Dosimetric assessment of a novel metal artifact reduction method in CT images

被引:19
|
作者
Spadea, Maria F. [1 ]
Verburg, Joost [2 ,3 ]
Baroni, Guido [4 ,5 ]
Seco, Joao [2 ,3 ]
机构
[1] Magna Graecia Univ Catanzaro, Dept Expt & Clin Med, I-88100 Catanzaro, Italy
[2] Harvard Univ, Sch Med, Dept Radiat Oncol, Cambridge, MA 02138 USA
[3] Massachusetts Gen Hosp, Boston, MA 02114 USA
[4] Politecn Milano Univ, Dept Bioengn, Milan, Italy
[5] Fdn CNAO, Bioengn Unit, Pavia, Italy
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2013年 / 14卷 / 01期
关键词
CT; metal artifact reduction; dose inaccuracies; IMRT; HIP PROSTHESES; MONTE-CARLO; SUPPRESSION; IMPLANTS;
D O I
10.1120/jacmp.v14i1.4027
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to assess the ability of metal artifact reduction (MAR) algorithm in restoring the CT image quality while correcting the tissue density information for the accurate estimation of the absorbed dose. A phantom filled with titanium (low-Z metal) and Cerrobend (high-Z metal) inserts was used for this purpose. The MAR algorithm was applied to phantom's CT dataset. Static intensity-modulated radiation therapy (IMRT) plans, including five beam angles, were designed and optimized on the uncorrected images to deliver 10 Gy on the simulated target. Monte Carlo dose calculation was computed on uncorrected, corrected, and ground truth image datasets. It was firstly verified that MAR methodology was able to correct HU errors due to the metal presence. In the worst situation (high-Z phantom), the image difference, uncorrected ground truth and corrected ground truth, went from -4.4 +/- 118.8 HU to 0.4 +/- 10.8 HU, respectively. Secondly, it was observed that the impact of dose errors estimation depends on the atomic number of the metal: low-Z inserts do not produce significant dose inaccuracies, while high-Z implants substantially influence the computation of the absorbed dose. In this latter case, dose errors in the PTV region were up to 23.56% (9.72% mean value) when comparing the uncorrected vs. the ground truth dataset. After MAR correction, errors dropped to 0.11% (0.10% mean value). In conclusion, it was assessed that the new MAR algorithm is able to restore image quality without distorting mass density information, thus producing a more accurate dose estimation.
引用
收藏
页码:299 / 304
页数:6
相关论文
共 50 条
  • [1] Novel 3-fold metal artifact reduction method for CT images
    Desai, Shrinivas D.
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2021, 70
  • [2] Advances in metal artifact reduction in CT images: A review of traditional and novel metal artifact reduction techniques
    Selles, Mark
    van Osch, Jochen A. C.
    Maas, Mario
    Boomsma, Martijn F.
    Wellenberg, Ruud H. H.
    EUROPEAN JOURNAL OF RADIOLOGY, 2024, 170
  • [3] Artifact reduction filtering method for CT images
    Matani, A
    Terakawa, K
    SECOND JOINT EMBS-BMES CONFERENCE 2002, VOLS 1-3, CONFERENCE PROCEEDINGS: BIOENGINEERING - INTEGRATIVE METHODOLOGIES, NEW TECHNOLOGIES, 2002, : 1035 - 1036
  • [4] Fast Method to Reconstruct and Enhance CT Images: Applied to Metal Artifact Reduction
    Benammar, A.
    Allag, A.
    Araar, I
    Merabti, H.
    Benyahia, A.
    Mazouz, N.
    RUSSIAN JOURNAL OF NONDESTRUCTIVE TESTING, 2021, 57 (07) : 600 - 608
  • [5] Fast Method to Reconstruct and Enhance CT Images: Applied to Metal Artifact Reduction
    A. Benammar
    A. Allag
    I. Araar
    H. Merabti
    A. Benyahia
    N. Mazouz
    Russian Journal of Nondestructive Testing, 2021, 57 : 600 - 608
  • [6] A pragmatic approach to metal artifact reduction in CT: merging of metal artifact reduced images
    Watzke, O
    Kalender, WA
    EUROPEAN RADIOLOGY, 2004, 14 (05) : 849 - 856
  • [7] A pragmatic approach to metal artifact reduction in CT: merging of metal artifact reduced images
    Oliver Watzke
    Willi A. Kalender
    European Radiology, 2004, 14 : 849 - 856
  • [8] A Method for Assessing the Effect of Metal Artifact Reduction Algorithms on Quantitative Characteristics of CT Images
    Yu. A. Vasilev
    D. S. Semenov
    E. S. Akhmad
    O. Yu. Panina
    K. A. Sergunova
    A. V. Petraikin
    Biomedical Engineering, 2020, 54 : 285 - 288
  • [9] A Method for Assessing the Effect of Metal Artifact Reduction Algorithms on Quantitative Characteristics of CT Images
    Vasilev, Yu A.
    Semenov, D. S.
    Akhmad, E. S.
    Panina, O. Yu
    Sergunova, K. A.
    Petraikin, A., V
    BIOMEDICAL ENGINEERING-MEDITSINSKAYA TEKNIKA, 2020, 54 (04): : 285 - 288
  • [10] Assessment of metal artifact reduction methods in pelvic CT
    Abdoli, Mehrsima
    Mehranian, Abolfazl
    Ailianou, Angeliki
    Becker, Minerva
    Zaidi, Habib
    MEDICAL PHYSICS, 2016, 43 (04) : 1588 - 1597