NC-stat sensory nerve conduction studies in the median and ulnar nerves of symptomatic patients

被引:29
作者
Kong, X
Gozani, SN
Hayes, MT
Weinberg, DH
机构
[1] NEUROMetrix Inc, Waltham, MA 02451 USA
[2] St Elizabeth Hosp, Dept Neurol, Boston, MA 02135 USA
关键词
NC-stat; inter-examiner; comparison; point-of-service; nerve conduction studies (NCS); median sensory nerve; ulnar sensory nerve;
D O I
10.1016/j.clinph.2005.10.019
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: This study evaluated validity and reliability of automated median and ulnar sensory nerve conduction study (NCS) measurements by the NC-stat((R)). Methods: Median and ulnar distal sensory latencies (DSL) and amplitudes (SNAP) were measured in sixty subjects with the NC-stat and by a neurologist (reference) using traditional instrumentation. The median-ulnar DSL differences (MUD) was calculated. Validity was quantified by the Pearson correlation. Reliability was evaluated by the intraclass correlation coefficient (ICC), Bland-Altman analysis, and inter-rater agreement of MUD abnormalities. Results: As a result of differences in electrode placement, NC-stat and reference mean values had systematic differences. The correlation ranged from 0.70 (ulnar DSL) to 0.91 (median DSL). The ICC ranged from 0.69 (ulnar DSL) to 0.91 (median DSL). In Bland-Altman analysis of DSLs, NC-stat measurements had a bias of 0.56 ms (median) and 0.31 ms (ulnar) and precision of 0.31 and 0.30 ms. Inter-rater agreement for MUD abnormalities was 93.8% (raw) and 0.80 (Kappa). Conclusions: NC-stat validity and reliability metrics were similar to traditional NCS. Use of the NC-stat would require applicable reference ranges. Significance: NC-stat median and ulnar NCS are valid and reliable. This device may be useful for increasing availability of NCS when clinically appropriate. (c) 2005 International Federation of Clinical Neurophysiology. Published by Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:405 / 413
页数:9
相关论文
共 33 条
  • [1] [Anonymous], 1998, INTRO BOOTSTRAP
  • [2] BISCHOFF C, 1990, RECOMMENDATIONS PR S, V52, P199
  • [3] COMPARING METHODS OF MEASUREMENT - WHY PLOTTING DIFFERENCE AGAINST STANDARD METHOD IS MISLEADING
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1995, 346 (8982): : 1085 - 1087
  • [4] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [5] Bril V, 1998, MUSCLE NERVE, V21, P1368, DOI 10.1002/(SICI)1097-4598(199811)21:11<1368::AID-MUS2>3.3.CO
  • [6] 2-G
  • [7] INTEREXAMINER AND INTRAEXAMINER RELIABILITY OF NERVE-CONDUCTION MEASUREMENTS IN PATIENTS WITH DIABETIC NEUROPATHY
    CHAUDHRY, V
    CORSE, AM
    FREIMER, ML
    GLASS, JD
    MELLITS, ED
    KUNCL, RW
    QUASKEY, SA
    CORNBLATH, DR
    [J]. NEUROLOGY, 1994, 44 (08) : 1459 - 1462
  • [8] INTER-EXAMINER AND INTRA-EXAMINER RELIABILITY OF NERVE-CONDUCTION MEASUREMENTS IN NORMAL SUBJECTS
    CHAUDHRY, V
    CORNBLATH, DR
    MELLITS, ED
    AVILA, O
    FREIMER, ML
    GLASS, JD
    REIM, J
    RONNETT, GV
    QUASKEY, SA
    KUNCL, RW
    [J]. ANNALS OF NEUROLOGY, 1991, 30 (06) : 841 - 843
  • [9] THE RELATIVE EXCITABILITY AND CONDUCTION VELOCITY OF SENSORY AND MOTOR NERVE FIBRES IN MAN
    DAWSON, GD
    [J]. JOURNAL OF PHYSIOLOGY-LONDON, 1956, 131 (02): : 436 - 451
  • [10] TREATMENT OF CARPAL-TUNNEL SYNDROME BY MEMBERS OF THE AMERICAN-SOCIETY-FOR-SURGERY-OF-THE-HAND - RESULTS OF A QUESTIONNAIRE
    DUNCAN, KH
    LEWIS, RC
    FOREMAN, KA
    NORDYKE, MD
    [J]. JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1987, 12A (03): : 384 - 391