Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures

被引:146
作者
Chida, K
Saito, H
Otani, H
Kohzuki, M
Takahashi, S
Yamada, S
Shirato, K
Zuguchi, M
机构
[1] Tohoku Univ, Sch Hlth Sci, Dept Radiol Technol, Aoba Ku, Sendai, Miyagi 9808575, Japan
[2] Tohoku Univ, Grad Sch Med, Dept Radiol, Sendai, Miyagi 9808574, Japan
[3] Tohoku Univ, Grad Sch Med, Dept Cardiovasc Med, Sendai, Miyagi 9808574, Japan
[4] Tohoku Univ, Grad Sch Med, Dept Internal Med & Rehabil Sci, Sendai, Miyagi 9808574, Japan
关键词
angiography; cardiovascular imaging; interventional radiology; radiation dose; radiation injury;
D O I
10.2214/AJR.04.1653
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. Real-time maximum dose monitoring of the skin is unavailable on many of the X-ray machines that are used for cardiac intervention procedures. Therefore, some reports have recommended that physicians record the fluoroscopic time for patients undergoing fluoroscopically guided intervention procedures. However, the relationship between the fluoroscopic time and the maximum radiation skin dose is not clear. This article describes the correlation between the maximum radiation skin dose and fluoroscopic time for patients undergoing cardiac intervention procedures. In addition, we examined whether the correlations between maximum radiation skin dose and body weight, fluoroscopic time, and dose-area product (DAP) were useful for estimating the maximum skin dose during cardiac intervention procedures. MATERIALS AND METHODS. Two hundred consecutive cardiac intervention procedures were studied: 172 percutaneous coronary interventions and 28 cardiac radiofrequency catheter ablation (RFCA) procedures. The patient skin dose and DAP were measured using Caregraph with skin-dose-mapping software. RESULTS. For the RFCA procedures, we found a good correlation between the maximum radiation skin dose and fluoroscopic time (r = 0.801, p < 0.0001), whereas we found a poor correlation between the maximum radiation skin dose and fluoroscopic time for the percutaneous coronary intervention procedures (r = 0.628, p < 0.0001). There was a strong correlation between the maximum radiation skin dose and DAP in RFCA procedures (r = 0.942, p < 0.0001). There was also a significant correlation between the maximum radiation skin dose and DAP (r = 0.724, p < 0.0001) and weight-fluoroscopic time product (WFP) (r = 0.709, p < 0.0001) in percutaneous coronary intervention procedures. CONCLUSION. The correlation between the maximum radiation skin dose with DAP is more striking than that with fluoroscopic time in both RFCA and percutaneous coronary intervention procedures. We recommend that physicians record the DAP when it can be monitored and that physicians record the fluoroscopic time when DAP cannot be monitored for estimating the maximum patient skin dose in RFCA procedures. For estimating the maximum patient skin dose in percutaneous coronary intervention procedures, we also recommend that physicians record DAP when it can be monitored and that physicians record WFP when DAP cannot be monitored.
引用
收藏
页码:774 / 778
页数:5
相关论文
共 23 条
  • [1] [Anonymous], 2001, ANN ICRP
  • [2] [Anonymous], REC INF PAT MED REC
  • [3] Baim DS, 2000, GROSSMANS CARDIAC CA, P211
  • [4] Radiation-induced skin injuries and fluoroscopy
    Berlin, L
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (01) : 21 - 25
  • [5] RADIATION EXPOSURE DURING RADIOFREQUENCY CATHETER ABLATION OF ACCESSORY ATRIOVENTRICULAR CONNECTIONS
    CALKINS, H
    NIKLASON, L
    SOUSA, J
    ELATASSI, R
    LANGBERG, J
    MORADY, F
    [J]. CIRCULATION, 1991, 84 (06) : 2376 - 2382
  • [6] Does digital acquisition reduce patients' skin dose in cardiac interventional procedures? An experimental study
    Chida, K
    Saito, H
    Zuguchi, M
    Shirotori, K
    Kumagai, S
    Nakayama, H
    Matsubara, K
    Kohzuki, M
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 183 (04) : 1111 - 1114
  • [7] Real-time quantification and display of skin radiation during coronary angiography and intervention
    den Boer, A
    de Feijter, PJ
    Serruys, PW
    Roelandt, JRTC
    [J]. CIRCULATION, 2001, 104 (15) : 1779 - 1784
  • [8] Comparison of four techniques to estimate radiation dose to skin during angiographic and interventional radiology procedures
    Fletcher, DW
    Miller, DL
    Balter, S
    Taylor, MA
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2002, 13 (04) : 391 - 397
  • [9] Hwang E, 1998, CATHETER CARDIO DIAG, V43, P367, DOI 10.1002/(SICI)1097-0304(199804)43:4<367::AID-CCD1>3.0.CO
  • [10] 2-6