Influence of different dose calculation algorithms on the estimate of NTCP for lung complications

被引:28
作者
Hedin, Emma [1 ]
Back, Anna [1 ,2 ]
机构
[1] Univ Gothenburg, Dept Radiat Phys, S-41345 Gothenburg, Sweden
[2] Sahlgrens Univ Hosp, Gothenburg, Sweden
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2013年 / 14卷 / 05期
关键词
radiation therapy; pneumonitis; dose calculation algorithms; NTCP; RADIATION PNEUMONITIS; MODELS; CANCER; TISSUE; IRRADIATION; THERAPY; VALUES;
D O I
10.1120/jacmp.v14i5.4316
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Due to limitations and uncertainties in dose calculation algorithms, different algorithms can predict different dose distributions and dose-volume histograms for the same treatment. This can be a problem when estimating the normal tissue complication probability (NTCP) for patient-specific dose distributions. Published NTCP model parameters are often derived for a different dose calculation algorithm than the one used to calculate the actual dose distribution. The use of algorithm-specific NTCP model parameters can prevent errors caused by differences in dose calculation algorithms. The objective of this work was to determine how to change the NTCP model parameters for lung complications derived for a simple correction-based pencil beam dose calculation algorithm, in order to make them valid for three other common dose calculation algorithms. NTCP was calculated with the relative seriality (RS) and Lyman-Kutcher-Burman (LKB) models. The four dose calculation algorithms used were the pencil beam (PB) and collapsed cone (CC) algorithms employed by Oncentra, and the pencil beam convolution (PBC) and anisotropic analytical algorithm (AAA) employed by Eclipse. Original model parameters for lung complications were taken from four published studies on different grades of pneumonitis, and new algorithm-specific NTCP model parameters were determined. The difference between original and new model parameters was presented in relation to the reported model parameter uncertainties. Three different types of treatments were considered in the study: tangential and locoregional breast cancer treatment and lung cancer treatment. Changing the algorithm without the derivation of new model parameters caused changes in the NTCP value of up to 10 percentage points for the cases studied. Furthermore, the error introduced could be of the same magnitude as the confidence intervals of the calculated NTCP values. The new NTCP model parameters were tabulated as the algorithm was varied from PB to PBC, AAA, or CC. Moving from the PB to the PBC algorithm did not require new model parameters; however, moving from PB to AAA or CC did require a change in the NTCP model parameters, with CC requiring the largest change. It was shown that the new model parameters for a given algorithm are different for the different treatment types.
引用
收藏
页码:127 / 139
页数:13
相关论文
共 14 条
  • [1] Sensitivity of NTCP parameter values against a change of dose calculation algorithm
    Brink, Carsten
    Berg, Martin
    Nielsen, Morten
    [J]. MEDICAL PHYSICS, 2007, 34 (09) : 3579 - 3586
  • [2] Incorporating an improved dose-calculation algorithm in conformal radiotherapy of lung cancer: re-evaluation of dose in normal lung tissue
    De Jaeger, K
    Hoogeman, MS
    Engelsman, M
    Seppenwoolde, Y
    Damen, EMF
    Mijnheer, BJ
    Boersma, LJ
    Lebesque, JV
    [J]. RADIOTHERAPY AND ONCOLOGY, 2003, 69 (01) : 1 - 10
  • [3] On the dosimetric behaviour of photon dose calculation algorithms in the presence of simple geometric heterogeneities:: comparison with Monte Carlo calculations
    Fogliata, Antonella
    Vanetti, Eugenio
    Albers, Dirk
    Brink, Carsten
    Clivio, Alessandro
    Knoos, Tommy
    Nicolini, Giorgia
    Cozzi, Luca
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (05) : 1363 - 1385
  • [4] Radiation pneumonitis after breast cancer irradiation:: Analysis of the complication probability using the relative seriality model
    Gagliardi, G
    Bjöhle, J
    Lax, I
    Ottolenghi, A
    Eriksson, F
    Liedberg, A
    Lind, P
    Rutqvist, LE
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 46 (02): : 373 - 381
  • [5] TUMOR AND NORMAL TISSUE RESPONSES TO FRACTIONATED NONUNIFORM DOSE DELIVERY
    KALLMAN, P
    AGREN, A
    BRAHME, A
    [J]. INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 1992, 62 (02) : 249 - 262
  • [6] Comparison of dose calculation algorithms for treatment planning in external photon beam therapy for clinical situations
    Knoos, Tommy
    Wieslander, Elinore
    Cozzi, Luca
    Brink, Carsten
    Fogliata, Antonella
    Albers, Dirk
    Nystrom, Hakan
    Lassen, Soren
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (22) : 5785 - 5807
  • [7] CALCULATION OF COMPLICATION PROBABILITY FACTORS FOR NON-UNIFORM NORMAL TISSUE IRRADIATION - THE EFFECTIVE VOLUME METHOD
    KUTCHER, GJ
    BURMAN, C
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 16 (06): : 1623 - 1630
  • [8] LYMAN JT, 1985, RAD RES S, V8, pS13, DOI DOI 10.2307/3576626
  • [9] Influence of dose calculation algorithms on the predicted dose distributions and NTCP values for NSCLC patients
    Nielsen, Tine B.
    Wieslander, Elinore
    Fogliata, Antonella
    Nielsen, Morten
    Hansen, Olfred
    Brink, Carsten
    [J]. MEDICAL PHYSICS, 2011, 38 (05) : 2412 - 2418
  • [10] Niemierko A., 1999, Medical Physics, V26, P1100