Near infrared spectroscopy (NIRS) reflectance technology for determination of tocopherols in animal feeds

被引:9
作者
González-Martín, I [1 ]
Bustamante-Rangel, JMGCM [1 ]
Delgado-Zamarreño, MMD [1 ]
机构
[1] Fac Ciencias Quim, Dept Quim Analit Nutr & Bromatol, Salamanca 37008, Spain
关键词
vitamin E; tocopherols; near infrared spectroscopy; fibre-optic probe; animal feed;
D O I
10.1016/j.aca.2005.10.085
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The Vitamin E (alpha-, (beta+gamma)- and delta-tocopherol) contents present in feeds of different textures used in animal feed were analysed using near infrared spectroscopy (NIRS) technology together with a remote reflectance fibre-optic probe. The range of Vitamin E was 0.06 and 1.87 mg/100 g. The regression method employed was modified partial least squares (MPLS). The equations developed using the fiber-optic probe for 44 samples to determine the content of Vitamin E in feeds, in prediction, had multiple correlation coefficients (RSQ) and corrected standard errors (SEP (C)): 0.792, 0.223 mg/100 g for alpha-tocopherol, of 0.766 and 0.022 mg/100 g for (beta+gamma)-tocopherol, and of 0.789 and 0.112 mg/100 g for delta-tocopherol. The predicted values of Vitamin E in feeds using NIRS technology applying the fibre-optic probe directly on the sample with neither previous treatment nor manipulation are comparable to the values obtained using the chemical method that includes alkaline hydrolysis and extraction with hexane, of the vitamin from the unsaponifiable fraction before chromatographic determination. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:132 / 136
页数:5
相关论文
共 26 条
[1]   Determination of vitamins A and E in infant milk formulae by high-performance liquid chromatography [J].
AlbalaHurtado, S ;
NovellaRodriguez, S ;
VecianaNogues, MT ;
MarineFont, A .
JOURNAL OF CHROMATOGRAPHY A, 1997, 778 (1-2) :243-246
[2]   STANDARD NORMAL VARIATE TRANSFORMATION AND DE-TRENDING OF NEAR-INFRARED DIFFUSE REFLECTANCE SPECTRA [J].
BARNES, RJ ;
DHANOA, MS ;
LISTER, SJ .
APPLIED SPECTROSCOPY, 1989, 43 (05) :772-777
[3]   Analysis of vitamin E isomers in seeds and nuts with and without coupled hydrolysis by liquid chromatography and coulometric detection [J].
Delgado-Zamarreño, MM ;
Bustamante-Rangel, M ;
Sánchez-Pérez, A ;
Hernández-Méndez, J .
JOURNAL OF CHROMATOGRAPHY A, 2001, 935 (1-2) :77-86
[4]   DIRECTLY COUPLED SAMPLE TREATMENT HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY FOR ONLINE AUTOMATIC-DETERMINATION OF LIPOSOLUBLE VITAMINS IN MILK [J].
DELGADOZAMARRENO, MM ;
SANCHEZPEREZ, A ;
GOMEZPEREZ, MC ;
HERNANDEZMENDEZ, J .
JOURNAL OF CHROMATOGRAPHY A, 1995, 694 (02) :399-406
[5]   ON THE SCALES ASSOCIATED WITH NEAR-INFRARED REFLECTANCE DIFFERENCE SPECTRA [J].
DHANOA, MS ;
LISTER, SJ ;
BARNES, RJ .
APPLIED SPECTROSCOPY, 1995, 49 (06) :765-772
[6]   LINEARIZATION AND SCATTER-CORRECTION FOR NEAR-INFRARED REFLECTANCE SPECTRA OF MEAT [J].
GELADI, P ;
MACDOUGALL, D ;
MARTENS, H .
APPLIED SPECTROSCOPY, 1985, 39 (03) :491-500
[7]   Determination of fatty acids in the subcutaneous fat of Iberian breed swine by near infrared spectroscopy (NIRS) with a fibre-optic probe [J].
González-Martín, I ;
González-Pérez, C ;
Hernández-Méndez, J ;
Alvarez-García, N .
MEAT SCIENCE, 2003, 65 (02) :713-719
[8]   Determination of vitamin E in aquatic organisms by high-performance liquid chromatography with fluorescence detection [J].
Huo, JZ ;
Nelis, HJ ;
Lavens, P ;
Sorgeloos, P ;
DeLeenheer, AP .
ANALYTICAL BIOCHEMISTRY, 1996, 242 (01) :123-128
[10]  
Konings EJM, 1996, J AOAC INT, V79, P902