Experimental validation of a deforming grid 4D dose calculation for PBS proton therapy

被引:30
作者
Krieger, Miriam [1 ,2 ]
Klimpki, Grischa [1 ,2 ]
Fattori, Giovanni [1 ]
Hrbacek, Jan [1 ]
Oxley, David [1 ]
Safai, Sairos [1 ]
Weber, Damien C. [1 ,3 ]
Lomax, Antony J. [1 ,2 ]
Zhang, Ye [1 ]
机构
[1] Paul Scherrer Inst, Ctr Proton Therapy, CH-5232 Villigen, Switzerland
[2] Swiss Fed Inst Technol, Dept Phys, CH-8092 Zurich, Switzerland
[3] Univ Hosp Zurich, Dept Radiat Oncol, CH-8091 Zurich, Switzerland
基金
瑞士国家科学基金会;
关键词
proton therapy; 4D dose calculation; pencil beam scanning; liver tumour; SCANNED PARTICLE THERAPY; MOTION MITIGATION; BEAM THERAPY; EXPERIMENTAL-VERIFICATION; RESPIRATORY MOTION; LOG FILES; RECONSTRUCTION; DELIVERY; SIMULATION; INTERPLAY;
D O I
10.1088/1361-6560/aaad1e
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of this study was to verify the temporal accuracy of the estimated dose distribution by a 4D dose calculation (4DDC) in comparison to measurements. A single-field plan (0.6 Gy), optimised for a liver patient case (CTV volume: 403cc), was delivered to a homogeneous PMMA phantom and measured by a high resolution scintillating-CCD system at two water equivalent depths. Various motion scenarios (no motion and motions with amplitude of 10 mm and two periods: 3.7 s and 4.4 s) were simulated using a 4D Quasar phantom and logged by an optical tracking system in real-time. Three motion mitigation approaches (single delivery, 6x layered and volumetric rescanning) were applied, resulting in 10 individual measurements. 4D dose distributions were retrospectively calculated in water by taking into account the delivery log files (retrospective) containing information on the actually delivered spot positions, fluences, and time stamps. Moreover, in order to evaluate the sensitivity of the 4DDC inputs, the corresponding prospective 4DDCs were performed as a comparison, using the estimated time stamps of the spot delivery and repeated periodical motion patterns. 2D gamma analyses and dose-difference-histograms were used to quantify the agreement between measurements and calculations for all pixels with >5% of the maximum calculated dose. The results show that a mean gamma score of 99.2% with standard deviation 1.0% can be achieved for 3%/3 mm criteria and all scenarios can reach a score of more than 95%. The average area with more than 5% dose difference was 6.2%. Deviations due to input uncertainties were obvious for single scan deliveries but could be smeared out once rescanning was applied. Thus, the deforming grid 4DDC has been demonstrated to be able to predict the complex patterns of 4D dose distributions for PBS proton therapy with high dosimetric and geometric accuracy, and it can be used as a valid clinical tool for 4D treatment planning, motion mitigation selection, and eventually 4D optimisation applications if the correct temporal information is available.
引用
收藏
页数:11
相关论文
共 42 条
[1]   Experimental verification of IMPT treatment plans in an anthropomorphic phantom in the presence of delivery uncertainties [J].
Albertini, F. ;
Casiraghi, M. ;
Lorentini, S. ;
Rombi, B. ;
Lomax, A. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (14) :4415-4431
[2]   A 4D dose computation method to investigate motion interplay effects in scanned ion beam prostate therapy [J].
Ammazzalorso, F. ;
Jelen, U. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (11) :N91-N99
[3]   Treatment log files as a tool to identify treatment plan sensitivity to inaccuracies in scanned proton beam delivery [J].
Belosi, Maria Francesca ;
van der Meer, Robert ;
de Acilu Laa, Paz Garcia ;
Bolsi, Alessandra ;
Weber, Damien C. ;
Lomax, Antony J. .
RADIOTHERAPY AND ONCOLOGY, 2017, 125 (03) :514-519
[4]   Comparative study of layered and volumetric rescanning for different scanning speeds of proton beam in liver patients [J].
Bernatowicz, K. ;
Lomax, A. J. ;
Knopf, A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (22) :7905-7920
[5]   Advanced treatment planning using direct 4D optimisation for pencil-beam scanned particle therapy [J].
Bernatowicz, Kinga ;
Zhang, Ye ;
Perrin, Rosalind ;
Weber, Damien C. ;
Lomax, Antony J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (16) :6595-6609
[6]   Four-Dimensional Dose Reconstruction for Scanned Proton Therapy Using Liver 4DCT-MRI [J].
Bernatowicz, Kinga ;
Peroni, Marta ;
Perrin, Rosalind ;
Weber, Damien C. ;
Lomax, Antony .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 95 (01) :216-223
[7]   Motion in radiotherapy: particle therapy [J].
Bert, C. ;
Durante, M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (16) :R113-R114
[8]   Quantification of interplay effects of scanned particle beams and moving targets [J].
Bert, Christoph ;
Groezinger, Sven O. ;
Rietzel, Eike .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (09) :2253-2265
[9]   4D treatment planning for scanned ion beams [J].
Bert, Christoph ;
Rietzel, Eike .
RADIATION ONCOLOGY, 2007, 2 (1)
[10]   Scanned carbon beam irradiation of moving films: comparison of measured and calculated response [J].
Bert, Christoph ;
Richter, Daniel ;
Durante, Marco ;
Rietzel, Eike .
RADIATION ONCOLOGY, 2012, 7