Shielding Artificially Increases the Attenuation of Water: Study of CT Gradient Attenuation Induced by Shielding (CT GAINS)

被引:3
作者
Goldin, Dmitry [1 ]
Emerson, Scott [1 ]
Farah, Michael C. [1 ]
Sokhandon, Farnoosh [1 ]
Nandalur, Kiran R. [1 ]
Kumar, Shawn [2 ]
Mezwa, Duane G. [1 ]
机构
[1] Beaumont Hlth Syst, Dept Radiol, Royal Oak, MI 48073 USA
[2] Wayne State Univ, Sch Med, Detroit, MI 48201 USA
关键词
Bismuth; shields; shielding; attenuation; water; computed tomography; BISMUTH BREAST SHIELD; COMPUTED-TOMOGRAPHY; IMAGE QUALITY; DOSE REDUCTION; RADIATION-EXPOSURE; FEMALE BREAST; NECK; MDCT;
D O I
10.1016/j.acra.2012.07.016
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: Quantitatively analyze the computed tomography (CT) attenuation effects caused by bismuth shields, which are used to reduce superficial organ dose. Materials and Methods: The solid water uniformity section of the American College of Radiology CT phantom was scanned with a modified chest CT protocol. Scans were performed with a bismuth breast shield in multiple configurations, emphasizing three clinically relevant orientations. Attenuation effects were measured as changes in mean Hounsfield unit (HU) values of equal midsagittal regions of interest (ROI). Multiple statistical techniques were used in regression analysis. Results: Bismuth shielding resulted in significant positive shifts of the expected Hounsfield unit values. The mean nonshielded CT attenuation was -0.16 +/- 0.75 HU. Based on the clinically relevant ROI distance from the shield (similar to 3-16 cm), the shielded values ranged from 43.8-4 HU, 45.8-10.1 HU, and 50.6-4.5 HU for shields 1, 2, and 3, respectively. All shield configurations displayed a statistically significant shift (P < .0001) at all distance ranges. The best fitting regression model was a quadratic-function of distance versus logarithmic function of HU. A prediction table of the approximate shift in water HU values as a function of ROI distance from the shield was generated per shield type from their respective close-fitting regressions. Conclusions: The data support the claim that bismuth shields increase the attenuation of water, which can cause inaccurate characterization of simple fluid, giving the appearance of complex fluid or even solid density. However, there is potential for anticipation of the attenuation effects to validate continued use of these shields for dose reduction.
引用
收藏
页码:231 / 237
页数:7
相关论文
共 18 条
  • [1] A Comprehensive Approach to CT Radiation Dose Reduction: One Institution's Experience
    Antypas, Elias J.
    Sokhandon, Farnoosh
    Farah, Michael
    Emerson, Scott
    Bis, Kostaki G.
    Tien, Hai
    Mezwa, Duane
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 197 (04) : 935 - 940
  • [2] Current concepts - Computed tomography - An increasing source of radiation exposure
    Brenner, David J.
    Hall, Eric J.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) : 2277 - 2284
  • [3] Dose reduction in multislice CT by means of bismuth shields: results of in vivo measurements and computed evaluation
    Catuzzo, P.
    Aimonetto, S.
    Fanelli, G.
    Marchisio, P.
    Meloni, T.
    Mistretta, L.
    Pasquino, M.
    Richetta, E.
    Tofani, S.
    [J]. RADIOLOGIA MEDICA, 2010, 115 (01): : 152 - 169
  • [4] Clinical usefulness of computed tomography study without contrast injection in the evaluation of acute pulmonary embolism
    Cobelli, R
    Zompatori, M
    De Luca, G
    Chiari, G
    Bresciani, P
    Marcato, C
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2005, 29 (01) : 6 - 12
  • [5] Colombo Paola, 2004, Radiol Med, V108, P560
  • [6] In-plane bismuth breast shields for pediatric CT: Effects on radiation dose and image quality using experimental and clinical data
    Fricke, BL
    Donnelly, LF
    Frush, DP
    Yoshizumi, T
    Varchena, V
    Poe, SA
    Lucaya, J
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 180 (02) : 407 - 411
  • [7] Frush Donald P, 2004, J Am Coll Radiol, V1, P113, DOI 10.1016/j.jacr.2003.11.012
  • [8] Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality
    Geleijns, J.
    Salvado Artells, M.
    Veldkamp, W. J. H.
    Lopez Tortosa, M.
    Calzado Cantera, A.
    [J]. EUROPEAN RADIOLOGY, 2006, 16 (10) : 2334 - 2340
  • [9] In-Plane Shielding for CT: Effect of Off-Centering, Automatic Exposure Control and Shield-to-Surface Distance
    Kalra, Mannudeep K.
    Dang, Pragya
    Singh, Sarabjeet
    Saini, Sanjay
    Shepard, Jo-Anne O.
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2009, 10 (02) : 156 - 163
  • [10] DOSE REDUCTION AND IMAGE QUALITY ASSESSMENT IN MDCT USING AEC (D-DOM & Z-DOM) AND IN-PLANE BISMUTH SHIELDING
    Lee, Kibaek
    Lee, Wonho
    Lee, Junhyup
    Lee, Boram
    Oh, Gyubum
    [J]. RADIATION PROTECTION DOSIMETRY, 2010, 141 (02) : 162 - 167