Whole-body dose equivalent including neutrons is similar for 6 MV and 15 MV IMRT, VMAT, and 3D conformal radiotherapy

被引:20
作者
Hauri, Pascal [1 ,2 ]
Schneider, Uwe [1 ,2 ]
机构
[1] Univ Zurich, Dept Phys, Zurich, Switzerland
[2] Hirslanden Med Ctr, Radiotherapy Hirslanden, Aarau, Switzerland
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2019年 / 20卷 / 03期
关键词
radiotherapy; 15 MV vs 6 MV; out-of-field dose; neutron dose equivalent; MODULATED RADIATION-THERAPY; DYNAMIC MLC IMRT; PHOTON BEAMS; CONSTITUENT COMPONENTS; DELIVERY TECHNIQUES; TREATMENT MACHINES; SPECTRA; MODEL; DOSIMETERS; SCATTER;
D O I
10.1002/acm2.12543
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose This study investigates the difference in whole-body dose equivalent between 6 and 15 MV image-guided radiotherapy (IGRT) for the treatment of a rhabdomyosarcoma in the prostate. Methods A previously developed model for stray radiation of the primary beam was improved and used to calculate the photon dose and photon energy in the out-of-field region for a radiotherapy patient. The dose calculated by the treatment planning system was fused with the model-calculated out-of-field dose, resulting in a whole-body photon dose distribution. The peripheral neutron dose equivalent was calculated using an analytical model from the literature. A daily cone beam CT dose was added to the neutron and photon dose equivalents. The calculated 3D dose distributions were compared to independent measurements conducted with thermoluminescence dosimeters and an anthropomorphic phantom. The dose contributions from the IGRT treatments of three different techniques applied with two nominal X-ray energies were compared using dose equivalent volume histograms (DEVHs). Results The calculated and measured out-of-field whole-body dose equivalents for the IGRT treatments agreed within (9 +/- 10) % (mean and type A SD). The neutron dose equivalent was a minor contribution to the total out-of-field dose up to 50 cm from the isocenter. Further from the isocenter, head leakage was dominating inside the patient body, whereas the neutron dose equivalent contribution was important close to the surface. There were small differences between the whole-body DEVHs of the 6 and 15 MV treatments applied with the same technique, although the single scatter contributions showed large differences. Independent of the beam energy, the out-of-field dose of the volumetric-modulated arc therapy (VMAT) treatment was significantly lower than the dynamic intensity-modulated radiation therapy (IMRT) treatment. Conclusion The calculated whole-body dose helped to understand the importance of the dose contributions in different areas of the patient. Regarding radiation protection of the patient for IGRT treatments, the choice of beam energy is not important, whereas the treatment technique has a large influence on the out-of-field dose. If the patient is treated with intensity-modulated beams, VMAT should be used instead of dynamic IMRT in terms of radiation protection of the patient. In general, the developed models for photon and neutron dose equivalent calculation can be used for any patient geometry, tumor location, and linear accelerator.
引用
收藏
页码:56 / 70
页数:15
相关论文
共 44 条
  • [11] Hauri P, 2017, THESIS
  • [12] Whole-body dose and energy measurements in radiotherapy by a combination of LiF:Mg,Cu,P and LiF:Mg,Ti
    Hauri, Pascal
    Schneider, Uwe
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2018, 28 (02): : 96 - 109
  • [13] Technical note: No increase in effective dose from half compared to full rotation pelvis cone beam CT
    Hauri, Pascal
    Halg, Roger A.
    Schneider, Uwe
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (05): : 364 - 368
  • [14] Technical Note: Comparison of peripheral patient dose from MR-guided 60Co therapy and 6 MV linear accelerator IGRT
    Hauri, Pascal
    Halg, Roger A.
    Schneider, Uwe
    [J]. MEDICAL PHYSICS, 2017, 44 (07) : 3788 - 3793
  • [15] A general model for stray dose calculation of static and intensity-modulated photon radiation
    Hauri, Pascal
    Haelg, Roger A.
    Besserer, Juergen
    Schneider, Uwe
    [J]. MEDICAL PHYSICS, 2016, 43 (04) : 1955 - 1968
  • [16] HOROWITZ YS, 1988, RADIAT PROT DOSIM, V23, P401
  • [17] Calculation of effective dose from measurements of secondary neutron spectra and scattered photon dose from dynamic MLC IMRT for 6 MV, 15 MV, and 18 MV beam energies
    Howell, RM
    Hertel, NE
    Wang, ZL
    Hutchinson, J
    Fullerton, GD
    [J]. MEDICAL PHYSICS, 2006, 33 (02) : 360 - 368
  • [18] Investigation of secondary neutron dose for 18 MV dynamic MLC IMRT delivery
    Howell, RM
    Ferenci, MS
    Hertel, NE
    Fullerton, GD
    [J]. MEDICAL PHYSICS, 2005, 32 (03) : 786 - 793
  • [19] International Atomic Energy Agency (IAEA), 2001, TECHNICAL REPORTS SE, V398
  • [20] Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations
    Kim, Sangroh
    Alaei, Parham
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 153 - 164