Evaluation of different sized blood sampling tubes for thromboelastometry, platelet function, and platelet count

被引:13
|
作者
Andreasen, Jo Bonding [2 ]
Pistor-Riebold, Thea Unger [3 ]
Knudsen, Ingrid Hell [3 ]
Ravn, Hanne Berg [2 ]
Hvas, Anne-Mette [1 ]
机构
[1] Aarhus Univ Hosp, Dept Clin Biochem, Ctr Haemophilia & Thrombosis, DK-8200 Aarhus N, Denmark
[2] Aarhus Univ Hosp, Dept Anaesthesiol & Intens Care, DK-8200 Aarhus N, Denmark
[3] Aarhus Univ, Fac Hlth Sci, Aarhus, Denmark
关键词
blood sampling; coagulation; paediatric; platelet aggregation; preanalytical variables; QUALITY STANDARDS; AGGREGOMETRY; TRANSFUSIONS; COLLECTION; TRANSPORT; EDTA; ROTEM(R); SYSTEM;
D O I
10.1515/cclm-2013-0836
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: To minimise the volume of blood used for diagnostic procedures, especially in children, we investigated whether the size of sample tubes affected whole blood coagulation analyses. Methods: We included 20 healthy individuals for rotational thromboelastometry (RoTEM (R)) analyses and compared three blood sampling tubes of different size: 1.8, 2.7, and 3.6 mL. All tubes were made of plastic and contained 3.2% sodium-citrate as anticoagulant. Platelet aggregation was investigated in 12 healthy individuals employing the Multiplate (R) Analyser comparing tubes of 3.6 mL and 1.8 mL. Platelet count was determined for each of the sampling tubes after 10, 60, and 120 min. Results: No significant differences were found in any of the RoTEM (R) measurements between the three sizes of tubes (p=0.07-0.53). Platelet aggregation was significantly lower when using smaller tubes (p=0.0004). The platelet count remained stable using a 3.6 mL tube during the entire observation period of 120 min (p=0.74), but decreased significantly after 60 min when using tubes smaller than 3.6 mL (p<0.0001). Conclusions: RoTEM (R) analyses were not affected by the size of blood sampling tubes. Therefore, 1.8 mL tubes should be preferred for RoTEM (R) analyses in order to minimise the volume of blood drawn. With regard to platelet aggregation analysed by impedance aggregometry tubes of different size cannot be used interchangeably. If platelet count is determined later than 10 min after blood sampling using tubes containing citrate as anticoagulant, a 3.6 mL tube should be preferred to avoid false low values.
引用
收藏
页码:701 / 706
页数:6
相关论文
共 50 条
  • [31] Evaluation of platelet function using multiple electrode platelet aggregometry in dogs with septic peritonitis
    Li, Ronald H. L.
    Chan, Daniel L.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2016, 26 (05) : 630 - 638
  • [32] Structure and Function of Platelet Receptors Initiating Blood Clotting
    Gardiner, Elizabeth E.
    Andrews, Robert K.
    SYSTEMS BIOLOGY APPROACH TO BLOOD, 2014, 844 : 263 - 275
  • [33] Platelet Function Assay Using Dielectric Blood Coagulometry
    Hayashi, Yoshihito
    Yamamoto, Yudai
    Murata, Aya
    Machida, Kenzo
    Katsumoto, Yoichi
    Uchida, Tokujiro
    ANALYTICAL CHEMISTRY, 2025, 97 (04) : 2036 - 2043
  • [34] Prostaglandin E2 levels and platelet function are different in cord blood compared to adults
    Schlagenhauf, Axel
    Haidl, Harald
    Leschnik, Bettina
    Leis, Hans-Joerg
    Heinemann, Akos
    Muntean, Wolfgang
    THROMBOSIS AND HAEMOSTASIS, 2015, 113 (01) : 97 - 106
  • [35] Evaluation of angiographic contrast media and platelet function with impedance aggregometry and the PFA-100™ 'platelet function analyser'
    Dalby, MCD
    Dewar, A
    Davidson, SJ
    Burman, JF
    Davies, SW
    PLATELETS, 1999, 10 (05) : 293 - 297
  • [36] Evaluation of platelet function in horses undergoing colic surgery using the PFA-100 platelet function analyser
    Iwaszko-Simonik, A.
    Graczyk, S.
    VETERINARNI MEDICINA, 2015, 60 (09) : 476 - 482
  • [37] Association of cord blood platelet count and volume with hemoglobin in healthy term infants
    Eskola, M.
    Juutistenaho, S.
    Aranko, K.
    Sainio, S.
    Kekomaki, R.
    JOURNAL OF PERINATOLOGY, 2011, 31 (04) : 258 - 262
  • [38] Investigations of Platelet Function in Whole Blood with BAPA as Anticoagulant
    Tauer, Josephine Tabea
    Lohse, Judith
    Koenig, Silvia
    Knoefler, Ralf
    TRANSFUSION MEDICINE AND HEMOTHERAPY, 2010, 37 (05) : 284 - 288
  • [39] Changes in platelet function, volume and count during labor and 24 hours postpartum
    Lurie, S
    Rigini, N
    Zabeeda, D
    Sadan, O
    Ezri, T
    Glezerman, M
    PLATELETS, 2003, 14 (06) : 355 - 358
  • [40] Peripheral versus central venous blood sampling does not influence the assessment of platelet activation in cirrhosis
    Brusilovskaya, Ksenia
    Simbrunner, Benedikt
    Lee, Silvia
    Eichelberger, Beate
    Bauer, David
    Zinober, Kerstin
    Schwabl, Philipp
    Mandorfer, Mattias
    Panzer, Simon
    Reiberger, Thomas
    Gremmel, Thomas
    PLATELETS, 2022, 33 (06) : 879 - 886