Strength of poly-ether-ether-ketone: Effects of sterilisation and thermal ageing

被引:21
作者
Xin, H. [1 ]
Shepherd, D. E. T. [1 ]
Dearn, K. D. [1 ]
机构
[1] Univ Birmingham, Sch Mech Engn, Birmingham B15 2TT, W Midlands, England
关键词
Fatigue strength; Flexural strength; Gamma sterilisation; Poly-ether-ether-ketone (PEEK); Thermal ageing; FATIGUE FRACTURE-BEHAVIOR; TENSION-TENSION FATIGUE; MECHANICAL-PROPERTIES; REINFORCED POLYETHERETHERKETONE; SPINAL IMPLANTS; PEEK; COMPOSITES; POLY(ETHERETHERKETONE); BIOMATERIALS; TEMPERATURE;
D O I
10.1016/j.polymertesting.2013.05.012
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
This paper investigates the strength of polyether-ether-ketone (PEEK) after sterilisation and thermal ageing. PEEK specimens were divided into five groups, according to whether the specimens had been annealed, sterilised or aged. Specimens were subjected to either static or dynamic three-point bend tests. Static tests involved loading the specimens until a maximum displacement of 40 mm was reached. Dynamic tests involved applying a sinusoidally varying force at a frequency of 5 Hz. The maximum force applied to a specimen was based on a percentage of the static yield strength. Testing continued until failure or run out of 10 million cycles. Sterilisation and ageing resulted in no significant change in the static yield strength. Annealing was found to significantly increase the yield strength. For the dynamic tests, the fatigue strength was in the range 99.4 to 107.4 MPa; sterilisation and thermal ageing were found to have no effect on fatigue strength. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1001 / 1005
页数:5
相关论文
共 31 条
  • [1] Tensile properties, tension-tension fatigue and biological response of polyetheretherketone-hydroxyapatite composites for load-bearing orthopedic implants
    Abu Bakar, MS
    Cheng, MHW
    Tang, SM
    Yu, SC
    Liao, K
    Tan, CT
    Khor, KA
    Cheang, P
    [J]. BIOMATERIALS, 2003, 24 (13) : 2245 - 2250
  • [2] [Anonymous], PEEK PROC GUID
  • [3] [Anonymous], 2008, F200302 ASTM
  • [4] [Anonymous], MAT PROP GUID
  • [5] [Anonymous], 2011, F234605 ASTM
  • [6] Compression and tension fatigue resistance of medical grade ultra high molecular weight polyethylene: the effect of morphology, sterilization, aging and temperature
    Baker, DA
    Hastings, RS
    Pruitt, L
    [J]. POLYMER, 2000, 41 (02) : 795 - 808
  • [7] Beland S., 1990, HIGH PERFORMANCE THE
  • [8] FATIGUE FRACTURE-BEHAVIOR OF PEEK .1. EFFECTS OF LOAD LEVEL
    BRILLHART, M
    GREGORY, BL
    BOTSIS, J
    [J]. POLYMER, 1991, 32 (09) : 1605 - 1611
  • [9] FATIGUE FRACTURE-BEHAVIOR OF PEEK .2. EFFECTS OF THICKNESS AND TEMPERATURE
    BRILLHART, M
    BOTSIS, J
    [J]. POLYMER, 1992, 33 (24) : 5225 - 5232
  • [10] FATIGUE-CRACK GROWTH ANALYSIS IN PEEK
    BRILLHART, M
    BOTSIS, J
    [J]. INTERNATIONAL JOURNAL OF FATIGUE, 1994, 16 (02) : 134 - 140