Independent verification of gantry angle for pre-treatment VMAT QA using EPID

被引:17
作者
Adamson, Justus [1 ]
Wu, Qiuwen [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiat Oncol, Durham, NC 27707 USA
关键词
MODULATED ARC THERAPY; DOSIMETRIC VERIFICATION; QUALITY-ASSURANCE; IMRT; RADIATION; DELIVERY; IMAT; TOOL;
D O I
10.1088/0031-9155/57/20/6587
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We propose a method to incorporate independent verification of gantry angle for electronic portal imaging device (EPID)-based pre-treatment quality assurance (QA) of clinical volumetric modulated arc therapy (VMAT) plans. Gantry angle is measured using projections in the EPID of a custom phantom placed on the couch and the treatment plan is modified so as to be incident on the phantom with a portion of the beam that is collimated in the clinical plan. For our implementation, collimator and couch angles were set to zero and the inferior jaw and two most inferior multi-leaf collimator pairs were opened for the entire QA delivery. A phantom containing five gold coils was used to measure the gantry rotation through which each portal image was acquired. We performed the EPID QA for ten clinical plans and evaluated accuracy of gantry angle measurement, scatter incident on the imager due to the phantom, inter-image pixel linearity and inter-and intra-image noise. The gantry angle could be measured to within 0.0 +/- 0.3 degrees for static gantry and 0.2 +/- 0.2 degrees for arc acquisitions. Scatter due to the presence of the phantom was negligible. The procedure was shown to be feasible and adds gantry angle to the treatment planning parameters that can be verified by EPID-based pre-treatment VMAT QA.
引用
收藏
页码:6587 / 6600
页数:14
相关论文
共 50 条
  • [31] In-vivo and pre-treatment quality assurance software validation and verification
    Vanzi, E.
    Reversi, L.
    Giani, G.
    Arilli, C.
    Bonomo, P.
    Casati, M.
    Compagnucci, A.
    Greto, D.
    Marrazzo, L.
    Scoccianti, S.
    Pallotta, S.
    Bucciolini, M.
    Talamonti, C.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 671 - 673
  • [32] Investigation into the radiobiological consequences of pre-treatment verification imaging with megavoltage X-rays in radiotherapy
    Hyland, W. B.
    McMahon, S. J.
    Butterworth, K. T.
    Cole, A. J.
    King, R. B.
    Redmond, K. M.
    Prise, K. M.
    Hounsell, A. R.
    McGarry, C. K.
    BRITISH JOURNAL OF RADIOLOGY, 2014, 87 (1036)
  • [33] Comparison of MLC error sensitivity of various commercial devices for VMAT pre-treatment quality assurance
    Saito, Masahide
    Sano, Naoki
    Shibata, Yuki
    Kuriyama, Kengo
    Komiyama, Takafumi
    Marino, Kan
    Aoki, Shinichi
    Ashizawa, Kazunari
    Yoshizawa, Kazuya
    Onishi, Hiroshi
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (03): : 87 - 93
  • [34] Examination of the properties of IMRT and VMAT beams and evaluation against pre-treatment quality assurance results
    Crowe, S. B.
    Kairn, T.
    Middlebrook, N.
    Sutherland, B.
    Hill, B.
    Kenny, J.
    Langton, C. M.
    Trapp, J. V.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (06) : 2587 - 2601
  • [35] Retrospective analysis of portal dosimetry pre-treatment quality assurance of intracranial SRS/SRT VMAT treatment plans
    Osei, Ernest
    Graves, Sarah
    Darko, Johnson
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2022, 21 (04) : 540 - 552
  • [36] Practical guidelines for routine intensity-modulated radiotherapy verification: pre-treatment verification with portal dosimetry and treatment verification with in vivo dosimetry
    Vinall, A. J.
    Williams, A. J.
    Currie, V. E.
    Van Esch, A.
    Huyskens, D.
    BRITISH JOURNAL OF RADIOLOGY, 2010, 83 (995) : 949 - 957
  • [37] Dosimetric verification of pre-treatment intensity modulated radiation therapy in the commissioning process
    Esmaili, G.
    Mandavi, S. R.
    Nikoofar, A. R.
    Fadavi, P.
    Ameri, A.
    Nazari, V
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2018, 16 (04): : 493 - 497
  • [38] Evaluation of gantry angle during respiratory-gated VMAT using triggered kilovoltage x-ray image
    Miyasaka, Ryohei
    Saitoh, Hidetoshi
    Kawachi, Toru
    Katayose, Tetsurou
    Cho, SangYong
    Yamauchi, Ryohei
    Hara, Ryusuke
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (08): : 98 - 104
  • [39] Comparison of the Efficacy of 2D Dosimetry Systems in the Pre-treatment Verification of IMRT
    Hong, Chae Seon
    Lim, Jongsoo
    Ju, Sang Gyu
    Shin, Eunhyuk
    Han, Youngyih
    Ahn, Yong Chan
    RADIATION ONCOLOGY JOURNAL, 2009, 27 (02): : 91 - 102
  • [40] DVH analysis using a transmission detector and model-based dose verification system as a comprehensive pretreatment QA tool for VMAT plans: Clinical experience and results
    Yoosuf, Ahamed B. Mohamed
    AlShehri, Salem
    Alhadab, Abdulrahman
    Alqathami, Mamdooh
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (11): : 80 - 87