Evaluation of different sized blood sampling tubes for thromboelastometry, platelet function, and platelet count

被引:13
|
作者
Andreasen, Jo Bonding [2 ]
Pistor-Riebold, Thea Unger [3 ]
Knudsen, Ingrid Hell [3 ]
Ravn, Hanne Berg [2 ]
Hvas, Anne-Mette [1 ]
机构
[1] Aarhus Univ Hosp, Dept Clin Biochem, Ctr Haemophilia & Thrombosis, DK-8200 Aarhus N, Denmark
[2] Aarhus Univ Hosp, Dept Anaesthesiol & Intens Care, DK-8200 Aarhus N, Denmark
[3] Aarhus Univ, Fac Hlth Sci, Aarhus, Denmark
关键词
blood sampling; coagulation; paediatric; platelet aggregation; preanalytical variables; QUALITY STANDARDS; AGGREGOMETRY; TRANSFUSIONS; COLLECTION; TRANSPORT; EDTA; ROTEM(R); SYSTEM;
D O I
10.1515/cclm-2013-0836
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: To minimise the volume of blood used for diagnostic procedures, especially in children, we investigated whether the size of sample tubes affected whole blood coagulation analyses. Methods: We included 20 healthy individuals for rotational thromboelastometry (RoTEM (R)) analyses and compared three blood sampling tubes of different size: 1.8, 2.7, and 3.6 mL. All tubes were made of plastic and contained 3.2% sodium-citrate as anticoagulant. Platelet aggregation was investigated in 12 healthy individuals employing the Multiplate (R) Analyser comparing tubes of 3.6 mL and 1.8 mL. Platelet count was determined for each of the sampling tubes after 10, 60, and 120 min. Results: No significant differences were found in any of the RoTEM (R) measurements between the three sizes of tubes (p=0.07-0.53). Platelet aggregation was significantly lower when using smaller tubes (p=0.0004). The platelet count remained stable using a 3.6 mL tube during the entire observation period of 120 min (p=0.74), but decreased significantly after 60 min when using tubes smaller than 3.6 mL (p<0.0001). Conclusions: RoTEM (R) analyses were not affected by the size of blood sampling tubes. Therefore, 1.8 mL tubes should be preferred for RoTEM (R) analyses in order to minimise the volume of blood drawn. With regard to platelet aggregation analysed by impedance aggregometry tubes of different size cannot be used interchangeably. If platelet count is determined later than 10 min after blood sampling using tubes containing citrate as anticoagulant, a 3.6 mL tube should be preferred to avoid false low values.
引用
收藏
页码:701 / 706
页数:6
相关论文
共 50 条
  • [1] A new anticoagulant for blood sampling tubes for platelet function tests
    Wolf, Johanna
    Sigl, Martin
    Giannitsis, Evangelos
    Ivandic, Boris
    LABORATORIUMSMEDIZIN-JOURNAL OF LABORATORY MEDICINE, 2010, 34 (03): : 149 - 153
  • [2] Evaluation of the Platelet Mapping Assay on rotational thromboelastometry ROTEM
    Scharbert, G.
    Auer, A.
    Kozek-Langenecker, S.
    PLATELETS, 2009, 20 (02) : 125 - 130
  • [3] Beyond the platelet count: immature platelet fraction and thromboelastometry correlate with bleeding in patients with immune thrombocytopenia
    Greene, Lindsey A.
    Chen, Siqi
    Seery, Caroline
    Imahiyerobo, Allison M.
    Bussel, James B.
    BRITISH JOURNAL OF HAEMATOLOGY, 2014, 166 (04) : 592 - 600
  • [4] Impact of incorrect filling of citrate blood sampling tubes on thromboelastometry
    Haas, Thorsten
    Spielmann, Nelly
    Cushing, Melissa
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2015, 75 (08) : 717 - 719
  • [5] Evaluation of platelet function in thrombocytopenia
    Skipper, Mette Tiedemann
    Rubak, Peter
    Stentoft, Jesper
    Hvas, Anne-Mette
    Larsen, Ole Halfdan
    PLATELETS, 2018, 29 (03) : 270 - 276
  • [6] A whole blood model of thrombocytopenia that controls platelet count and hematocrit
    Bercovitz, R. S.
    Brenner, M. K.
    Newman, D. K.
    ANNALS OF HEMATOLOGY, 2016, 95 (11) : 1887 - 1894
  • [7] Diagnostic evaluation of platelet function disorders
    Hayward, Catherine P. M.
    BLOOD REVIEWS, 2011, 25 (04) : 169 - 173
  • [8] ON THE USE OF VACUATED CLOSED SYSTEM BLOOD-SAMPLING FOR EXVIVO EVALUATION OF PLATELET-FUNCTION IN MAN
    NAESH, O
    HINDBERG, I
    THROMBOSIS RESEARCH, 1990, 57 (04) : 665 - 668
  • [9] Correlation of rotational thromboelastometry (ROTEM) parameters with platelet count and their ability to predict thrombocytopenia in dogs
    Turner, Jennifer S.
    Kutter, Annette P. N.
    Sigrist, Nadja E.
    RESEARCH IN VETERINARY SCIENCE, 2019, 126 : 45 - 50
  • [10] Influence of platelet count, platelet mass index, and platelet function on the spontaneous closure of ductus arteriosus in the prematurity
    Kahvecioglu, Dilek
    Erdeve, Omer
    Akduman, Hasan
    Ucar, Tayfun
    Alan, Serdar
    Cakir, Ufuk
    Yildiz, Duran
    Atasay, Begum
    Arsan, Saadet
    Atalay, Semra
    PEDIATRICS AND NEONATOLOGY, 2018, 59 (01) : 53 - 57