Rotational Stability and Visual Quality in Eyes With Rayner Toric Intraocular Lens Implantation

被引:10
作者
Alberdi, Txomin [1 ]
Macias-Murelaga, Beatriz [1 ]
Bascaran, Lucia [1 ]
Goni, Nahia [1 ]
Saez de Arregui, Sandra [1 ]
Mendicute, Javier [1 ]
机构
[1] Donostia Hosp, Dept Ophthalmol, San Sebastian, Spain
关键词
ASTIGMATISM;
D O I
10.3928/1081597X-20120921-04
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To assess astigmatic reduction and rotational stability of Rayner T-flex toric intraocular lenses (IOL) (573T and 623T; Rayner Intraocular Lenses Ltd) in a series of cataract surgery patients with corneal astigmatism > 1.50 diopters (D). METHODS: This prospective, observational study included 27 eyes from 22 consecutive patients with preoperative regular corneal astigmatism >= 1.50 D in which Rayner T-flex toric IOL implantation was performed after phacoemulsification. Uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), residual refractive sphere, residual keratometric and refractive cylinder, and deviation from attempted toric IOL axis were measured. RESULTS: Mean UDVA 3 months after surgery was 0.1042 +/- 0.1290 with 20/40 or better in 96% of eyes and 20/25 or better in 52% of eyes. Mean refractive postoperative sphere was 0.25 +/- 0.37 D and mean refractive postoperative cylinder was 0.52 +/- 0.63 D. Vector analysis of attempted versus achieved visual correction showed that 100% of eyes were within +/- 1.00 D and 87% of eyes were within +/- 0.50 D. Mean change in keratometric astigmatism was 0.54 +/- 1.33 for J(0) and -0.24 +/- 1.17 for J(45) (P >.05). Twenty-five (92.6%) eyes had IOL rotation < 10 degrees. CONCLUSIONS: Rayner T-flex toric IOL implantation is an effective and reliable option to correct preexisting astigmatism in cataract surgery. [J Refract Surg. 2012;28(10):696-700.] doi:10.3928/1081597X-20120921-04
引用
收藏
页码:696 / +
页数:6
相关论文
共 23 条
[1]   Visual function and patient experience after bilateral implantation of toric intraocular lenses [J].
Ahmed, Iqbal Ike K. ;
Rocha, Guillermo ;
Slomovic, Allan R. ;
Climenhaga, Harold ;
Gohill, Jit ;
Gregoire, Alain ;
Ma, Joseph .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (04) :609-616
[2]  
Becker KA, 2004, OPHTHALMOLOGE, V101, P600, DOI 10.1007/s00347-003-0951-7
[3]   Assessment of toric intraocular lens alignment by a refractive power/corneal analyzer system and slitlamp observation [J].
Carey, Paul J. ;
Leccisotti, Antonio ;
McGilligan, Victoria E. ;
Goodall, Ed A. ;
Moore, C. B. Tara .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (02) :222-229
[4]   Comparative rotational stability of single-piece open-loop acrylic and plate-haptic silicone toric intraocular lenses [J].
Chang, David F. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (11) :1842-1847
[5]   Evaluation of a toric intraocular lens with a Z-haptic [J].
De Silva, Don Julian ;
Ramkissoon, Yashin D. ;
Bloom, Philip A. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (09) :1492-1498
[6]   Injectable 1-piece hydrophilic acrylic toric intraocular lens for cataract surgery: Efficacy and stability [J].
Entabi, Molham ;
Harman, Francesca ;
Lee, Nicholas ;
Bloom, Philip A. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (02) :235-240
[7]   Toric Intraocular Lens Misalignment Inducing Astigmatism After Refractive Surgery [J].
Forseto, Adriana S. ;
Nose, Ricardo M. ;
Nose, Walton .
JOURNAL OF REFRACTIVE SURGERY, 2011, 27 (09) :690-692
[8]  
Gerten G, 2001, OPHTHALMOLOGE, V98, P715, DOI 10.1007/s003470170077
[9]   BIOMETRY OF 7,500 CATARACTOUS EYES [J].
HOFFER, KJ .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1980, 90 (03) :360-368
[10]  
Kim Myung Hun, 2010, Korean J Ophthalmol, V24, P207, DOI 10.3341/kjo.2010.24.4.207