NASCET Percent Stenosis Semi-Automated Versus Manual Measurement on CTA

被引:9
作者
Lian, Kevin [1 ]
White, Jeremy H. [1 ]
Bartlett, Eric S. [1 ]
Bharatha, Aditya [1 ]
Aviv, Richard I. [1 ]
Fox, Allan J. [1 ]
Symons, Sean P. [1 ]
机构
[1] Univ Toronto, Sunnybrook Hlth Sci Ctr, Div Neuroradiol, Dept Med Imaging, Toronto, ON M4N 3M5, Canada
关键词
CAROTID-ARTERY STENOSIS; ANGIOGRAPHY; REPRODUCIBILITY; ENDARTERECTOMY; QUANTIFICATION; OCCLUSION;
D O I
10.1017/S0317167100013482
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Purpose: To compare North American Symptomatic Carotid Endarterectomy Trial (NASCET) stenosis values and NASCET grade categorization (mild, moderate, severe) of semi-automated vessel analysis software versus manual measurements on computed tomography angiography (CTA). Methods: There were four observers. Two independently analyzed 8 I carotid artery CTAs using semi-automated vessel analysis software according to a blinded protocol. The software measured the narrowest stenosis in millimeters (mm), distal internal carotid artery (ICA) in mm, and calculated percent stenosis based on NASCET criteria. One of these two observers performed this task twice on each carotid, the second analysis was delayed two months in order to mitigate recall bias. Two other observers manually measured the narrowest stenosis in mm, distal ICA in mm, and calculated NASCET percent stenosis in a blinded fashion. The calculated NASCET stenoses were categorized into mild, moderate, or severe. Chi square and analysis of variance (ANOVA) were used to test for statistical differences. Results: ANOVA did not find a statistically significant difference in the mean percent stenosis when comparing the two manual measurements, the two semi-automated measurements, and the repeat semi-automated. Chi square demonstrated that the distribution of grades of stenosis were statistically different (p<0.05) between the manual and semi-automated grades. Semi-automated vessel analysis tended to underestimate the degree of stenosis compared to manual measurement. Conclusion: The mean percentage stenosis determined by semi-automated vessel analysis is not significantly different from manual measurement. However, when the data is categorized into mild, moderate and severe stenosis, there is a significant difference between semi-automated and manual measurements. The semi-automated software tends to underestimate the stenosis grade compared to manual measurement.
引用
收藏
页码:343 / 346
页数:4
相关论文
共 25 条
  • [21] Intra- and inter-observer reproducibility of volume measurement of knee cartilage segmented from the OAI MR image set using a novel semi-automated segmentation method
    Bae, K. T.
    Shim, H.
    Tao, C.
    Chang, S.
    Wang, J. H.
    Boudreau, R.
    Kwoh, C. K.
    OSTEOARTHRITIS AND CARTILAGE, 2009, 17 (12) : 1589 - 1597
  • [22] Individually adapted, interactive multiplanar reformations vs. semi-automated coronary segmentation and curved planar reformations for stenosis detection in coronary computed tomography angiography
    Anders, Katharina
    Ropers, Ulrike
    Kuettner, Axel
    Wechsel, Martin
    Daniel, Werner G.
    Uder, Michael
    Achenbach, Stephan
    EUROPEAN JOURNAL OF RADIOLOGY, 2011, 80 (01) : 89 - 95
  • [23] Intra- and interoperator variability of lobar pulmonary volumes and emphysema scores in patients with chronic obstructive pulmonary disease and emphysema: comparison of manual and semi-automated segmentation techniques
    Molinari, Francesco
    Pirronti, Tommaso
    Sverzellati, Nicola
    Diciotti, Stefano
    Amato, Michele
    Paolantonio, Guglielmo
    Gentile, Luigia
    Parapatt, George K.
    D'Argento, Francesco
    Kuhnigk, Jan-Martin
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2013, 19 (04) : 279 - 285
  • [24] Inter-observer and intra-observer reliability between manual segmentation and semi-automated segmentation for carotid vessel wall volume measurements on three-dimensional ultrasonography
    Chan, Chun Wai
    Chow, Sze Chai Christy
    Kwok, Man Hei
    Ngan, Ka Ching Tiffany
    Or, Tsun Hei
    Gunda, Simon Takadiyi
    Ying, Michael
    ULTRASONOGRAPHY, 2023, 42 (02) : 214 - 226
  • [25] Agreement between manual relaxometry and semi-automated scanner-based multi-echo Dixon technique for measuring liver T2*in a pediatric and young adult population
    Serai, Suraj D.
    Smith, Ethan A.
    Trout, Andrew T.
    Dillman, Jonathan R.
    PEDIATRIC RADIOLOGY, 2018, 48 (01) : 94 - 100