Electrochemical characterization of screen-printed carbonaceous electrodes for the determination of peroxidase activity in novel screen-printed flow-through modules

被引:19
作者
Stiene, M [1 ]
Bilitewski, U [1 ]
机构
[1] German Res Ctr Biotechnol, D-38124 Braunschweig, Germany
关键词
screen-printed flow-through electrode; thin layer affinity reactor; electron transfer rate; p-benzoquinone detection; African Swine Fever Virus;
D O I
10.1007/s00216-001-1208-4
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel totally screen-printed flow-through cell for immunoanalysis is presented. It contained screen-printed carbonaceous electrodes, which allowed the determination of peroxidase activity through the electrochemical reduction of p-benzoquinone. As different electrode materials differ strongly in their electrochemical properties, electrodes resulting from various screen-printable carbonaceous pastes were characterized using the hydroquinone/ p-benzoquinone redox couple. For most of the electrodes, cyclic voltammogram peak separations of between 550 and 670 mV were observed indicating only quasi-reversible electrochemical behavior. This was confirmed by variation of the peak separation with scan rate. Heterogeneous electron transfer rates of ca. 0.5-1x10(-3) cm s(-1) and electrochemical activation energies of ca. 20 kJ mol(-1) were found. These flow-through cells were not only applied to electrochemical peroxidase activity determinations but also, in combination with a separate detector, as affinity reactors. After biotinylation of screen-printed layers, streptavidin and then biotinylated peroxidase could be bound. However, as signals were only 10-20% of those obtained with a column filled with biotinylated glass beads, only the screen-printed electrochemical detector was applied to the detection of antibodies against the African Swine Fever Virus.
引用
收藏
页码:240 / 247
页数:8
相关论文
共 45 条
[1]   Direct electrochemistry of cytochrome c at plain and membrane modified screen-printed carbon electrodes [J].
Abass, AK ;
Hart, JP .
ELECTROCHIMICA ACTA, 2001, 46 (06) :829-836
[2]   A disposable multielectrode biosensor for rapid simultaneous detection of the insecticides paraoxon and carbofuran at high resolution [J].
Bachmann, TT ;
Schmid, RD .
ANALYTICA CHIMICA ACTA, 1999, 401 (1-2) :95-103
[3]   Electrochemistry in nanovials fabricated by combining screen printing and laser micromachining [J].
Ball, JC ;
Scott, DL ;
Lumpp, JK ;
Daunert, S ;
Wang, J ;
Bachas, LG .
ANALYTICAL CHEMISTRY, 2000, 72 (03) :497-501
[4]  
Bard A. J., 1980, ELECTROCHEMICAL METH
[5]   Zeptomole-detecting biosensor for alkaline phosphatase in an electrochemical immunoassay for 2,4-dichlorophenoxyacetic acid [J].
Bauer, CG ;
Eremenko, AV ;
EhrentreichForster, E ;
Bier, FF ;
Makower, A ;
Halsall, HB ;
Heineman, WR ;
Scheller, FW .
ANALYTICAL CHEMISTRY, 1996, 68 (15) :2453-2458
[6]   A redox-label immunosensor on basis of a bi-enzyme electrode [J].
Bier, FF ;
EhrentreichForster, E ;
Dolling, R ;
Eremenko, AV ;
Scheller, FW .
ANALYTICA CHIMICA ACTA, 1997, 344 (1-2) :119-124
[7]   ENZYME ELECTRODES FOR THE DETERMINATION OF CARBOHYDRATES IN FOOD [J].
BILITEWSKI, U ;
JAGER, A ;
RUGER, P ;
WEISE, W .
SENSORS AND ACTUATORS B-CHEMICAL, 1993, 15 (1-3) :113-118
[8]  
BILITEWSKI U, 1998, Patent No. 9800167
[9]   Amperometric immunosensor for granulocyte-macrophage colony-stimulating factor using screen-printed electrodes [J].
Crowley, E ;
O'Sullivan, C ;
Guilbault, GG .
ANALYTICA CHIMICA ACTA, 1999, 389 (1-3) :171-178
[10]   An immunomagnetic electrochemical sensor based an a perfluorosulfonate-coated screen-printed electrode for the determination of 2,4-dichlorophenoxyacetic acid [J].
Dequaire, M ;
Degrand, C ;
Limoges, B .
ANALYTICAL CHEMISTRY, 1999, 71 (13) :2571-2577