Ultrasonography-based 2D motion-compensated HIFU sonication integrated with reference-free MR temperature monitoring: a feasibility study ex vivo

被引:39
作者
Auboiroux, Vincent [1 ]
Petrusca, Lorena [1 ]
Viallon, Magalie [2 ]
Goget, Thomas [1 ]
Becker, Christoph D. [2 ]
Salomir, Rares [1 ,2 ]
机构
[1] Univ Geneva, Fac Med, CH-1211 Geneva 4, Switzerland
[2] Univ Hosp Geneva, Dept Radiol, Geneva, Switzerland
关键词
GUIDED FOCUSED ULTRASOUND; BONE METASTASES; SURGERY MRGFUS; ABLATION; PALLIATION; THERAPY; LIVER; PAIN;
D O I
10.1088/0031-9155/57/10/N159
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Magnetic resonance imaging (MRI) and ultrasonography have been used simultaneously in this ex vivo study for the image-guidance of high intensity focused ultrasound (HIFU) treatment in moving tissue. A ventilator-driven balloon produced periodic and non-rigid (i.e. breathing-like) motion patterns in phantoms. MR-compatible ultrasound (US) imaging enabled near real-time 2D motion tracking based on optical flow detection, while near-harmonic reference-free proton resonance frequency shift (PRFS) MR thermometry (MRT) was used to monitor the thermal buildup on line. Reference-free MRT was applied to gradient-echo echo-planar imaging phase maps acquired at the frame rate of 250 to 300 ms/slice with voxel size 1.25 x 1.25 x 5 mm(3). The MR-US simultaneous imaging was completely free of mutual interferences while minor RF interferences from the HIFU device were detected in the far field of the US images. The effective duty-cycle of the HIFU sonication was close to 100 % and no off-interval was required to temporally decouple it from the ultrasonography. The motion compensation of the HIFU sonication was achieved with an 8 Hz frame rate and sub-millimeter spatial accuracy, both for single-focus mode and for an iterated multi-foci line scan. Near harmonic reference-less PRFSMRT delivered motion-robust thermal maps perpendicular or parallel to the HIFU beam (0.7 degrees C precision, 0.5 degrees C absolute accuracy). Out-of-plane motion compensation was not addressed in this study.
引用
收藏
页码:N159 / N171
页数:13
相关论文
共 41 条
[1]  
[Anonymous], P ISMRM 19 SCI M EXH
[2]  
[Anonymous], ULTRASON S
[3]  
[Anonymous], APPL PHYS LETT
[4]  
[Anonymous], 1999, INTEL CORPORATION MI
[5]   An MR-compliant phased-array HIFU transducer with augmented steering range, dedicated to abdominal thermotherapy [J].
Auboiroux, Vincent ;
Dumont, Erik ;
Petrusca, Lorena ;
Viallon, Magalie ;
Salomir, Rares .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (12) :3563-3582
[6]  
Bracewell Ronald N., 2000, The Fourier Transform and Its Applications
[7]   MR-guided focused ultrasound surgery (MRgFUS) for the palliation of pain in patients with bone metastases - preliminary clinical experience [J].
Catane, R. ;
Beck, A. ;
Inbar, Y. ;
Rabin, T. ;
Shabshin, N. ;
Hengst, S. ;
Pfeffer, R. M. ;
Hanannel, A. ;
Dogadkin, O. ;
Liberman, B. ;
Kopelman, D. .
ANNALS OF ONCOLOGY, 2007, 18 (01) :163-167
[8]   Thermal ablation of uterine fibroids using MR-guided focused ultrasound-a truly non-invasive treatment modality [J].
Chapman, Alexander ;
ter Haar, Gail .
EUROPEAN RADIOLOGY, 2007, 17 (10) :2505-2511
[9]   Rapid motion correction in MR-guided high-intensity focused ultrasound heating using real-time ultrasound echo information [J].
de Oliveira, Philippe Lourenco ;
de Senneville, Baudouin Denis ;
Dragonu, Lulius ;
Moonen, Chrit T. W. .
NMR IN BIOMEDICINE, 2010, 23 (09) :1103-1108
[10]   Real-time adaptive methods for treatment of mobile organs by MRI-controlled high-intensity focused ultrasound [J].
de Senneville, Baudouin Denis ;
Mougenot, Charles ;
Moonen, Chrit T. W. .
MAGNETIC RESONANCE IN MEDICINE, 2007, 57 (02) :319-330