A magnetic fluid seal for rotary blood pumps: Behaviors of magnetic fluids in a magnetic fluid seal

被引:29
作者
Mitamura, Yoshinori [1 ]
Yano, Tetsuya [2 ]
Nakamura, Wataru [3 ]
Okamoto, Eiji [3 ]
机构
[1] Hokkaido Univ, Grad Sch Informat Sci & Technol, Sapporo, Hokkaido, Japan
[2] Akita Prefectural Univ, Dept Machine Intelligence & Syst Engn, Fac Syst Sci & Technol, Yuri Honjo, Japan
[3] Tokai Univ, Sch Biol Sci & Engn, Dept Human Sci & Informat, Sapporo, Hokkaido, Japan
关键词
Magnetic fluid; magnetic fluid seal; rotary blood pump; computational fluid dynamics; MECHANICAL CIRCULATORY SUPPORT; VENTRICULAR ASSIST DEVICE; LIQUID;
D O I
10.3233/BME-120733
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A magnetic fluid (MF) seal has excellent durability. The performance of an MF seal, however, has been reported to decrease in liquids (several days). We have developed an MF seal that has a shield mechanism. The seal was perfect for 275 days in water. To investigate the effect of a shield, behaviors of MFs in a seal in water were studied both experimentally and computationally. (a) Two kinds of MF seals, one with a shield and one without a shield, were installed in a centrifugal pump. Behaviors of MFs in the seals in water were observed with a video camera and high-speed microscope. In the seal without a shield, the surface of the water in the seal waved and the turbulent flow affected behaviors of the MFs. In contrast, MFs rotated stably in the seal with a shield in water even at high rotational speeds. (b) Computational fluid dynamics analysis revealed that a stationary secondary flow pattern in the seal and small velocity difference between magnetic fluid and water at the interface. These MF behaviors prolonged the life of an MF seal in water.
引用
收藏
页码:63 / 74
页数:12
相关论文
共 13 条
[1]   A prospective, multicenter trial of the VentrAssist left ventricular assist device for bridge to transplant: Safety and efficacy [J].
Esmore, Donald ;
Kaye, David ;
Spratt, Phillip ;
Larbalestier, Robert ;
Ruygrok, Peter ;
Tsui, Steven ;
Meyers, Deborah ;
Fiane, Arnt E. ;
Woodard, John .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2008, 27 (06) :579-588
[2]  
Kitahora T., 1994, T JWRI B, V60, P3086, DOI [10.1299/kikaib.60.30, DOI 10.1299/KIKAIB.60.30]
[3]   SEALING LIQUIDS WITH MAGNETIC LIQUIDS [J].
KURFESS, J ;
MULLER, HK .
JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 1990, 85 (1-3) :246-252
[4]   Design optimization of seal structure for sealing liquid by magnetic fluids [J].
Liu, TG ;
Cheng, YS ;
Yang, ZY .
JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2005, 289 :411-414
[5]   Application of a magnetic fluid seal to rotary blood pumps [J].
Mitamura, Y. ;
Arioka, S. ;
Sakota, D. ;
Sekine, K. ;
Azegami, M. .
JOURNAL OF PHYSICS-CONDENSED MATTER, 2008, 20 (20)
[6]   A magnetic fluid seal for rotary blood pumps: effects of seal structure on long-term performance in liquid [J].
Mitamura, Yoshinori ;
Takahashi, Sayaka ;
Amari, Shuichi ;
Okamoto, Eiji ;
Murabayashi, Shun ;
Nishimura, Ikuya .
JOURNAL OF ARTIFICIAL ORGANS, 2011, 14 (01) :23-30
[7]   A magnetic fluid seal for rotary blood pumps: Long-term performance in liquid [J].
Mitamura, Yoshinori ;
Takahashi, Sayaka ;
Amari, Shuichi ;
Okamoto, Eiji ;
Murabayashi, Shun ;
Nishimura, Ikuya .
12TH INTERNATIONAL CONFERENCE ON MAGNETIC FLUIDS ICMF12, 2010, 9 :229-233
[8]   Long-term circulatory support with the DuraHeart™ Mag-Lev centrifugal left ventricular assist system for advanced heart failure patients eligible to transplantation:: European experiences [J].
Nojiri, C. ;
Fey, O. ;
Jaschke, F. ;
Medevedev, A. ;
Munjal, D. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2008, 27 (02) :S245-S245
[9]   Extended Mechanical Circulatory Support With a Continuous-Flow Rotary Left Ventricular Assist Device [J].
Pagani, Francis D. ;
Miller, Leslie W. ;
Russell, Stuart D. ;
Aaronson, Keith D. ;
John, Ranjit ;
Boyle, Andrew J. ;
Conte, John V. ;
Bogaev, Roberta C. ;
MacGillivray, Thomas E. ;
Naka, Yoshifumi ;
Mancini, Donna ;
Massey, H. Todd ;
Chen, Leway ;
Klodell, Charles T. ;
Aranda, Juan M. ;
Moazami, Nader ;
Ewald, Gregory A. ;
Farrar, David J. ;
Frazier, O. Howard .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 54 (04) :312-321
[10]  
Rosensweig R.E., 1997, FERROHYDRODYNAMICS, P202