Safety and effectiveness of repeat arterial closure using the StarClose vascular closure device in patients with hepatic malignancy

被引:4
作者
Wang, Z. [1 ]
Li, X. [1 ]
Chen, J. [1 ]
Shi, H. [1 ]
Pan, J. [1 ]
Zhang, X. [1 ]
Jin, Z. [1 ]
机构
[1] Beijing Union Med Coll Hosp, Dept Radiol, Beijing 100730, Peoples R China
关键词
RANDOMIZED CONTROLLED-TRIAL; HEPATOCELLULAR-CARCINOMA; MANUAL COMPRESSION; ANGIOSEAL; EFFICACY; CHEMOEMBOLIZATION; TUMORS;
D O I
10.1016/j.crad.2013.04.001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
AIM: To evaluate the safety and effectiveness of the use of the StarClose vascular closure device for repeat arterial closure in patients with hepatic malignancy undergoing transarterial chemoembolization (TACE). MATERIALS AND METHODS: A retrospective analysis of patients with hepatic malignancy who had undergone arterial closure with the StarClose device was performed in Peking Union Medical College Hospital between January 2009 and March 2012. A total of 165 patients (94 men, 71 women; mean age 60.1 +/- 17.2 years) had arterial puncture closure after TACE (using a 5 F sheath). Percutaneous closure of the common femoral artery with the StarClose device was performed in accordance with the manufacturer's recommendations. The patients were examined for complications on follow-up. Device success was defined as haemostasis achieved immediately after StarClose device deployment with or without applying 3 min or less of manual compression. All the patients were evaluated clinically for wound complications and follow-up angiograms of the puncture site were obtained at subsequent TACE sessions. RESULTS: There were a total of 593 closures using StarClose devices in 165 patients who underwent TACE, including 194 (32.7%) single closures and 399 (67.3%) repeat closures for the same femoral access. The number of repeat closures for the same femoral access ranged from one to nine (median 2). Haemostasis was achieved in 571 closures, and success rate was 96.3% (96.9% in single closures and 96% in repeat closures). There was no statistically different haemostasis success rate between single closures or repeat closures. No severe complications occurred during follow-up. CONCLUSION: The repeat use of the StarClose device is safe and effective in patients with hepatic malignancy undergoing TACE. (C) 2013 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:E498 / E501
页数:4
相关论文
共 22 条
[1]   The safety and efficacy of Angioseal in therapeutic endovascular interventions [J].
Aksoy, M. ;
Becquemin, J. P. ;
Desgranges, P. ;
Allaire, E. ;
Kobeiter, H. .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2006, 32 (01) :90-93
[2]  
Ananthakrishnan Ashwin, 2006, Semin Intervent Radiol, V23, P47, DOI 10.1055/s-2006-939841
[3]   Sequential preoperative arterial and portal venous embolizations in patients with hepatocellular carcinoma [J].
Aoki, T ;
Imamura, H ;
Hasegawa, K ;
Matsukura, A ;
Sano, K ;
Sugawara, Y ;
Kokudo, N ;
Makuuchi, M .
ARCHIVES OF SURGERY, 2004, 139 (07) :766-774
[4]   Results of surgical and nonsurgical treatment for small-sized hepatocellular carcinomas: A retrospective and nationwide survey in Japan [J].
Arii, S ;
Yamaoka, Y ;
Futagawa, S ;
Inoue, K ;
Kobayashi, K ;
Kojiro, M ;
Makuuchi, M ;
Nakamura, Y ;
Okita, K ;
Yamada, R .
HEPATOLOGY, 2000, 32 (06) :1224-1229
[5]   Closure of the femoral artery after cardiac catheterization: A comparison of Angio-Seal, StarClose, and manual compression [J].
Deuling, J. H. H. ;
Vermeulen, R. P. ;
Anthonio, R. A. ;
van den Heuvel, A. F. M. ;
Jaarsma, T. ;
Jessurun, G. ;
de Smet, B. J. G. L. ;
Tan, E. S. ;
Zijistra, F. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2008, 71 (04) :518-523
[6]   Hepatocellular carcinoma - An epidemiologic view [J].
El-Serag, HB .
JOURNAL OF CLINICAL GASTROENTEROLOGY, 2002, 35 (05) :S72-S78
[7]   Safety and efficacy of StarClose SE Vascular Closure System in high-risk liver interventional oncology patients [J].
Gaba, Ron C. ;
Parvinian, Ahmad ;
Trinos, Estrellita M. ;
Padayao, Suzanne V. ;
Francisco, Rechel M. ;
Yap, Felix Y. ;
Knuttinen, M. Grace ;
Owens, Charles A. ;
Bui, James T. .
JOURNAL OF VASCULAR ACCESS, 2012, 13 (04) :415-420
[8]   The StarClose® Vascular Closure System:: Interventional results from the CLIP study [J].
Hermiller, J. B. ;
Simonton, C. ;
Hinohara, T. ;
Lee, D. ;
Cannon, L. ;
Mooney, M. ;
O'Shaughnessy, C. ;
Carlson, H. ;
Fortuna, R. ;
Zapien, M. ;
Fletcher, D. R. ;
DiDonato, K. ;
Chou, T. M. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2006, 68 (05) :677-683
[9]  
Hermiller James, 2005, J Invasive Cardiol, V17, P504
[10]   Safety and Effectiveness of Repeat Arterial Closure Using the AngioSeal Device in Patients with Hepatic Malignancy [J].
Hieb, Robert A. ;
Neisen, Melissa J. ;
Hohenwalter, Eric J. ;
Molnar, Jim A. ;
Rilling, William S. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2008, 19 (12) :1704-1708