Singular value description of a digital radiographic detector: Theory and measurements

被引:14
作者
Kyprianou, Iacovos S. [1 ]
Badano, Aldo [1 ]
Gallas, Brandon D. [1 ]
Myers, Kyle J. [1 ]
机构
[1] US FDA, NIBIB CDRH, Lab Assessment Med Imaging Syst, Silver Spring, MD 20993 USA
关键词
H matrix; MTF; system response; line spread function; edge method;
D O I
10.1118/1.2975222
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The H operator represents the deterministic performance of any imaging system. For a linear, digital imaging system, this system operator can be written in terms of a matrix, H, that describes the deterministic response of the system to a set of point objects. A singular value decomposition of this matrix results in a set of orthogonal functions (singular vectors) that form the system basis. A linear combination of these vectors completely describes the transfer of objects through the linear system, where the respective singular values associated with each singular vector describe the magnitude with which that contribution to the object is transferred through the system. This paper is focused on the measurement, analysis, and interpretation of the H matrix for digital x-ray detectors. A key ingredient in the measurement of the H matrix is the detector response to a single x ray (or infinitestimal x-ray beam). The authors have developed a method to estimate the 2D detector shift-variant, asymmetric ray response function (RRF) from multiple measured line response functions (LRFs) using a modified edge technique. The RRF measurements cover a range of x-ray incident angles from 0 degrees (equivalent location at the detector center) to 30 degrees (equivalent location at the detector edge) for a standard radiographic or cone-beam CT geometric setup. To demonstrate the method, three beam qualities were tested using the inherent, Lu/Er, and Yb beam filtration. The authors show that measures using the LRF, derived from an edge measurement, underestimate the system's performance when compared with the H matrix derived using the RRF. Furthermore, the authors show that edge measurements must be performed at multiple directions in order to capture rotational asymmetries of the RRF. The authors interpret the results of the H matrix SVD and provide correlations with the familiar MTF methodology. Discussion is made about the benefits of the H matrix technique with regards to signal detection theory, and the characterization of shift-variant imaging systems.
引用
收藏
页码:4744 / 4756
页数:13
相关论文
共 25 条
[1]  
BADANO A, 2006, P SPIE, V6142
[2]   Anisotropic imaging performance in indirect x-ray imaging detectors [J].
Badano, Aldo ;
Kyprianou, Iacovos S. ;
Sempau, Josep .
MEDICAL PHYSICS, 2006, 33 (08) :2698-2713
[3]   OBJECTIVE ASSESSMENT OF IMAGE QUALITY .2. FISHER INFORMATION, FOURIER CROSSTALK, AND FIGURES OF MERIT FOR TASK-PERFORMANCE [J].
BARRETT, HH ;
DENNY, JL ;
WAGNER, RF ;
MYERS, KJ .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1995, 12 (05) :834-852
[4]  
BARRETT HH, 1991, PROG CLIN BIOL RES, V363, P211
[5]  
Barrett HH., 2004, FDN IMAGE SCI
[6]   AN ANALYTICAL EDGE SPREAD FUNCTION MODEL FOR COMPUTER FITTING AND SUBSEQUENT CALCULATION OF THE LSF AND MTF [J].
BOONE, JM ;
SEIBERT, JA .
MEDICAL PHYSICS, 1994, 21 (10) :1541-1545
[7]   Simple method for modulation transfer function determination of digital imaging detectors from edge images [J].
Buhr, E ;
Günther-Kohfahl, S ;
Neitzel, U .
MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 :877-884
[8]   A METHOD FOR MODULATION TRANSFER-FUNCTION DETERMINATION FROM EDGE PROFILES WITH CORRECTION FOR FINITE-ELEMENT DIFFERENTIATION [J].
CUNNINGHAM, IA ;
FENSTER, A .
MEDICAL PHYSICS, 1987, 14 (04) :533-537
[9]  
Dainty J.C., 1974, Image Science: Principles, Analysis and Evaluation of Photographic-type Imaging Processes
[10]   REDUCTION OF ALIASING IN MODULATION TRANSFER-FUNCTION MEASUREMENTS [J].
EHRHARDT, JC .
MEDICAL PHYSICS, 1986, 13 (05) :658-662