US probes: Risk of cross infection and ways to reduce it - Comparison of cleaning methods

被引:46
作者
Fowler, C
McCracken, D
机构
[1] Univ Wales Hosp, Dept Microbiol, Cardiff CF4 4XW, S Glam, Wales
[2] Univ Wales Hosp, Dept Radiol, Cardiff CF4 4XW, S Glam, Wales
关键词
ultrasound; (US); biological effects;
D O I
10.1148/radiology.213.1.r99au41299
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
After their use at ultrasonography (US) in the intensive therapy unit, probes were used to directly inoculate blood agar plates before and after various cleaning procedures. The uncleaned probes transmitted large numbers of clinically important microbes. Simple cleaning methods were effective in reducing transmission among certain patients: fit patients, double paper wipe; patients at risk of contracting infection, single paper wipe followed by alcohol wipe; patients with a potential source of infection, single paper wipe followed by alcohol wipe.
引用
收藏
页码:299 / 300
页数:2
相关论文
共 3 条
[1]   CAN ULTRASOUND PROBES AND COUPLING GEL BE A SOURCE OF NOSOCOMIAL INFECTION IN PATIENTS UNDERGOING SONOGRAPHY - AN IN-VIVO AND IN-VITRO STUDY [J].
MURADALI, D ;
GOLD, WL ;
PHILLIPS, A ;
WILSON, S .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 164 (06) :1521-1524
[2]   ULTRASOUND SCANNING OF POSTOPERATIVE WOUNDS - THE RISKS OF CROSS-INFECTION [J].
SPENCER, P ;
SPENCER, RC .
CLINICAL RADIOLOGY, 1988, 39 (03) :245-246
[3]   Sonography machines as a source of infection [J].
Tesch, C ;
Froschle, G .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1997, 168 (02) :567-568