    Khalil, Saif
    [J]. CLINICAL BIOMECHANICS, 2012, 27 (01) : 64 - 70
  • [2] THE H, P AND H-P VERSION OF THE FINITE-ELEMENT METHOD - BASIS THEORY AND APPLICATIONS
    BABUSKA, I
    GUO, BQ
    [J]. ADVANCES IN ENGINEERING SOFTWARE, 1992, 15 (3-4) : 159 - 174
  • [3] Bertrand S, 2005, THESIS ECOLE NATL SU
  • [4] The effect of posterior polyester tethers on the biomechanics of proximal junctional kyphosis: a finite element analysis
    Bess, Shay
    Harris, Jeffrey E.
    Turner, Alexander W. L.
    LaFage, Virginie
    Smith, Justin S.
    Shaffrey, Christopher I.
    Schwab, Frank J.
    Haid, Regis W., Jr.
    [J]. JOURNAL OF NEUROSURGERY-SPINE, 2017, 26 (01) : 125 - 133
  • [5] Decompression Alone Versus Decompression and Fusion for Lumbar Degenerative Spondylolisthesis: A Meta-Analysis
    Chen, Zihao
    Xie, Peigen
    Feng, Feng
    Chhantyal, Kishor
    Yang, Yang
    Rong, Limin
    [J]. WORLD NEUROSURGERY, 2018, 111 : E158 - E170
  • [6] Denoziere G, 2004, MEMOIRE DE MASTER
  • [7] Comparison of eight published static finite element models of the intact lumbar spine: Predictive power of models improves when combined together
    Dreischarf, M.
    Zander, T.
    Shirazi-Adl, A.
    Puttlitz, C. M.
    Adam, C. J.
    Chen, C. S.
    Goel, V. K.
    Kiapour, A.
    Kim, Y. H.
    Labus, K. M.
    Little, J. P.
    Park, W. M.
    Wang, Y. H.
    Wilke, H. J.
    Rohlmann, A.
    Schmidt, H.
    [J]. JOURNAL OF BIOMECHANICS, 2014, 47 (08) : 1757 - 1766
  • [8] Laminectomy plus Fusion versus Laminectomy Alone for Lumbar Spondylolisthesis
    Ghogawala, Zoher
    Dziura, James
    Butler, William E.
    Dai, Feng
    Terrin, Norma
    Magge, Subu N.
    Coumans, Jean-Valery C. E.
    Harrington, J. Fred
    Amin-Hanjani, Sepideh
    Schwartz, J. Sanford
    Sonntag, Volker K. H.
    Barker, Fred G., II
    Benzel, Edward C.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2016, 374 (15) : 1424 - 1434
  • [9] Moment-rotation responses of the human lumbosacral spinal column
    Guan, Yabo
    Yoganandan, Narayan
    Moore, Jason
    Pintar, Frank A.
    Zhang, Jiangyue
    Maiman, Dennis J.
    Laud, Purushottam
    [J]. JOURNAL OF BIOMECHANICS, 2007, 40 (09) : 1975 - 1980
  • [10] Iatrogenic spondylolisthesis following laminectomy for degenerative lumbar stenosis: systematic review and current concepts
    Guha, Daipayan
    Heary, Robert F.
    Shamji, Mohammed F.