Systematic analysis of bias and variability of morphologic features for lung lesions in computed tomography

被引:11
|
作者
Hoye, Jocelyn [1 ,2 ,3 ,4 ]
Solomon, Justin [1 ,2 ,3 ,4 ]
Sauer, Thomas J. [1 ,2 ,3 ,4 ]
Robins, Marthony [1 ,2 ,3 ,4 ]
Samei, Ehsan [1 ,2 ,3 ,4 ]
机构
[1] Duke Univ, Carl E Ravin Adv Imaging Labs, Durham, NC 27708 USA
[2] Duke Univ, Med Phys Grad Program, Durham, NC 27708 USA
[3] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
[4] Duke Univ, Med Ctr, Durham, NC 27708 USA
关键词
morphology; lung lesions; quantitative imaging; computed tomography; imaging conditions; RADIOMICS; RECONSTRUCTION; IMAGES; MODEL;
D O I
10.1117/1.JMI.6.1.013504
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We propose to characterize the bias and variability of quantitative morphology features of lung lesions across a range of computed tomography (CT) imaging conditions. A total of 15 lung lesions were simulated (five in each of three spiculation classes: low, medium, and high). For each lesion, a series of simulated CT images representing different imaging conditions were synthesized by applying three-dimensional blur and adding correlated noise based on the measured noise and resolution properties of five commercial multislice CT systems, representing three dose levels (CTDIvol of 1.90, 3.75, 7.50 mGy), three slice thicknesses (0.625, 1.25, 2.5 mm), and 33 clinical reconstruction kernels from five clinical scanners. The images were segmented using three segmentation algorithms and each algorithm was evaluated by computing a Sorensen-Dice coefficient between the ground truth and the segmentation. A series of 21 shape-based morphology features were extracted from both "ground truth" (i.e., preblur without noise) and "image rendered" lesions (i.e., postblur and with noise). For each morphology feature, the bias was quantified by comparing the percentage relative error in the morphology metric between the imaged lesions and the ground-truth lesions. The variability was characterized by calculating the average coefficient of variation averaged across repeats and imaging conditions. The active contour segmentation had the highest average Dice coefficient of 0.80 followed by 0.63 for threshold, and 0.39 for fuzzy c-means. The bias of the features was segmentation algorithm and feature-dependent, with sharper kernels being less biased and smoother kernels being more biased in general. The feature variability from simulated images ranged from 0.30% to 10% for repeats of the same condition and from 0.74% to 25.3% for different lesions in the same spiculation class. In conclusion, the bias of morphology features is dependent on the acquisition protocol in combination with the segmentation algorithm used and the variability is primarily dependent on the segmentation algorithm. (C) 2019 Society of Photo-Optical Instrumentation Engineers (SPIE)
引用
收藏
页数:14
相关论文
共 50 条
  • [31] Comprehensive Computed Tomography Radiomics Analysis of Lung Adenocarcinoma for Prognostication
    Lee, Geewon
    Park, Hyunjin
    Sohn, Insuk
    Lee, Seung-Hak
    Song, So Hee
    Kim, Hyeseung
    Lee, Kyung Soo
    Shim, Young Mog
    Lee, Ho Yun
    ONCOLOGIST, 2018, 23 (07) : 806 - 813
  • [32] TEXTURE ANALYSIS OF ADENOMATOUS AND METASTATIC ADRENAL LESIONS ON NATIVE AND CONTRAST-ENHANCED COMPUTED TOMOGRAPHY
    Belyanova, Magdalena
    Krupev, Martin
    COMPTES RENDUS DE L ACADEMIE BULGARE DES SCIENCES, 2023, 76 (10): : 1617 - 1624
  • [33] Study on Computed Tomography Guided Fine Needle Aspiration Cytology of Peripheral Lung Lesions
    Das, Krishnendu
    JOURNAL OF EVOLUTION OF MEDICAL AND DENTAL SCIENCES-JEMDS, 2019, 8 (31): : 2475 - 2479
  • [34] Extrapolation in the analysis of lung aeration by computed tomography: a validation study
    Reske, Andreas W.
    Rau, Anna
    Reske, Alexander P.
    Koziol, Manja
    Gottwald, Beate
    Alef, Michaele
    Ionita, Jean-Claude
    Spieth, Peter M.
    Hepp, Pierre
    Seiwerts, Matthias
    Beda, Alessandro
    Born, Silvia
    Scheuermann, Gerik
    Amato, Marcelo B. P.
    Wrigge, Hermann
    CRITICAL CARE, 2011, 15 (06):
  • [35] Modern radiopharmaceuticals for lung cancer imaging with positron emission tomography/computed tomography scan: A systematic review
    Theodoropoulos, Athanasios S.
    Gkiozos, Ioannis
    Kontopyrgias, Georgios
    Charpidou, Adrianni
    Kotteas, Elias
    Kyrgias, George
    Tolia, Maria
    SAGE OPEN MEDICINE, 2020, 8
  • [36] Primary solid lung cancerous nodules with different sizes: computed tomography features and their variations
    Zhi-gang Chu
    Yan Zhang
    Wang-jia Li
    Qi Li
    Yi-neng Zheng
    Fa-jin Lv
    BMC Cancer, 19
  • [37] Primary solid lung cancerous nodules with different sizes: computed tomography features and their variations
    Chu, Zhi-gang
    Zhang, Yan
    Li, Wang-jia
    Li, Qi
    Zheng, Yi-neng
    Lv, Fa-jin
    BMC CANCER, 2019, 19 (01)
  • [38] Computed tomography imaging and clinical features of congenital hepatoblastoma A retrospective analysis
    Li, Li
    Liu, Wen
    Wen, Rong
    Jin, Ke
    MEDICINE, 2020, 99 (31) : E21174
  • [39] Low dose computed tomography of the lung for detection and grading of interstitial lung disease: A systematic simulation study
    Ley, S.
    Fidler, L.
    Schenk, H.
    Durand, M.
    Marras, T.
    Paul, N.
    Shapera, S.
    Mittoo, S.
    PULMONOLOGY, 2021, 27 (01): : 14 - 25
  • [40] Quantifying invasiveness of clinical stage IA lung adenocarcinoma with computed tomography texture features
    Qiu, Zhen-Bin
    Zhang, Chao
    Chu, Xiang-Peng
    Cai, Fei-Yue
    Yang, Xue-Ning
    Wu, Yi-Long
    Zhong, Wen-Zhao
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2022, 163 (03) : 805 - +