A 50μW Fully Differential Interface Amplifier With a Current Steering Class AB Output Stage for PPG and NIRS Recordings

被引:15
作者
Song, Shuang [1 ]
Lin, Qiuyang [1 ,2 ]
van Hoof, Chris [1 ,2 ]
van Helleputte, Nick [1 ]
机构
[1] imec, BECHS, B-3001 Heverlee, Belgium
[2] Katholieke Univ Leuven, Dept ESAT MICAS, B-3000 Leuven, Belgium
关键词
Interface amplifier; near infrared spectroscopy (NIRS); photoplethysmography (PPG); low-power; POWER; SOC; PHOTOPLETHYSMOGRAPHY; SYSTEM; ECG;
D O I
10.1109/TCSII.2020.3002952
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This brief presents a low-power fully differential optical sensor interface amplifier for photoplethysmography (PPG) and near infrared spectroscopy (NIRS) recordings. The proposed amplifier employs a stacked current reuse input stage that provides a DC biasing voltage for the photodiode, reducing the parasitic capacitance. The output stage exploits a fully differential current steering class AB topology with a self-regulated common mode voltage, providing a 70 mu A input current range with minimized quiescent current. Moreover, the interface amplifier can be reconfigured as either single-ended mode suitable for photo detectors biased at a high voltage or fully differential mode for those biased at a low voltage. This amplifier is implemented in a standard 0.18 mu m CMOS process and characterized experimentally. Measurement results show that, with a 1.5V supply, the amplifier can provide up to 1V biasing voltage for the photodiode and consumes a static power of around 50 mu W. It achieves a dynamic range of 97dB in fully differential mode with an input referred noise of 67pArms in 32Hz band and 102dB in single -ended mode with an input referred noise of 582pArms.
引用
收藏
页码:1564 / 1568
页数:5
相关论文
共 21 条
[1]   Photoplethysmography [J].
Alian, Aymen A. ;
Shelley, Kirk H. .
BEST PRACTICE & RESEARCH-CLINICAL ANAESTHESIOLOGY, 2014, 28 (04) :395-406
[2]   Photoplethysmography and its application in clinical physiological measurement [J].
Allen, John .
PHYSIOLOGICAL MEASUREMENT, 2007, 28 (03) :R1-R39
[3]  
[Anonymous], 2014, OSRAM SFH 7050 APPL
[4]   A Fully Integrated High-Sensitivity Wide Dynamic Range PPG Sensor With an Integrated Photodiode and an Automatic Dimming Control LED Driver [J].
Atef, Mohamed ;
Wang, Min ;
Wang, Guoxing .
IEEE SENSORS JOURNAL, 2018, 18 (02) :652-659
[5]   A Sub-mW Fully-Integrated Pulse Oximeter Front-End [J].
Glaros, Konstantinos N. ;
Drakakis, Emmanuel M. .
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2013, 7 (03) :363-375
[6]   Near-infrared diffuse optical tomography [J].
Hielscher, AH ;
Bluestone, AY ;
Abdoulaev, GS ;
Klose, AD ;
Lasker, J ;
Stewart, M ;
Netz, U ;
Beuthan, J .
DISEASE MARKERS, 2002, 18 (5-6) :313-337
[7]  
Kim HG, 2017, ESSCIRC 2017 - 43RD IEEE EUROPEAN SOLID STATE CIRCUITS CONFERENCE, P55, DOI 10.1109/ESSCIRC.2017.8094524
[8]   A Multi(bio)sensor Acquisition System With Integrated Processor, Power Management, 8 x 8 LED Drivers, and Simultaneously Synchronized ECG, BIO-Z, GSR, and Two PPG Readouts [J].
Konijnenburg, Mario ;
Stanzione, Stefano ;
Yan, Long ;
Jee, Dong-Woo ;
Pettine, Julia ;
van Wegberg, Roland ;
Kim, Hyejung ;
van Liempd, Chris ;
Fish, Ram ;
Schuessler, James ;
de Groot, Harmke ;
Van Hoof, Chris ;
Yazicioglu, Refet Firat ;
Van Helleputte, Nick .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2016, 51 (11) :2584-2595
[9]  
Lin QY, 2019, SYMP VLSI CIRCUITS, pC58, DOI [10.23919/vlsic.2019.8778004, 10.23919/VLSIC.2019.8778004]
[10]   A 1-V 8.1-μW PPG-Recording Front-End With >92-dB DR Using Light-to-Digital Conversion With Signal-Aware DC Subtraction and Ambient Light Removal [J].
Marefat, Fatemeh ;
Erfani, Reza ;
Mohseni, Pedram .
IEEE SOLID-STATE CIRCUITS LETTERS, 2020, 3 (01) :17-20