In this report we present the technical features of Gantry 2, the new second generation scanning system of PSI. On the basis of the experience and success with the first prototype, Gantry 1, built in the 90s for introducing pencil beam scanning and IMPT into the field of proton therapy, we have recently implemented a new system capable of offering much faster repainted conformal scanning for being able to treat moving targets with scanning under image guidance, the next challenge in the field of proton therapy. The new technical developments are conducted in parallel to the ongoing basic commissioning of Gantry 2, which should go into operation with usual discrete spot scanning for treating static targets in 2013. The innovative layout of Gantry 2 and the integration in the treatment area of the basic equipment for image guidance are presented. Noteworthy are the sliding CT within reach of the patient table and the unique new Beam's-Eye-View X-ray fluoroscopy system for taking images in the beam direction synchronized with the proton beam delivery. The first preliminary results with the development of much faster scanning modes look very encouraging. We can change the beam energy with the beam line within 80 ms for typical 0.5 cm range steps. We can deliver whole fluence-shaped energy layers within a time of the order of 100 ms. Dose lines are painted by changing the velocity of the scan magnets. The instantaneous dose rate of the pencil beam can be varied dynamically as well. The dose is precisely controlled with a feedback loop connecting the main gantry beam monitor with a vertical deflector plate at the ion source. These new fast scanning modes should be used for providing scanning with repainting, gating and tracking for treating moving targets. The goal is to develop pencil beam scanning as a universal beam delivery solution capable of treating optimally all possible clinical indications for proton therapy. Scanning could then completely replace the old beam delivery methods based on passive scattering from the market. The long term projects of Gantry 2 should represent the new contributions of PSI to the proton therapy field in the next 5-10 years, by providing direct translational cancer research from the physics laboratory into industry and clinics.
引用
收藏
页码:196 / 206
页数:11
相关论文
共 17 条
  • [1] MAGNETIC SCANNING SYSTEM FOR HEAVY-ION THERAPY
    HABERER, T
    BECHER, W
    SCHARDT, D
    KRAFT, G
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 330 (1-2) : 296 - 305
  • [2] Kawachi K, 1983, Acta Radiol Suppl, V364, P81
  • [3] Special report: Workshop on 4D-treatment planning in actively scanned particle therapy-Recommendations, technical challenges, and future research directions
    Knopf, Antje
    Bert, Christoph
    Heath, Emily
    Nill, Simeon
    Kraus, Kim
    Richter, Daniel
    Hug, Eugen
    Pedroni, Eros
    Safai, Sairos
    Albertini, Francesca
    Zenklusen, Silvan
    Boye, Dirk
    Sohn, Matthias
    Soukup, Martin
    Sobotta, Benjamin
    Lomax, Antony
    [J]. MEDICAL PHYSICS, 2010, 37 (09) : 4608 - 4614
  • [4] Lomax A.J., 2008, PROTON CHARGED PARTI
  • [5] Treatment planning and verification of proton therapy using spot scanning:: Initial experiences
    Lomax, AJ
    Böhringer, T
    Bolsi, A
    Coray, D
    Emert, F
    Goitein, G
    Jermann, M
    Lin, SX
    Pedroni, E
    Rutz, H
    Stadelmann, O
    Timmermann, B
    Verwey, J
    Weber, DC
    [J]. MEDICAL PHYSICS, 2004, 31 (11) : 3150 - 3157
  • [6] Intensity modulated proton therapy: A clinical example
    Lomax, AJ
    Boehringer, T
    Coray, A
    Egger, E
    Goitein, G
    Grossmann, M
    Juelke, P
    Lin, S
    Pedroni, E
    Rohrer, B
    Roser, W
    Rossi, B
    Siegenthaler, B
    Stadelmann, O
    Stauble, H
    Vetter, C
    Wisser, L
    [J]. MEDICAL PHYSICS, 2001, 28 (03) : 317 - 324
  • [7] Experimental characterization and physical modelling of the dose distribution of scanned proton pencil beams
    Pedroni, E
    Scheib, S
    Böhringer, T
    Coray, A
    Grossmann, M
    Lin, S
    Lomax, A
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (03) : 541 - 561
  • [8] Pencil beam characteristics of the next-generation proton scanning gantry of PSI: design issues and initial commissioning results
    Pedroni, E.
    Meer, D.
    Bula, C.
    Safai, S.
    Zenklusen, S.
    [J]. EUROPEAN PHYSICAL JOURNAL PLUS, 2011, 126 (07): : 1 - 27
  • [9] THE 200-MEV PROTON THERAPY PROJECT AT THE PAUL-SCHERRER-INSTITUTE - CONCEPTUAL DESIGN AND PRACTICAL REALIZATION
    PEDRONI, E
    BACHER, R
    BLATTMANN, H
    BOHRINGER, T
    CORAY, A
    LOMAX, A
    LIN, SX
    MUNKEL, G
    SCHEIB, S
    SCHNEIDER, U
    TOUROVSKY, A
    [J]. MEDICAL PHYSICS, 1995, 22 (01) : 37 - 53
  • [10] PEDRONI E, 1981, TREATMENT PLANNING E, P60