Fracture and migration of implantable venous access port catheters: Cause analysis and management of 4 cases

被引:10
作者
Xiao, Shu-ping [1 ]
Xiong, Bin [1 ]
Chu, Jun [2 ]
Li, Xiao-fang [1 ]
Yao, Qi [1 ]
Zheng, Chuan-sheng [1 ]
机构
[1] Huazhong Univ Sci & Technol, Union Hosp, Dept Intervent Radiol, Tongji Med Coll, Wuhan 430022, Peoples R China
[2] Shanghai Jiao Tong Univ, Dept Surg, Shanghai Childrens Med Ctr, Shanghai 200127, Peoples R China
关键词
implantable venous access port; catheter migration; intervention; nurse care; INTERNAL JUGULAR-VEIN; PINCH-OFF SYNDROME; PERCUTANEOUS RETRIEVAL; EMBOLIZATION; DEVICES; RARE;
D O I
10.1007/s11596-015-1504-4
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
This study aimed to investigate the causes and managements of the fractures and migrations of the implantable venous access port catheter (IVAPC). The fracture or migration of IVAPC occurred in 4 patients who were treated between May 2012 and January 2014 in Union Hospital, Wuhan, China. The port catheter leakage was found in 2 cases during drug infusion. Catheters that dislodged to the superior vena cava and right atrium were confirmed by port angiogram. The two dislodged catheters were successfully retrieved by interventional procedures. Catheter fracture occurred in two cases during port removal. One catheter was eventually removed from the subclavian vein through right clavicle osteotomy and subclavian venotomy, and the other removed by external jugular venotomy. Flushing the port in high pressure and injury of the totally implantable venous access port (TIVP) during implantation are usually responsible for catheter displacement. Interventional retrieval procedure can be used if the catheter dislodges to the vena cava and right atrium. Catheter fracture may occur during removal if clipping syndrome occurs or the catheter is sutured very tight during implantation.
引用
收藏
页码:763 / 765
页数:3
相关论文
共 18 条
  • [11] The mechanisms of failure of totally implantable central venous access system: Analysis of 73 cases with fracture of catheter
    Lin, C. H.
    Wu, H. S.
    Chan, D. C.
    Hsieh, C. B.
    Huang, M. H.
    Yu, J. C.
    [J]. EJSO, 2010, 36 (01): : 100 - 103
  • [12] Percutaneous retrieval of 20 centrally dislodged Port-A catheter fragments
    Liu, JC
    Tseng, HS
    Chen, CY
    Chern, MS
    Chang, CY
    [J]. CLINICAL IMAGING, 2004, 28 (03) : 223 - 229
  • [13] Catheter fracture: a rare complication of totally implantable venous devices
    Maisey, NR
    Sacks, N
    Johnston, SRD
    [J]. BREAST, 2003, 12 (04) : 287 - 289
  • [14] Is catheter rupture rare after totally implantable access port implantation via the right internal jugular vein? Report of a case
    Nagasawa, Yoshinobu
    Shimizu, Tomoharu
    Sonoda, Hiromichi
    Chou, Hirotomi
    Mekata, Eiji
    Tani, Tohru
    [J]. SURGERY TODAY, 2014, 44 (07) : 1346 - 1349
  • [15] Ultrasound-guided infraclavicular axillary vein puncture is effective to avoid pinch-off syndrome: a long-term follow-up study
    Osawa, Hideki
    Hasegawa, Junichi
    Yamakawa, Kazuma
    Matsunami, Nobuki
    Mikata, Shoki
    Shimizu, Junzo
    Kim, Yong Kook
    Morishima, Hirotaka
    Hirota, Masaki
    Souma, Yoshihito
    Kim, Ho Min
    Sawada, Genta
    Nezu, Riichiro
    [J]. SURGERY TODAY, 2013, 43 (07) : 745 - 750
  • [16] A Case of Catheter Fracture of a Totally Implantable Access Port Introduced Through the Right Internal Jugular Vein
    Shimizu, Tomoharu
    Mekata, Eiji
    Murata, Satoshi
    Yamamoto, Takashi
    Tani, Tohru
    [J]. JOURNAL OF SURGICAL ONCOLOGY, 2011, 103 (05) : 460 - 461
  • [17] Talwar V, 2003, J Vasc Access, V4, P158
  • [18] Percutaneous retrieval of PICC fractures via the femoral vein in six cancer patients
    Wang, Qi
    Xiong, Bin
    Zheng, ChuanSheng
    Feng, GanSheng
    Liang, Ming
    Liang, HuiMin
    [J]. JOURNAL OF VASCULAR ACCESS, 2015, 16 (01) : 47 - 51