An investigation of the validity of the Lifeshirt in comparison to standard polysomnography in the detection of obstructive sleep apnea

被引:16
|
作者
Goodrich, Suanne [1 ]
Orr, William C. [1 ,2 ]
机构
[1] Lynn Hlth Sci Inst, Oklahoma City, OK 73112 USA
[2] Univ Oklahoma, Hlth Sci Ctr, Oklahoma City, OK 73104 USA
关键词
Apnea; Measurement; Airflow; Respiratory inductance plethysmography; Polysomnography; Portable sleep study; DIAGNOSIS; AGREEMENT; ACCESS;
D O I
10.1016/j.sleep.2007.11.003
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background: Due to the cost and waiting times for traditional sleep studies, there is great interest in finding alternatives for the diagnosis of sleep apnea. Several alternatives exist, including ambulatory devices. Our laboratory utilized one such device, a form-fitting vest called the Lifeshirt, and compared its accuracy to that of polysomnography (PSG). Methods: PSG was performed with simultaneous utilization of the Lifeshirt on 50 individuals who met screening criteria for obstructive sleep apnea. Participants came to the sleep laboratory approximately 2 h before their normal bedtime. A sleep technician prepared each participant for PSG and Lifeshirt monitoring. "Lights Out" occurred when the participant was ready for bed, and time in bed was standardized to approximately 7 h. PSGs were scored by experienced personnel in our laboratory, while Lifeshirt data were sent electronically to the Lifeshirt manufacturer for analysis. The major variable that was compared between PSG and Lifeshirt was the apnea hypopnea index (AHI), or the number of apneas and hypopneas per hour. Due to incomplete data on two participants, analyses were completed on 48 individuals. Results: Sensitivity of the Lifeshirt ranged from .85 (AHI of >= 5) to 1.00 (AHI of >= 25). Specificity ranged from .67 to 1.00. Using the Bland-Altman technique of determining agreement, the mean difference between the AHI of the Lifeshirt and PSG was 1.02 (+/- 16.36). When these values are plotted, every case falls within the limits of agreement, with one exception. Conclusions: The Lifeshirt compared favorably with PSG and could be used with considerable confidence for the screening of patients with suspected obstructive sleep apnea. (C) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:118 / 122
页数:5
相关论文
共 50 条
  • [1] An investigation of the accuracy of the Lifeshirt in comparison to standard polysomnography
    Orr, W. C.
    Goodrich, S.
    SLEEP, 2006, 29 : A348 - A349
  • [2] Comparison of a simple obstructive sleep apnea screening device with standard in-laboratory polysomnography
    Assefa, Samson Z.
    Diaz-Abad, Montserrat
    Korotinsky, Arkady
    Tom, Sarah E.
    Scharf, Steven M.
    SLEEP AND BREATHING, 2016, 20 (02) : 537 - 541
  • [3] Comparison of a simple obstructive sleep apnea screening device with standard in-laboratory polysomnography
    Samson Z. Assefa
    Montserrat Diaz-Abad
    Arkady Korotinsky
    Sarah E. Tom
    Steven M. Scharf
    Sleep and Breathing, 2016, 20 : 537 - 541
  • [4] COMPARISON OF A DISPOSABLE HOME SLEEP APNEA TEST TO POLYSOMNOGRAPHY IN PATIENTS REFERRED FOR OBSTRUCTIVE SLEEP APNEA INVESTIGATION.
    Tellez, Daniela
    Uhles, Matthew
    Alisic, Sabina
    Pinson, Mary-Beth
    Willes, Leslee
    Decker, Meredith
    Benjafield, Adam
    SLEEP, 2022, 45 : A305 - A305
  • [5] A comparison of polysomnography and the WatchPAT in the diagnosis of obstructive sleep apnea
    Pang, Kenny P.
    Gourin, Christine G.
    Terris, David J.
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2007, 137 (04) : 665 - 668
  • [6] Comparison of OPPO Watch Sleep Analyzer and Polysomnography for Obstructive Sleep Apnea Screening
    Zhou, Guangxin
    Zhao, Wei
    Zhang, Yi
    Zhou, Wenli
    Yan, Haizhou
    Wei, Yongli
    Tang, Yuming
    Zeng, Zijing
    Cheng, Hanrong
    NATURE AND SCIENCE OF SLEEP, 2024, 16 : 125 - 141
  • [7] Investigation for the dates of polysomnography from obstructive sleep apnea-hypopnea
    Hou, Yajun
    Gao, Feng
    Zhao, Tingting
    INFORMATION TECHNOLOGY, 2015, : 235 - 237
  • [8] Is the OSA-18 predictive of obstructive sleep apnea: Comparison to polysomnography
    Ishman, Stacey L.
    Yang, Christina J.
    Cohen, Aliza P.
    Benke, James R.
    Meinzen-Derr, Jareen K.
    Anderson, Rebecca M.
    Madden, Marie E.
    Tabangin, Meredith E.
    LARYNGOSCOPE, 2015, 125 (06): : 1491 - 1495
  • [9] Computed Tomography Characterization and Comparison With Polysomnography for Obstructive Sleep Apnea Evaluation
    Chousangsuntorn, Khaisang
    Bhongmakapat, Thongchai
    Apirakkittikul, Navarat
    Sungkarat, Witaya
    Supakul, Nucharin
    Laothamatas, Jiraporn
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2018, 76 (04) : 854 - 872
  • [10] COMPARISON OF LABORATORY POLYSOMNOGRAPHY AND AN AMBULATORY SLEEP APNEA MONITOR FOR DETECTING OBSTRUCTIVE SLEEP APNEA IN PREGNANT WOMEN
    Sharkey, K. M.
    Millman, R. P.
    Bourjeily, G.
    SLEEP, 2011, 34 : A321 - A321