Electrocardiogram Provides a Continuous Heart Rate Faster Than Oximetry During Neonatal Resuscitation

被引:106
|
作者
Katheria, Anup [1 ]
Rich, Wade [1 ]
Finer, Neil [1 ]
机构
[1] Univ Calif San Diego, Div Neonatol, UCSD Med Ctr, San Diego, CA 92103 USA
关键词
ECG; heart rate; pulse oximetry; resuscitation; very low birth weight; PULSE OXIMETRY; DELIVERY ROOM; ACCURACY; INFANTS;
D O I
10.1542/peds.2012-0784
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
OBJECTIVE: To compare the time required to obtain a continuous audible heart rate signal from an electrocardiogram (ECG) monitor and pulse oximeter (PO) in infants requiring resuscitation. METHODS: Infants who had both ECG and PO placed during resuscitation were analyzed using video and analog recordings. The median times from arrival until the ECG electrodes and PO sensor were placed, and the time to achieve audible tones from the devices, were compared. RESULTS: Forty-six infants had ECG and PO data. Thirty infants were very low birth weight (23-30 weeks). There was a difference in the median total time to place either device (26 vs 38 seconds; P = .04), and a difference (P < .001) in the time to achieve an audible heart rate signal after ECG lead (2 seconds) versus PO probe (24 seconds) placement. In infants weighing >1500 g (n = 16), the median time (interquartile range) to place the ECG was 20 seconds (14-43) whereas the time to place the PO was 36 seconds (28-56) (P = .74). The median times (interquartile range) to acquire a signal from the ECG and PO were 4 seconds (1-6) and 32 seconds (15-40, P = .001), respectively. During the first minutes of resuscitation, 93% of infants had an ECG heart rate compared with only 56% for PO. CONCLUSIONS: Early application of ECG electrodes during infant resuscitation can provide the resuscitation team with a continuous audible heart rate, and its use may improve the timeliness of appropriate critical interventions. Pediatrics 2012;130:e1177-e1181
引用
收藏
页码:E1177 / E1181
页数:5
相关论文
共 50 条
  • [1] Electrocardiogram shows reliable heart rates much earlier than pulse oximetry during neonatal resuscitation
    Mizumoto, Hiroshi
    Tomotaki, Seiichi
    Shibata, Hirofumi
    Ueda, Kazutoshi
    Akashi, Ryoko
    Uchio, Hiroko
    Hata, Daisuke
    PEDIATRICS INTERNATIONAL, 2012, 54 (02) : 205 - 207
  • [2] Heart Rate Assessment during Neonatal Resuscitation
    Johnson, Peter A.
    Schmolzer, Georg M.
    HEALTHCARE, 2020, 8 (01)
  • [3] Pulse oximetry versus electrocardiogram for heart rate assessment during resuscitation of the preterm infant
    Iglesias, B.
    Rodriguez, M. J.
    Aleo, E.
    Criado, E.
    Herranz, G.
    Moro, M.
    Martinez Orgado, J.
    Arruza, L.
    ANALES DE PEDIATRIA, 2016, 84 (05): : 271 - 277
  • [4] Evaluation of a Tap-Based Smartphone App for Heart Rate Assessment During Asphyxia in a Porcine Model of Neonatal Resuscitation
    Johnson, Peter A.
    Morina, Nicolo
    O'Reilly, Megan
    Lee, Tze-Fun
    Cheung, Po-Yin
    Schmolzer, Georg M.
    FRONTIERS IN PEDIATRICS, 2019, 7
  • [5] Reliability of Pulse Oximetry during Cardiopulmonary Resuscitation in a Piglet Model of Neonatal Cardiac Arrest
    Hassan, Mohammad Ahmad
    Mendler, Marc
    Maurer, Miriam
    Waitz, Markus
    Huang, Li
    Hummler, Helmut D.
    NEONATOLOGY, 2015, 107 (02) : 113 - 119
  • [6] Pedi-cap color change precedes a significant increase in heart rate during neonatal resuscitation
    Blank, Doug
    Rich, Wade
    Leone, Tina
    Garey, Donna
    Finer, Neil
    RESUSCITATION, 2014, 85 (11) : 1568 - 1572
  • [7] A Pilot Randomized Trial of Heart Rate Monitoring Using Conventional Versus a New Electrocardiogram Algorithm during Neonatal Resuscitation at Birth
    Katheria, Anup C.
    Morales, Ana
    Shashank, Sanjay
    Rich, Wade D.
    Finer, Neil N.
    JOURNAL OF PEDIATRICS, 2022, 242 : 245 - +
  • [8] Compliance with Guidelines and Efficacy of Heart Rate Monitoring during Newborn Resuscitation: A Prospective Video Study
    Bjorland, Peder Aleksander
    Ersdal, Hege Langli
    Oymar, Knut
    Rettedal, Siren Irene
    NEONATOLOGY, 2020, 117 (02) : 175 - 181
  • [9] Reliability of Pulse Oximetry during Progressive Hypoxia, Cardiopulmonary Resuscitation, and Recovery in a Piglet Model of Neonatal Hypoxic Cardiac Arrest
    Hassan, Mohammad Ahmad
    Weber, Claudia
    Waitz, Markus
    Huang, Li
    Hummler, Helmut D.
    Mendler, Marc Robin
    NEONATOLOGY, 2017, 112 (01) : 40 - 46
  • [10] 3-lead electrocardiogram is more reliable than pulse oximetry to detect bradycardia during stabilisation at birth of very preterm infants
    Iglesias, Beatriz
    Jose Rodriguez, Maria
    Aleo, Esther
    Criado, Enrique
    Martinez-Orgado, Jose
    Arruza, Luis
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2018, 103 (03): : F233 - F237