Cluster-Randomized Trial of Devices to Prevent Catheter-Related Bloodstream Infection

被引:35
作者
Brunelli, Steven M. [1 ]
Van Wyck, David B. [2 ]
Njord, Levi [2 ]
Ziebol, Robert J. [3 ]
Lynch, Laurie E. [3 ]
Killion, Douglas P. [3 ]
机构
[1] DaVita Clin Res, Minneapolis, MN USA
[2] DaVita Inc, Denver, CO USA
[3] Pursuit Vasc Inc, 6901 E Fish Lake Rd,Ste 166, Maple Grove, MN 55369 USA
来源
JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY | 2018年 / 29卷 / 04期
关键词
TRISODIUM CITRATE; LOCKING SOLUTIONS; SPILLAGE;
D O I
10.1681/ASN.2017080870
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Central venous catheters (CVCs) contribute disproportionately to bloodstream infection (BSI) and, by extension, to infection-related hospitalization, mortality, and health care costs in patients undergoing dialysis. Recent product advancements may reduce BSIs, but a sufficiently powered comparative-effectiveness study is needed to facilitate evidence-based patient care decisions. In a 13-month, prospective, cluster-randomized, open-label trial, we compared BSI rates in facilities using ClearGuard HD antimicrobial barrier caps (ClearGuard group) with those in facilities using Tego hemodialysis connectors plus Curos disinfecting caps (Tego+Curos group). Forty DaVita dialysis facilities in the United States were pair-matched by BSI rate, number of patients using CVCs, and geographic location, and then cluster randomized 1:1. We enrolled all adult patients undergoing dialysis with CVCs at these facilities, except those allergic to heparin or chlorhexidine. Overall, 1671 patients participated in the study, accruing >183,000 CVC-days. The study outcome was positive blood culture (PBC) rate as an indicator of BSI rate. We calculated results at the cluster level and adjusted for the facility cluster effect. During a 3-month run-in period immediately before study interventions, the groups had similar BSI rates (P=0.8). During the 13-month intervention period that immediately followed, the ClearGuard group had a BSI rate significantly lower than that of the Tego+Curos group (0.28 versus 0.75 PBCs per 1000 CVC-days, respectively; P=0.001). No device-related adverse events were reported. In conclusion, compared with Tego connectors plus Curos caps, ClearGuard HD antimicrobial barrier caps significantly lowered the rate of catheter-related BSIs in patients undergoing hemodialysis using CVCs, representing an important advancement in hemodialysis patient care.
引用
收藏
页码:1336 / 1343
页数:8
相关论文
共 26 条
[1]  
3M, 2017, PROT PORTS ENS PEAC
[2]  
[Anonymous], 2017, NAT HEALTHC SAF NETW
[3]   Use of the Tego needlefree connector is associated with reduced incidence of catheter-related bloodstream infections in hemodialysis patients [J].
Brunelli, Steven M. ;
Njord, Levi ;
Hunt, Abigail E. ;
Sibbel, Scott P. .
INTERNATIONAL JOURNAL OF NEPHROLOGY AND RENOVASCULAR DISEASE, 2014, 7 :131-139
[4]  
Centers for Disease Control and Prevention, DIAL EV SURV MAN
[5]  
Centers for Disease Control and Prevention, 2015, DIAL EV PROT
[6]  
[FDA Food and Drug Administration], 2000, 14T00T16 FDA
[7]  
Food and Drug Administration, 2013, K131060 CLEARGUARD H
[8]  
Grudzinski Alexa, 2015, Can J Kidney Health Dis, V2, P13, DOI 10.1186/s40697-015-0040-2
[9]   Dialysis Catheter-Related Bloodstream Infections: A Cluster-Randomized Trial of the ClearGuard HD Antimicrobial Barrier Cap [J].
Hymes, Jeffrey L. ;
Mooney, Ann ;
Van Zandt, Carly ;
Lynch, Laurie ;
Ziebol, Robert ;
Killion, Douglas .
AMERICAN JOURNAL OF KIDNEY DISEASES, 2017, 69 (02) :220-227
[10]   A meta-analysis of hemodialysis catheter locking solutions in the prevention of catheter-related infection [J].
Jaffer, Yasmin ;
Selby, Nicholas M. ;
Taal, Maarten W. ;
Fluck, Richard J. ;
McIntyre, Christopher W. .
AMERICAN JOURNAL OF KIDNEY DISEASES, 2008, 51 (02) :233-241