Bolus timing in high-pitch CT angiography of the aorta

被引:14
作者
Beeres, Martin [1 ]
Loch, Matthias [1 ]
Schulz, Boris [1 ]
Kerl, Matthias [1 ]
Al-Butmeh, Firas [1 ]
Bodelle, Boris [1 ]
Herrmann, Eva [2 ]
Gruber-Rouh, Tatjana [1 ]
Lee, Clara [1 ]
Jacobi, Volkmar [1 ]
Vogl, Thomas J. [1 ]
Bauer, Ralf W. [1 ]
机构
[1] Clin Goethe Univ, Dept Diagnost & Intervent Radiol, D-60590 Frankfurt, Germany
[2] Clin Goethe Univ, Dept Biostat, D-60590 Frankfurt, Germany
关键词
CT angiography; Aorta; High-pitch CT; Dual-source CT; Bolus geometry; IMAGE QUALITY; INJECTION; PROTOCOLS;
D O I
10.1016/j.ejrad.2013.01.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To investigate the bolus geometry in high-pitch CT angiography (CTA) of the aorta without ECG synchronisation in comparison to single-source CT. Methods: Overall 160 consecutive patients underwent CTA either in conventional single-source mode with a pitch of 1.2 (group 1), or in dual-source mode with a pitch of 3.0 (groups 2, 3 and 4) using different contrast media timings with bolus triggering at 140HU (5 s, group 1; 10 s, group 2; 12 s, group 3; 14 s, group 4). Contrast material, saline flush, flow rate and kV/mAs settings were kept equal for optimum comparability. Aortic attenuation was measured along the z-axis of the patient at different anatomic landmarks and subjective image quality was compared. Results: The most homogeneous enhancement of the aorta was reached with a delay of 10 s after reaching the trigger threshold. The imaging length was not significantly different, but the examination time was significantly (p < 0.001) shorter in the high-pitch group (7.7 s vs. 1.7 s for group 1 vs. 2, 3 and 4). Conclusion: In high-pitch CT angiography using a start delay of 10 s after a trigger threshold of 140HU in the descending aorta is reached, a homogenous contrast along the z-axis is accomplished. (C) 2013 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:1028 / 1033
页数:6
相关论文
共 17 条
  • [1] American Association of Physicists in Medicine (AAPM), 2011, Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations: the report of AAPM Task Group 204
  • [2] Intravenous Contrast Medium Administration and Scan Timing at CT: Considerations and Approaches
    Bae, Kyongtae T.
    [J]. RADIOLOGY, 2010, 256 (01) : 32 - 61
  • [3] High-pitch dual-source CT angiography of the whole aorta without ECG synchronisation: Initial experience
    Beeres, Martin
    Schell, Boris
    Mastragelopoulos, Aristidis
    Herrmann, Eva
    Kerl, Josef Matthias
    Gruber-Rouh, Tatjana
    Lee, Clara
    Siebenhandl, Petra
    Bodelle, Boris
    Zangos, Stephan
    Vogl, Thomas J.
    Jacobi, Volkmar
    Bauer, Ralf W.
    [J]. EUROPEAN RADIOLOGY, 2012, 22 (01) : 129 - 137
  • [4] Image Quality, Contrast Enhancement, and Radiation Dose of ECG-Triggered High-Pitch CT Versus Non-ECG-Triggered Standard-Pitch CT of the Thoracoabdominal Aorta
    Bolen, Michael A.
    Popovic, Zoran B.
    Tandon, Navdeep
    Flamm, Scott D.
    Schoenhagen, Paul
    Halliburton, Sandra S.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 198 (04) : 931 - 938
  • [5] Intravenous contrast material administration at 16-detector row helical CT coronary angiography: Test bolus versus bolus-tracking technique
    Cademartiri, F
    Nieman, K
    van der Lugt, A
    Raaijmakers, RH
    Mollet, N
    Pattynama, PMT
    de Feyter, PJ
    Krestin, GP
    [J]. RADIOLOGY, 2004, 233 (03) : 817 - 823
  • [6] Size-specific Dose Estimates for Adult Patients at CT of the Torso
    Christner, Jodie A.
    Braun, Natalie N.
    Jacobsen, Megan C.
    Carter, Rickey E.
    Kofler, James M.
    McCollough, Cynthia H.
    [J]. RADIOLOGY, 2012, 265 (03) : 841 - 847
  • [7] Comparison of image quality and radiation dose of different pulmonary CTA protocols on a 128-slice CT: high-pitch dual source CT, dual energy CT and conventional spiral CT
    De Zordo, Tobias
    von Lutterotti, Klemens
    Dejaco, Christian
    Soegner, Peter F.
    Frank, Renate
    Aigner, Friedrich
    Klauser, Andrea S.
    Pechlaner, Christoph
    Schoepf, U. Joseph
    Jaschke, Werner R.
    Feuchtner, Gudrun M.
    [J]. EUROPEAN RADIOLOGY, 2012, 22 (02) : 279 - 286
  • [8] How to design injection protocols for multiple detector-row CT angiography (MDCTA)
    Fleischmann, D
    [J]. EUROPEAN RADIOLOGY, 2005, 15 : E60 - E65
  • [9] Fleischmann D, 2010, RADIOL CLIN N AM, V48, pvii
  • [10] CT Angiography: Injection and Acquisition Technique
    Fleischmann, Dominik
    [J]. RADIOLOGIC CLINICS OF NORTH AMERICA, 2010, 48 (02) : 237 - +