Fracture of totally implanted central venous access devices: a propensity-score-matched comparison of risks for Groshong silicone versus polyurethane catheters

被引:10
作者
Kojima, Soichi [1 ]
Hiraki, Takao [1 ]
Gobara, Hideo [1 ]
Iguchi, Toshihiro [1 ]
Fujiwara, Hiroyasu [1 ]
Matsui, Yusuke [1 ]
Mitsuhashi, Toshiharu [2 ]
Kanazawa, Susumu [1 ]
机构
[1] Okayama Univ, Sch Med, Dept Radiol, 2-5-1 Kita Ku Shikata Cho, Okayama 7008558, Japan
[2] Okayama Univ, Sch Med, Dept Epidemiol, Okayama, Japan
基金
日本学术振兴会;
关键词
Breakage; Central venous catheter; Fracture; Polyurethane; Risk; Silicone; INTERNAL JUGULAR-VEIN; COMPLICATIONS; BREAKAGE; RATES; PORT;
D O I
10.5301/jva.5000606
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Purpose: To evaluate retrospectively the fracture risk of totally implanted venous access devices connected to Groshong silicone (SC) versus polyurethane (PU) catheters, inserted via the internal jugular vein. Materials and methods: The study population comprised 384 SC and 221 PU central venous catheters implanted via the internal jugular vein. The presence of catheter fracture was evaluated. Variables possibly related to catheter fracture were evaluated. First, in order to determine the factors associated with fracture, fracture rates were compared with the log-rank test between the two groups divided by each of the variables. Then, in order to adjust for potential confounders, propensity-score matching of the variables was employed in the two catheter groups. Finally, the rates of fracture were compared between the two propensity-score-matched catheter groups. Results: There were 16 cases of catheter fracture, for an overall fracture percentage of 2.6% (16/605). All 16 cases of fracture occurred in the SC catheter group. Smaller patient body mass index (p = 0.039), deeper catheter tip position (p = 0.022), and SC catheters (p = 0.019) were significantly associated with fracture. With the propensity-score- matching method, 180 cases were selected in each catheter group. Comparison of the two propensity-score- matched groups showed that fracture rates for SC catheters remained significantly (p = 0.018) higher than those for PU catheters. Conclusions: Ports connected to Groshong SC catheters - when implanted via the internal jugular vein - posed a higher risk of fracture than did ports connected to PU catheters.
引用
收藏
页码:535 / 541
页数:7
相关论文
共 21 条
[1]   Fractures of totally implantable central venous ports: more than fortuity. A three-year single center experience [J].
Balsorano, Paolo ;
Galducci, Giulia ;
De Fanti, Ilaria ;
Evans, Samuel Kagan ;
De Gaudio, Angelo Raffaele ;
Pelagatti, Cecilia .
JOURNAL OF VASCULAR ACCESS, 2014, 15 (05) :391-395
[2]  
Chan L. L., 1999, Annals Academy of Medicine Singapore, V28, P481
[3]   Percutaneous retrieval of dislodged totally implantable central venous access system in 92 cases: Experience in a single hospital [J].
Cheng, Cheng-Chung ;
Tsai, Tsung-Neng ;
Yang, Chung-Chi ;
Han, Chih-Lu .
EUROPEAN JOURNAL OF RADIOLOGY, 2009, 69 (02) :346-350
[4]   Silicone and Polyurethane Tunneled Infusion Catheters: A Comparison of Durability and Breakage Rates [J].
Cohen, Aaron B. ;
Dagli, Mandeep ;
Stavropoulos, S. William, Jr. ;
Mondschein, Jeffrey I. ;
Soulen, Michael C. ;
Shlansky-Goldberg, Richard D. ;
Solomon, Jeffrey A. ;
Chittams, Jesse L. ;
Trerotola, Scott O. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2011, 22 (05) :638-641
[5]  
Di Giacomo Michele, 2009, Br J Nurs, V18, P8
[6]   PINCH-OFF SYNDROME - A COMPLICATION OF IMPLANTABLE SUBCLAVIAN VENOUS ACCESS DEVICES [J].
HINKE, DH ;
ZANDTSTASTNY, DA ;
GOODMAN, LR ;
QUEBBEMAN, EJ ;
KRZYWDA, EA ;
ANDRIS, DA .
RADIOLOGY, 1990, 177 (02) :353-356
[7]   Tunneled infusion catheter breakage: Rrequency and repair kit outcomes [J].
Hwang, Freeman R. ;
Stavropoulos, S. William ;
Shlansky-Goldberg, Richard D. ;
Mondschein, Jeffrey I. ;
Patel, Aalpen A. ;
Solomon, Jeffrey A. ;
Itkin, Maxim ;
Soulen, Michael C. ;
Chittams, Jesse L. ;
Trerotola, Scott O. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2008, 19 (02) :201-206
[8]   Complications of central venous catheterization [J].
Kusminsky, Roberto E. .
JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 2007, 204 (04) :681-696
[9]   The mechanisms of failure of totally implantable central venous access system: Analysis of 73 cases with fracture of catheter [J].
Lin, C. H. ;
Wu, H. S. ;
Chan, D. C. ;
Hsieh, C. B. ;
Huang, M. H. ;
Yu, J. C. .
EJSO, 2010, 36 (01) :100-103
[10]   Comparison of technical success and outcome of tunneled catheters inserted via the jugular and subclavian approaches [J].
Macdonald, S ;
Watt, AJB ;
McNally, D ;
Edwards, RD ;
Moss, JG .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (02) :225-231