Conditions for efficient on-chip magnetic bead detection via magnetoresistive sensors

被引:26
作者
Albisetti, E. [1 ,2 ]
Petti, D. [1 ]
Cantoni, M. [1 ]
Damin, F. [2 ]
Torti, A. [1 ]
Chiari, M. [2 ]
Bertacco, R. [1 ]
机构
[1] Politecn Milan, Dipartimento Fis, L NESS, I-22100 Como, Italy
[2] CNR, Ist Chim Riconoscimento Mol, I-20131 Milan, Italy
关键词
Magnetic biosensor; Magnetic tunneling junction; Magnetic bead; SHAPE-ANISOTROPY; TUNNEL-JUNCTIONS; MICROARRAY;
D O I
10.1016/j.bios.2013.03.016
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A commonly used figure of merit of magnetoresistive sensors employed to detect magnetic beads labeling biomolecules in lab-on-chip applications is the sensor sensitivity (So) to external magnetic fields in the linear region of the sensor. In this paper we show that, in case of lock-in detection and bead excitation by a small AC magnetic field, S-0 is not the good figure of merit to optimize. Indeed, the highest sensitivity to the magnetic beads is achieved biasing the sensor in the region of its characteristics where the product between the DC bias field and the second derivative of the resistance with respect to the magnetic field is maximum. The validity of this criterion, derived from a phenomenological model of bead detection, is proved in case of magnetic tunneling junction sensors detecting magnetic beads with 250 nm diameter. This work paves the way to the development of a new generation of sensors properly designed to maximize the bead sensitivity. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:213 / 217
页数:5
相关论文
共 14 条
[1]  
Cardoso F., 2011, THESIS U TECNICA LIS
[2]   Single molecule actuation and detection on a lab-on-a-chip magnetoresistive platform [J].
Chaves, R. C. ;
Bensimon, D. ;
Freitas, P. P. .
JOURNAL OF APPLIED PHYSICS, 2011, 109 (06)
[3]   On-chip measurement of the Brownian relaxation frequency of magnetic beads using magnetic tunneling junctions [J].
Donolato, M. ;
Sogne, E. ;
Dalslet, B. T. ;
Cantoni, M. ;
Petti, D. ;
Cao, J. ;
Cardoso, F. ;
Cardoso, S. ;
Freitas, P. P. ;
Hansen, M. F. ;
Bertacco, R. .
APPLIED PHYSICS LETTERS, 2011, 98 (07)
[4]   Effect of spin-valve sensor magnetostatic fields on nanobead detection for biochip applications [J].
Ferreira, HA ;
Feliciano, N ;
Graham, DL ;
Freitas, PP .
JOURNAL OF APPLIED PHYSICS, 2005, 97 (10)
[5]  
Gaster RS, 2011, NAT NANOTECHNOL, V6, P314, DOI [10.1038/NNANO.2011.45, 10.1038/nnano.2011.45]
[6]   A novel zero-drift detection method for highly sensitive GMR biochips [J].
Han, S. -J. ;
Xu, L. ;
Wilson, R. J. ;
Wang, S. X. .
IEEE TRANSACTIONS ON MAGNETICS, 2006, 42 (10) :3560-3562
[7]   Theoretical study of in-plane response of magnetic field sensor to magnetic beads magnetized by the sensor self-field [J].
Hansen, Troels Borum Grave ;
Damsgaard, Christian Danvad ;
Dalslet, Bjarke Thomas ;
Hansena, Mikkel Fougt .
JOURNAL OF APPLIED PHYSICS, 2010, 107 (12)
[8]   Shape-anisotropy-controlled magnetoresistive response in magnetic tunnel junctions [J].
Lu, Y ;
Altman, RA ;
Marley, A ;
Rishton, SA ;
Trouilloud, PL ;
Xiao, G ;
Gallagher, WJ ;
Parkin, SSP .
APPLIED PHYSICS LETTERS, 1997, 70 (19) :2610-2612
[9]   Femtomolar limit of detection with a magnetoresistive biochip [J].
Martins, V. C. ;
Cardoso, F. A. ;
Germano, J. ;
Cardoso, S. ;
Sousa, L. ;
Piedade, M. ;
Freitas, P. P. ;
Fonseca, L. P. .
BIOSENSORS & BIOELECTRONICS, 2009, 24 (08) :2690-2695
[10]   Characterization of a polymeric adsorbed coating for DNA microarray glass slides [J].
Pirri, G ;
Damin, F ;
Chiari, M ;
Bontempi, E ;
Depero, LE .
ANALYTICAL CHEMISTRY, 2004, 76 (05) :1352-1358