Variation of Cardiac Troponin I and T Measured with Sensitive Assays in Emergency Department Patients with Noncardiac Chest Pain

被引:31
作者
Scharnhorst, Volkher [1 ]
Krasznai, Krisztina [2 ]
van 't Veer, Marcel [2 ]
Michels, Rolf H. [2 ]
机构
[1] Catharina Hosp, Clin Lab, Eindhoven, Netherlands
[2] Catharina Hosp, Dept Cardiol, Eindhoven, Netherlands
关键词
BIOLOGICAL VARIATION; PERFORMANCE; VALIDATION;
D O I
10.1373/clinchem.2011.179903
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BACKGROUND: New-generation high-sensitivity assays for cardiac troponin have lower detection limits and less imprecision than earlier assays. Reference 99th-percentile cutoff values for these new assays are also lower, leading to higher frequencies of positive test results. When cardiac troponin concentrations are minimally increased, serial testing allows discrimination of myocardial infarction from other causes of increased cardiac troponin. We assessed various measures of short-term variation, including absolute concentration changes, reference change values (RCVs), and indices of individuality (11) for 2 cardiac troponin assays in emergency department (ED) patients. METHODS: We collected blood from patients presenting with cardiac chest pain upon arrival in the ED and 2, 6, and 12 h later. Cardiac troponin was measured with the high-sensitivity cardiac troponin T (hs-cTnT) assay (Roche Diagnostics) and a sensitive cTnI assay (Siemens Diagnostics). Cardiac troponin results from 67 patients without acute coronary syndrome or stable angina were used in calculating absolute changes in cardiac troponin, RCVs, and II. RESULTS: The 95th percentiles for absolute change in cardiac troponin were 8.3 ng/L for hs-cTnT and 28 ng/L for cTnI. Within-individual and total CVs were 11% and 14% for hs-cTnT and 18% and 21% for cTnI, respectively. RCVs were 38% (hs-cTnT) and 57% (cTnI). The corresponding log-normal RCVs were +46%/-32% for hs-cTnT and +76%/-43% for cTnI. II values were 0.31 (cTnI) and 0.12 (hs-cTnT). CONCLUSIONS: The short-term variations and IIs of cardiac troponin were low in ED patients free of ischemic myocardial necrosis. The detection of cardiac troponin variation exceeding reference thresholds can help to identify ED patients with acute myocardial necrosis whereas variation within these limits renders acute coronary syndrome unlikely. (C) 2012 American Association for Clinical Chemistry
引用
收藏
页码:1208 / 1214
页数:7
相关论文
共 17 条
[1]  
[Anonymous], SIEMENS MED SOLUTION
[2]   Value of Cardiac Troponin I Cutoff Concentrations below the 99th Percentile for Clinical Decision-Making [J].
Eggers, Kai M. ;
Jaffe, Allan S. ;
Lind, Lars ;
Venge, Per ;
Lindahl, Bertil .
CLINICAL CHEMISTRY, 2009, 55 (01) :85-92
[3]   Biological Variation and Reference Change Value of High-Sensitivity Troponin T in Healthy Individuals during Short and Intermediate Follow-up Periods [J].
Frankenstein, Lutz ;
Wu, Alan H. B. ;
Haller-Mayer, Klaus ;
Wians, Frank H., Jr. ;
Giannitsis, Evangelos ;
Katus, Hugo A. .
CLINICAL CHEMISTRY, 2011, 57 (07) :1068-1071
[4]   Biological variation of high sensitive Troponin T in stable heart failure patients with ischemic or dilated cardiomyopathy [J].
Frankenstein, Lutz ;
Remppis, Andrew ;
Giannitis, Evangelos ;
Frankenstein, Joerdis ;
Hess, Georg ;
Zdunek, Dietmar ;
Doesch, Andreas ;
Zugck, Christian ;
Katus, Hugo A. .
CLINICAL RESEARCH IN CARDIOLOGY, 2011, 100 (08) :633-640
[5]   Analytical Validation of a High-Sensitivity Cardiac Troponin T Assay [J].
Giannitsis, Evangelos ;
Kurz, Kerstin ;
Hallermayer, Klaus ;
Jarausch, Jochen ;
Jaffe, Allan S. ;
Katus, Hugo A. .
CLINICAL CHEMISTRY, 2010, 56 (02) :254-261
[6]   ESC Guidelines for the management of acute coronary syndromes in patients presenting without persistent ST-segment elevation [J].
Hamm, Christian W. ;
Bassand, Jean-Pierre ;
Agewall, Stefan ;
Bax, Jeroen ;
Boersma, Eric ;
Bueno, Hector ;
Caso, Pio ;
Dudek, Dariusz ;
Gielen, Stephan ;
Huber, Kurt ;
Ohman, Magnus ;
Petrie, Mark C. ;
Sonntag, Frank ;
Uva, Miguel Sousa ;
Storey, Robert F. ;
Wijns, William ;
Zahger, Doron .
EUROPEAN HEART JOURNAL, 2011, 32 (23) :2999-3054
[7]   Analytical characteristics of the Roche highly sensitive troponin T assay and its application to a cardio-healthy population [J].
Koerbin, G. ;
Tate, J. R. ;
Hickman, P. E. .
ANNALS OF CLINICAL BIOCHEMISTRY, 2010, 47 :524-528
[8]   Reference Population and Marathon Runner Sera Assessed by Highly Sensitive Cardiac Troponin T and Commercial Cardiac Troponin T and I Assays [J].
Mingels, Alma ;
Jacobs, Leo ;
Michielsen, Etienne ;
Swaanenburg, Joost ;
Wodzig, Will ;
van Dieijen-Visser, Marja .
CLINICAL CHEMISTRY, 2009, 55 (01) :101-108
[9]   Utility of Absolute and Relative Changes in Cardiac Troponin Concentrations in the Early Diagnosis of Acute Myocardial Infarction [J].
Reichlin, Tobias ;
Irfan, Affan ;
Twerenbold, Raphael ;
Reiter, Miriam ;
Hochholzer, Willibald ;
Burkhalter, Hanna ;
Bassetti, Stefano ;
Steuer, Stephan ;
Winkler, Katrin ;
Peter, Federico ;
Meissner, Julia ;
Haaf, Philip ;
Potocki, Mihael ;
Drexler, Beatrice ;
Osswald, Stefan ;
Mueller, Christian .
CIRCULATION, 2011, 124 (02) :136-U66
[10]   Rapid Detection of Myocardial Infarction With a Sensitive Troponin Test [J].
Scharnhorst, Volkher ;
Krasznai, Krisztina ;
van't Veer, Marcel ;
Michels, Rolf .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2011, 135 (03) :424-428