The System Design and Evaluation of a 7-DOF Image-Guided Venipuncture Robot

被引:42
作者
Balter, Max L. [1 ,2 ]
Chen, Alvin I. [1 ,2 ]
Maguire, Timothy J. [1 ,2 ]
Yarmush, Martin L. [1 ,3 ]
机构
[1] Rutgers State Univ, Dept Biomed Engn, Piscataway, NJ 08854 USA
[2] VascuLogic LLC, Piscataway, NJ 08854 USA
[3] Massachusetts Gen Hosp, Ctr Engn Med, Boston, MA 02108 USA
关键词
Computer vision; kinematics; mechanism design; medical robots and systems; visual servoing;
D O I
10.1109/TRO.2015.2452776
中图分类号
TP24 [机器人技术];
学科分类号
080202 ; 1405 ;
摘要
Accessing the venous bloodstream to deliver fluids or obtain a blood sample is the most common clinical routine practiced in the U.S. Practitioners continue to rely on manual venipuncture techniques, but success rates are heavily dependent on clinician skill and patient physiology. In the U.S., failure rates can be as high as 50% in difficult patients, making venipuncture the leading cause of medical injury. To improve the rate of first-stick success, we have developed a portable autonomous venipuncture device that robotically servos a needle into a suitable vein under image guidance. The device operates in real time, combining near-infrared and ultrasound imaging, image analysis, and a 7-degree-of-freedom (DOF) robotic system to perform the venipuncture. The robot consists of a 3-DOF gantry to image the patient's peripheral forearm veins and a miniaturized 4-DOF serial arm to guide the cannula into the selected vein under closed-loop control. In this paper, we present the system architecture of the robot and evaluate the accuracy and precision through tracking, free-space positioning, and in vitro phantom cannulation experiments. The results demonstrate sub-millimeter accuracy throughout the operating workspace of the manipulator and a high rate of success when cannulating phantom veins in a skin-mimicking tissue model.
引用
收藏
页码:1044 / 1053
页数:10
相关论文
共 30 条
[1]   SEEDED REGION GROWING [J].
ADAMS, R ;
BISCHOF, L .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1994, 16 (06) :641-647
[2]  
[Anonymous], 2012, 31 9097 PHLEB
[3]   Optical properties of human skin, subcutaneous and mucous tissues in the wavelength range from 400 to 2000 nm [J].
Bashkatov, AN ;
Genina, EA ;
Kochubey, VI ;
Tuchin, VV .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2005, 38 (15) :2543-2555
[4]   A Novel Manipulator for Percutaneous Needle Insertion: Design and Experimentation [J].
Bassan, Harmanpreet S. ;
Patel, Rajni V. ;
Moallem, Mehrdad .
IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2009, 14 (06) :746-761
[5]   Medical Robots: Current Systems and Research Directions [J].
Beasley, Ryan A. .
JOURNAL OF ROBOTICS, 2012, 2012
[6]   FAST-SCAN CONVERSION ALGORITHMS FOR DISPLAYING ULTRASOUND SECTOR IMAGES [J].
BERKHOFF, AP ;
HUISMAN, HJ ;
THIJSSEN, JM ;
JACOBS, EMGP ;
HOMAN, RJF .
ULTRASONIC IMAGING, 1994, 16 (02) :87-108
[7]  
Bouguet J.-Y., 2013, CAMERA CALIBRATION T
[8]  
Campbell J, 1997, Prof Nurse, V12, pS10
[9]   Portable robot for autonomous venipuncture using 3D near infrared image guidance [J].
Chen, Alvin ;
Nikitczuk, Kevin ;
Nikitczuk, Jason ;
Maguire, Tim ;
Yarmush, Martin .
TECHNOLOGY, 2013, 1 (01) :72-87
[10]   Laboratory study on needle-tissue interaction: Towards the development of an instrument for automatic venipuncture [J].
De Boer, T. ;
Steinbuch, M. ;
Neerken, S. ;
Kharin, A. .
JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2007, 7 (03) :325-335