The stability of liquid-filled matrix ionization chamber electronic portal imaging devices for dosimetry purposes

被引:17
作者
Louwe, RJW [1 ]
Tielenburg, R [1 ]
van Ingen, KM [1 ]
Mijnheer, BJ [1 ]
van Herk, MB [1 ]
机构
[1] Antoni Van Leeuwenhoek Hosp, Netherlands Canc Inst, Dept Radiotherapy, NL-1066 CX Amsterdam, Netherlands
关键词
Radiotherapy; dose verification; EPID; portal dosimetry; quality assurance;
D O I
10.1118/1.1668411
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study was performed to determine the stability of liquid-filled matrix ionization chamber (LiFi-type) electronic portal imaging devices (EPID) for dosimetric purposes. The short- and long-term stability of the response was investigated, as well as the importance of factors influencing the response (e.g., temperature fluctuations, radiation damage, and the performance of the electronic hardware). It was shown that testing the performance of the electronic hardware as well as the short-term stability of the imagers may reveal the cause of a poor long-term stability of the imager response. In addition, the short-term stability was measured to verify the validity of the fitted dose-response curve immediately after beam startup. The long-term stability of these imagers could be considerably improved by correcting for room temperature fluctuations and gradual changes in response due to radiation damage. As a result, the reproducibility was better than 1% (I SD) over a period of two years. The results of this study were used to formulate recommendations for a quality control program for portal dosimetry. The effect of such a program was assessed by comparing the results of portal dosimetry and in vivo dosimetry using diodes during the treatment of 31 prostate patients. The improvement of the results for portal dosimetry was consistent with the deviations observed with the reproducibility tests in that particular period. After a correction for the variation in response of the imager, the average difference between the measured and prescribed dose during the treatment of prostate patients was -0.7% +/- 1.5% (1 SD), and -0.6% +/- 1.1 % (1 SD) for EPID and diode in vivo dosimetry, respectively. It can be concluded that a high stability of the response can be achieved for this type of EPID by applying a rigorous quality control program. (C) 2004 American Association of Physicists in Medicine.
引用
收藏
页码:819 / 827
页数:9
相关论文
共 13 条
[1]   The dose response relationship of a liquid-filled electronic portal imaging device [J].
Boellaard, R ;
vanHerk, M ;
Mijnheer, BJ .
MEDICAL PHYSICS, 1996, 23 (09) :1601-1611
[2]   Dosimetric characteristics of a liquid-filled electronic portal imaging device [J].
Essers, M ;
Hoogervorst, BR ;
vanHerk, M ;
Lanson, H ;
Mijnheer, BJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05) :1265-1272
[3]   Clinical use of electronic portal imaging: Report of AAPM radiation therapy committee Task Group 58 [J].
Herman, MG ;
Balter, JM ;
Jaffray, DA ;
McGee, KP ;
Munro, P ;
Shalev, S ;
Van Herk, M ;
Wong, JW .
MEDICAL PHYSICS, 2001, 28 (05) :712-737
[4]   In vivo dosimetry during conformal radiotherapy - Requirements for and findings of a routine procedure [J].
Lanson, JH ;
Essers, M ;
Meijer, GJ ;
Minken, AWH ;
Uiterwaal, GJ ;
Mijnheer, BJ .
RADIOTHERAPY AND ONCOLOGY, 1999, 52 (01) :51-59
[5]   Three-dimensional dose reconstruction of breast cancer treatment using portal imaging [J].
Louwe, RJW ;
Damen, EMF ;
van Herk, M ;
Minken, AWH ;
Törzsök, O ;
Mijnheer, BJ .
MEDICAL PHYSICS, 2003, 30 (09) :2376-2389
[6]   Commissioning and periodic quality assurance of a clinical electronic portal imaging device [J].
Low, DA ;
Klein, EE ;
Maag, DK ;
Umfleet, WE ;
Purdy, JA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (01) :117-123
[7]   Accurate in vivo dosimetry of a randomized trial of prostate cancer irradiation [J].
Meier, GJ ;
Minken, AWH ;
van Ingen, KM ;
Smulders, B ;
Uiterwaal, H ;
Mijnheer, BJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (05) :1409-1418
[8]   Image quality parameters for the production standard Eliav PORTpro portal imaging device [J].
Poynter, AJ .
BRITISH JOURNAL OF RADIOLOGY, 1999, 72 (860) :802-804
[9]   A quality control test for electronic portal imaging devices [J].
Rajapakshe, R ;
Luchka, K ;
Shalev, S .
MEDICAL PHYSICS, 1996, 23 (07) :1237-1244
[10]   DOSE EVALUATION TO PATIENTS IRRADIATED BY CO-60 BEAMS, BY MEANS OF DIRECT MEASUREMENT ON THE INCIDENT AND ON THE EXIT SURFACES [J].
RIZZOTTI, A ;
COMPRI, C ;
GARUSI, GF .
RADIOTHERAPY AND ONCOLOGY, 1985, 3 (03) :279-283