Evaluation of a Completely Robotized Neurosurgical Operating Microscope

被引:29
作者
Kantelhardt, Sven R. [1 ]
Finke, Markus [2 ]
Schweikard, Achim [2 ]
Giese, Alf [1 ]
机构
[1] Johannes Gutenberg Univ Mainz, Dept Neurosurg, D-55101 Mainz, Germany
[2] Med Univ Lubeck, Inst Robot & Cognit Syst, D-23538 Lubeck, Germany
关键词
Brain tumor; Neuronavigation; Operating microscope; Optical coherence tomography; Robot-assisted surgery; OPTICAL COHERENCE TOMOGRAPHY; TISSUE; RESECTION;
D O I
10.1227/NEU.0b013e31827235f8
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND: Operating microscopes are essential for most neurosurgical procedures. Modern robot-assisted controls offer new possibilities, combining the advantages of conventional and automated systems. OBJECTIVE: We evaluated the prototype of a completely robotized operating microscope with an integrated optical coherence tomography module. METHODS: A standard operating microscope was fitted with motors and control instruments, with the manual control mode and balance preserved. In the robot mode, the microscope was steered by a remote control that could be fixed to a surgical instrument. External encoders and accelerometers tracked microscope movements. The microscope was additionally fitted with an optical coherence tomography-scanning module. RESULTS: The robotized microscope was tested on model systems. It could be freely positioned, without forcing the surgeon to take the hands from the instruments or avert the eyes from the oculars. Positioning error was about 1 mm, and vibration faded in 1 second. Tracking of microscope movements, combined with an autofocus function, allowed determination of the focus position within the 3-dimensional space. This constituted a second loop of navigation independent from conventional infrared reflector-based techniques. In the robot mode, automated optical coherence tomography scanning of large surface areas was feasible. CONCLUSION: The prototype of a robotized optical coherence tomography-integrated operating microscope combines the advantages of a conventional manually controlled operating microscope with a remote-controlled positioning aid and a self-navigating microscope system that performs automated positioning tasks such as surface scans. This demonstrates that, in the future, operating microscopes may be used to acquire intra-operative spatial data, volume changes, and structural data of brain or brain tumor tissue.
引用
收藏
页码:A19 / A26
页数:8
相关论文
共 20 条
[1]   A robotic C-arm fluoroscope [J].
Binder, N ;
Matthäus, L ;
Burgkart, R ;
Schweikard, A .
INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2005, 1 (03) :108-116
[2]   Time-domain and spectral-domain optical coherence tomography in the analysis of brain tumor tissue [J].
Boehringer, H. J. ;
Boller, D. ;
Leppert, J. ;
Knopp, U. ;
Lankenau, E. ;
Reusche, E. ;
Huettmann, G. ;
Giese, A. .
LASERS IN SURGERY AND MEDICINE, 2006, 38 (06) :588-597
[3]   Automatic scanning of large tissue areas in neurosurgery using optical coherence tomography [J].
Finke, M. ;
Kantelhardt, S. ;
Schlaefer, A. ;
Bruder, R. ;
Lankenau, E. ;
Giese, A. ;
Schweikard, A. .
INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2012, 8 (03) :327-336
[4]   Motorization of a surgical microscope for intra-operative navigation and intuitive control [J].
Finke, M. ;
Schweikard, A. .
INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2010, 6 (03) :269-280
[5]  
Giese A, 2006, PROG BOIMED OPT IMAG, V7
[6]  
Giorgi C, 1995, J Image Guid Surg, V1, P158, DOI 10.1002/(SICI)1522-712X(1995)1:3<158::AID-IGS5>3.0.CO
[7]  
2-9
[8]   Robotics in child neurosurgery [J].
Giorgi, C ;
Sala, R ;
Riva, D ;
Cossu, A ;
Eisenberg, H .
CHILDS NERVOUS SYSTEM, 2000, 16 (10-11) :832-834
[9]   Axel Perneczky: A Remembrance [J].
Grotenhuis, J. Andre ;
Cohen, Alan R. .
NEUROSURGERY, 2010, 66 (06) :1036-1038
[10]   Perioperative course and accuracy of screw positioning in conventional, open robotic-guided and percutaneous robotic-guided, pedicle screw placement [J].
Kantelhardt, Sven Rainer ;
Martinez, Ramon ;
Baerwinkel, Stefan ;
Burger, Ralf ;
Giese, Alf ;
Rohde, Veit .
EUROPEAN SPINE JOURNAL, 2011, 20 (06) :860-868