A 0.338 cm3, Artifact-Free, 64-Contact Neuromodulation Platform for Simultaneous Stimulation and Sensing

被引:45
作者
Rozgic, Dejan [1 ]
Hokhikyan, Vahagn [2 ]
Jiang, Wenlong [2 ]
Akita, Ippei [3 ]
Basir-Kazeruni, Sina [2 ]
Chandrakumar, Hariprasad [2 ]
Markovic, Dejan [1 ]
机构
[1] Univ Calif Los Angeles, Parallel Data Architectures Grp, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Elect & Comp Engn Dept, Los Angeles, CA 90095 USA
[3] Natl Inst Adv Ind Sci & Technol, Tsukuba, Ibaraki 3058568, Japan
关键词
Artifact rejection; current-controlled stimulation; high dynamic range; implant-grid; neural probe; neural recording; neuromodulation; wireless power transfer; INTEGRATED-CIRCUIT; RECTIFIER; CHIP; DESIGN; SYSTEM; NOISE;
D O I
10.1109/TBCAS.2018.2889040
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Neuromodulation (NM) is the alteration of nervous tissue function through targeted delivery of a stimulus, such as electrical stimulation, into the affected neurological sites in the body. We present a bidirectional NM interface that features 100 mV(pp) linear input range and ability to sense data concurrent with stimulation (without blanking). The system includes a flexible 8-driver-to-64-contact custom waveform stimulator able to deliver up to 5.1 mA per driver and a 64-contact sensing unit with online blind artifact rejection unit. This artifact rejection unit removes stimulation artifacts from recorded data and allows extraction of neural biomarkers. The NM interface also features an efficient, integrated power management unit that can support various power delivery options. The proposed 64-contact interface satisfies design requirements of human-grade brain implants at unprecedented level of electronic miniaturization compared to state-of-the-art.
引用
收藏
页码:38 / 55
页数:18
相关论文
共 47 条
[1]   64-Channel UWB Wireless Neural Vector Analyzer SOC With a Closed-Loop Phase Synchrony-Triggered Neurostimulator [J].
Abdelhalim, Karim ;
Jafari, Hamed Mazhab ;
Kokarovtseva, Larysa ;
Velazquez, Jose Luis Perez ;
Genov, Roman .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2013, 48 (10) :2494-2510
[2]  
[Anonymous], 2012, VERCISE PC NEURAL NA
[3]  
[Anonymous], 2013, New medtronic deep brain stimulation system the first to sense and record brain activity while delivering therapy
[4]  
[Anonymous], 2015, RNS system user manual
[5]  
[Anonymous], 2012, RIPPLE GRAPEVINE
[6]  
[Anonymous], 2009, ACTIVE RC NEUROSTIMU
[7]   Massively-Parallel Neuromonitoring and Neurostimulation Rodent Headset With Nanotextured Flexible Microelectrodes [J].
Bagheri, Arezu ;
Gabran, S. R. I. ;
Salam, Muhammad Tariqus ;
Velazquez, Jose Luis Perez ;
Mansour, Raafat R. ;
Salama, M. M. A. ;
Genov, Roman .
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2013, 7 (05) :601-609
[8]  
Basir-Kazeruni S, 2017, I IEEE EMBS C NEUR E, P186, DOI 10.1109/NER.2017.8008322
[9]   A 16-Channel Patient-Specific Seizure Onset and Termination Detection SoC With Impedance-Adaptive Transcranial Electrical Stimulator [J].
Bin Altaf, Muhammad Awais ;
Zhang, Chen ;
Yoo, Jerald .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2015, 50 (11) :2728-2740
[10]   An HV-CMOS Integrated Circuit for Neural Stimulation in Prosthetic Applications [J].
Bisoni, Lorenzo ;
Carboni, Caterina ;
Raffo, Luigi ;
Carta, Nicola ;
Barbaro, Massimo .