CMOS Biomedical Sensor with In Situ Gold Reference Electrode for Urine Detection Application

被引:9
作者
Juang, Ying-Zong [1 ]
Lin, Chen-Fu [1 ]
Tsai, Hann-Huei [1 ]
Liao, Hsin-Hao [1 ]
Wang, Ruey-Lue [2 ]
机构
[1] Natl Appl Res Labs, Natl Chip Implementat Ctr, Hsinchu, Taiwan
[2] Natl Kaohsiung Univ, Dept Elect Engn, Kaohsiung, Taiwan
来源
26TH EUROPEAN CONFERENCE ON SOLID-STATE TRANSDUCERS, EUROSENSOR 2012 | 2012年 / 47卷
关键词
ENFET;
D O I
10.1016/j.proeng.2012.09.317
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this paper, it presents the fully CMOS biomedical sensor with in situ gold reference electrode in the chip. The traditional Ag/AgCl reference electrode is too large to form a biomedical microsystem. Using the on chip gold reference electrode to replace the Ag/AgCl reference electrode has been studied. The ISFET device with native aluminum oxide is used to detect the hydrogen ions. The shape of gold reference electrode is important to get the stable results. To place the surrounding gold reference electrode nearby the sensing membrane is better than the bar-type gold electrode. The sensitivity and linearity of this pH sensor with in situ surrounding gold reference electrode are 56mV/pH and 0.975, respectively. This work demonstrates the optimum gold electrode could be integrated into CMOS-based biomedical sensors and performs the good results.
引用
收藏
页码:1005 / 1008
页数:4
相关论文
共 9 条
[1]   Study on separative structure of EnFET to detect acetylcholine [J].
Chi, LL ;
Yin, LT ;
Chou, JC ;
Chung, WY ;
Sun, TP ;
Hsiung, KP ;
Hsiung, SK .
SENSORS AND ACTUATORS B-CHEMICAL, 2000, 71 (1-2) :68-72
[2]   A novel SnO2/Al discrete gate ISFET pH sensor with CMOS standard process [J].
Chin, YL ;
Chou, JC ;
Sun, TP ;
Liao, HK ;
Chung, WY ;
Hsiung, SK .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 75 (1-2) :36-42
[3]   ISFET characteristics in CMOS and their application to weak inversion operation [J].
Georgiou, Pantelis ;
Toumazou, Christofer .
SENSORS AND ACTUATORS B-CHEMICAL, 2009, 143 (01) :211-217
[4]  
JUANG YZ, 2009, EUR 23 C, V1, P1067, DOI DOI 10.1016/J.PROCHE.2009.07.266
[5]   CMOS Open-Gate Ion-Sensitive Field-Effect Transistors for Ultrasensitive Dopamine Detection [J].
Li, Dong-Che ;
Yang, Po-Hung ;
Lu, Michael S. -C. .
IEEE TRANSACTIONS ON ELECTRON DEVICES, 2010, 57 (10) :2761-2767
[6]   Multiple type biosensors fabricated using the CMOS BioMEMS platform [J].
Tsai, Hann-Huei ;
Lin, Chen-Fu ;
Juang, Ying-Zong ;
Wang, I-Long ;
Lin, Yu-Cheng ;
Wang, Ruey-Lue ;
Lin, Hung-Yin .
SENSORS AND ACTUATORS B-CHEMICAL, 2010, 144 (02) :407-412
[7]   THE EXTENDED GATE CHEMICALLY SENSITIVE FIELD-EFFECT TRANSISTOR AS MULTI-SPECIES MICROPROBE [J].
VANDERSPIEGEL, J ;
LAUKS, I ;
CHAN, P ;
BABIC, D .
SENSORS AND ACTUATORS, 1983, 4 (02) :291-298
[8]  
Yang Haigang, 2005, 5 INT WORKSH SYST ON
[9]   Glucose ENFET doped with MnO2 powder [J].
Yin, LT ;
Chou, JC ;
Chung, WY ;
Sun, TP ;
Hsiung, KP ;
Hsiung, SK .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 76 (1-3) :187-192