Characterization and in vivo evaluation of a bio-corrodible nitrided iron stent

被引:69
作者
Feng, Qimao [1 ]
Zhang, Deyuan [2 ]
Xin, Chaohua [2 ]
Liu, Xiangdong [2 ]
Lin, Wenjiao [2 ]
Zhang, Wanqian [2 ]
Chen, Sun [3 ]
Sun, Kun [3 ]
机构
[1] Shanghai Univ Tradit Chinese Med, Dept Cardiol, Shanghai Hosp Tradit Chinese Med, Shanghai 200071, Peoples R China
[2] Lifetech Sci Shenzhen Co Ltd, R&D Ctr, Shenzhen 518057, Peoples R China
[3] Shanghai Jiao Tong Univ, Sch Med, Childrens Heart Ctr, Xinhua Hosp, Shanghai 200092, Peoples R China
关键词
PURE IRON; CORROSION-RESISTANCE; BALLOON ANGIOPLASTY; PULMONARY-ARTERY; ELUTING STENTS; IMPLANTATION; TERM; COARCTATION; BIOCOMPATIBILITY; REDILATATION;
D O I
10.1007/s10856-012-4823-z
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A bio-corrodible nitrided iron stent was developed using a vacuum plasma nitriding technique. In the nitrided iron stents, the tensile strength, radial strength, stiffness and in vitro electrochemical corrosion rate were significantly increased compared with those of the control pure iron stent. To evaluate its performance in vivo, the deployment of the nitrided iron stents in juvenile pig iliac arteries was performed. At 3 or 6 months postoperatively, the stented vessels remained patent well; however, slight luminal loss resulting from intimal hyperplasia and relative stenosis of the stented vessel segment with piglets growth were observed by 12 months; no thrombosis or local tissue necrosis was found. At 1 month postoperatively, a nearly intact layer of endothelial cells formed on the stented vessel wall. Additionally, a decreased inflammation scoring, considerably corroded struts and corrosion products accumulation were seen. These findings indicate the potential of this nitrided iron stent as an attractive biodegradable stent.
引用
收藏
页码:713 / 724
页数:12
相关论文
共 45 条
[1]   Medical progress: Disorders of iron metabolism [J].
Andrews, NC .
NEW ENGLAND JOURNAL OF MEDICINE, 1999, 341 (26) :1986-1995
[2]   Stent implantation and balloon angioplasty for treatment of branch pulmonary artery stenosis in children [J].
Baerlocher, Lorenz ;
Kretschmar, Oliver ;
Harpes, Paul ;
Arbenz, Urs ;
Berger, Felix ;
Knirsch, Walter .
CLINICAL RESEARCH IN CARDIOLOGY, 2008, 97 (05) :310-317
[3]  
Bayer U, 2011, US Patent, Patent No. [20110112628A1, 20110112628]
[4]  
Bornapour M, 2012, ACTA BIOMATER
[5]   Covered Cheatham-Platinum Stents for Serial Dilation of Severe Native Aortic Coarctation [J].
Bruckheimer, Elchanan ;
Dagan, Tamir ;
Arnir, Gabriel ;
Birk, Einat .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2009, 74 (01) :117-123
[6]   The microstructure and properties of commercial pure iron modified by plasma nitriding [J].
Chen, Chang-Zi ;
Shi, Xing-Hua ;
Zhang, Peng-Cheng ;
Bai, Bin ;
Leng, Yong-Xiang ;
Huang, Nan .
SOLID STATE IONICS, 2008, 179 (21-26) :971-974
[7]   Improved hardness and corrosion resistance of iron by Ti/TiN multilayer coating and plasma nitriding duplex treatment [J].
Chen, Chang-Zi ;
Li, Qiang ;
Leng, Yong-Xiang ;
Chen, J. Y. ;
Zhang, Peng-Cheng ;
Bai, Bin ;
Huang, Nan .
SURFACE & COATINGS TECHNOLOGY, 2010, 204 (18-19) :3082-3086
[8]   The efficacy and safety of stent redilatation in congenital heart disease [J].
Duke, C ;
Rosenthal, E ;
Qureshi, SA .
HEART, 2003, 89 (08) :905-912
[9]  
El-Omar Magdi M., 2001, Curr Interv Cardiol Rep, V3, P296
[10]   Novel growth Stent for the permanent treatment of vessel stenosis in growing children: An experimental study [J].
Ewert, P ;
Riesenkampff, E ;
Neuss, M ;
Kretschmar, O ;
Nagdyman, N ;
Lange, PE .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 62 (04) :506-510